Neonatal/infant validation study of the CAS model 740 noninvasive blood pressure monitor with the Orion/MaxIQ NIBP module.
MedLine Citation:
PMID:  24618916     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
OBJECTIVE: Blood pressure monitoring is an essential vital sign when caring for critically ill children. Invasive monitoring is considered the gold standard, but is not always feasible. The following study compared the CAS model 740 noninvasive blood pressure monitor with the Orion/MaxIQ NIBP module with the reference (invasive arterial measurement). We evaluated the validity of the system using the criteria provided by the Association for the Advancement of Medical Instrumentation.
RESULTS: We performed paired measurements of 29 critically ill neonates and children. For individual paired comparisons, the mean difference in the systolic blood pressure was 2.42 mmHg (SD ±6.3). The mean difference in the diastolic blood pressure was -1.29 mmHg (SD ±5.45). The percentage of readings within 5, 10, and 15 mmHg for systolic blood pressure was 65.6, 86.0, and 96.8%, respectively. The percentage of readings within 5, 10, and 15 mmHg for diastolic blood pressure was 77.7, 93, and 95.5%, respectively.
CONCLUSION: The CAS model 740 noninvasive blood pressure monitor with the Orion/MaxIQ NIBP module fulfills the accuracy performance criteria of the Association for the Advancement of Medical Instrumentation. This model may allow for rapid and accurate noninvasive blood pressure monitoring in neonates and children.
Authors:
Sean M Lang; John S Giuliano; Christopher L Carroll; Ted S Rosenkrantz; Leonard Eisenfeld
Related Documents :
22446226 - Distributed high-temperature pressure sensing using air-hole microstructural fibers.
24622426 - Correlation between balloon release pressure and no-reflow in patients with acute myoca...
8368936 - Transplantation and reanimation of hearts removed from donors 30 minutes after warm, as...
19559416 - Natural orifice transluminal endoscopic surgery gastrotomy closure with an over-the-end...
1815486 - Synthesis and pharmacological evaluation of 6-(7-theophylline)-3(2h)-pyridazinone.
20857796 - Safety, efficacy and acceptability of implanon a single rod implantable contraceptive (...
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2014-3-10
Journal Detail:
Title:  Blood pressure monitoring     Volume:  -     ISSN:  1473-5725     ISO Abbreviation:  Blood Press Monit     Publication Date:  2014 Mar 
Date Detail:
Created Date:  2014-3-12     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9606438     Medline TA:  Blood Press Monit     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Rational selection of substrates to improve color intensity and uniformity on microfluidic paper-bas...
Next Document:  Genetic Diversity and Population Demography of the Chinese Crocodile Lizard (Shinisaurus crocodiluru...