Is there an adverse outcome from abandoned pacing leads?
MedLine Citation:
PMID:  11046188     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
BACKGROUND: Indications for extraction of an abandoned pacemaker lead (APL) are controversial. The purpose of this study was to determine whether or not APLs should be extracted in the absence of pacemaker-related problems. METHODS AND RESULTS: We retrospectively reviewed, from 1977 through 1998, all patients with retained, non-functional leads and identified 433-266 males and 167 females. Mean age at initial pacemaker implantation was 68[emsp4 ]years. These patients received a total of 259 atrial and 948 ventricular leads. Of the total of 1,207 leads, 611 became non-functional. A total of 531 non-functional leads were abandoned, of which 18 were later extracted: one APL in 345 patients, two in 78, and three in 10. Indications for new lead placement when non-functional leads were abandoned included capture and/or sensing failure (243), lead recall (177), lead fracture (86), pacing system replacement to the contralateral side (11), accommodating patient growth (5), pacemaker function upgrade (5), replacement with implantable cardioverter defibrillator (ICD, 2), interference with ICD (1), and unknown (1). Complications that were associated with pacemakers were found in 24 patients (5.5%)-pacemaker system infection (8 patients) and venous occlusion at the time of a subsequent procedure of new lead placement when APLs had already been in place (16) which resulted in APL extraction (7) or transfer of the pacemaker system to the contralateral side (9). Neither venous thrombosis nor other complications were found in the remaining 409 patients (94.5%). The incidence of complications was higher in patients with three APLs than in patients with two or fewer APLs (40% vs. 4.7%, P=1x10(-6)), in patients with four or more total lead implantations than in patients with three or fewer total lead implantations (26.2% vs. 0. 6%, P<1x10(-10)), and in patients with three or more procedures of new lead placements than in patients with two or fewer procedures of new lead placements (36.4% vs. 3.9%, P=1x10(-10)). Patients with complications were younger than those without complications both at the time of initial pacemaker implantation (59+/-16 vs. 68+/-17 y, P=0.01) and when non-functional leads were abandoned (63+/-15 vs. 71+/-16 y, P=0.04). Mean numbers of APLs, total leads implanted, and procedures of new lead placement were significantly larger in patients with complications than in those without complications (1.58+/-0.78 vs. 1.2+/-0.44, 4.96+/-1.23 vs. 2.66+/-0.8, and 2.13+/-0.85 vs. 1.25+/-0.53, P=0.03, 4x10(-9) and 4x10(-5), respectively). CONCLUSIONS: 1. With only 5.5% of patients having had pacemaker-related complications, the adverse outcome of APL is small. 2. Clinical clues to the possible occasion for pacemaker-related complications include three or more APLs, four or more total leads, three or more procedures of new lead placement, and a younger age at initial pacemaker implantation. 3. Patients with a large number of APLs, total lead implantations, and procedures of new lead placement should be carefully observed to detect possible pacemaker-associated complications.
Authors:
C Suga; D L Hayes; L K Hyberger; M A Lloyd
Related Documents :
3528678 - Surgical ablation of posterior septal accessory pathways in the wolff-parkinson-white s...
6120648 - Reflex mechanisms responsible for early spontaneous termination of paroxysmal supravent...
6528268 - Spontaneous rotation of permanent pacemaker--a cause of muscle stimulation and myopoten...
9080488 - Interactions between transvenous nonthoracotomy cardioverter defibrillator systems and ...
452968 - Carcinoembryonic antigen (cea) levels in patients with brain tumours.
19403888 - Hematopoietic stem cell transplantation in griscelli syndrome type 2: a single-center r...
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing     Volume:  4     ISSN:  1383-875X     ISO Abbreviation:  J Interv Card Electrophysiol     Publication Date:  2000 Oct 
Date Detail:
Created Date:  2001-01-11     Completed Date:  2001-01-11     Revised Date:  2007-11-15    
Medline Journal Info:
Nlm Unique ID:  9708966     Medline TA:  J Interv Card Electrophysiol     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  493-9     Citation Subset:  IM    
Affiliation:
Division of Cardiovascular Diseases and Internal Medicine, Mayo Clinic and Mayo Foundation, Rochester, Minnesota 55905, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Aged
Arrhythmias, Cardiac / therapy*
Chi-Square Distribution
Equipment Design
Equipment Failure*
Equipment Safety
Female
Humans
Male
Middle Aged
Monitoring, Physiologic
Pacemaker, Artificial / adverse effects*
Probability
Retrospective Studies
Risk Assessment

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Demonstration of a His-Purkinje fatigue phenomenon with programmed stimulation of the right ventricu...
Next Document:  Activation effects of single-site and dual-site right atrial pacing in canine.