A systematic approach to designing reliable VV optimization methodology: Assessment of internal validity of echocardiographic, electrocardiographic and haemodynamic optimization of cardiac resynchronization therapy.
Jump to Full Text
MedLine Citation:
PMID:  22459364     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
BACKGROUND: In atrial fibrillation (AF), VV optimization of biventricular pacemakers can be examined in isolation. We used this approach to evaluate internal validity of three VV optimization methods by three criteria. METHODS AND RESULTS: Twenty patients (16 men, age 75±7) in AF were optimized, at two paced heart rates, by LVOT VTI (flow), non-invasive arterial pressure, and ECG (minimizing QRS duration). Each optimization method was evaluated for: singularity (unique peak of function), reproducibility of optimum, and biological plausibility of the distribution of optima. The reproducibility (standard deviation of the difference, SDD) of the optimal VV delay was 10ms for pressure, versus 8ms (p=ns) for QRS and 34ms (p<0.01) for flow. Singularity of optimum was 85% for pressure, 63% for ECG and 45% for flow (Chi(2)=10.9, p<0.005). The distribution of pressure optima was biologically plausible, with 80% LV pre-excited (p=0.007). The distributions of ECG (55% LV pre-excitation) and flow (45% LV pre-excitation) optima were no different to random (p=ns). The pressure-derived optimal VV delay is unaffected by the paced rate: SDD between slow and fast heart rate is 9ms, no different from the reproducibility SDD at both heart rates. CONCLUSIONS: Using non-invasive arterial pressure, VV delay optimization by parabolic fitting is achievable with good precision, satisfying all 3 criteria of internal validity. VV optimum is unaffected by heart rate. Neither QRS minimization nor LVOT VTI satisfy all validity criteria, and therefore seem weaker candidate modalities for VV optimization. AF, unlinking interventricular from atrioventricular delay, uniquely exposes resynchronization concepts to experimental scrutiny.
Authors:
Andreas Kyriacou; Matthew E Li Kam Wa; Punam A Pabari; Beth Unsworth; Resham Baruah; Keith Willson; Nicholas S Peters; Prapa Kanagaratnam; Alun D Hughes; Jamil Mayet; Zachary I Whinnett; Darrel P Francis
Related Documents :
22380084 - The dependence of extracted current on discharge gas pressure in neutral beam ion sourc...
8333234 - Effects of volume loading, propranolol, and heart rate changes on pump function and sys...
10445744 - Two step cardiomyoplasty with vascular delay: effect of stimulation of latissimus dorsi...
22369924 - Cardiovascular and electrocortical markers of anger and motivation during a simulated d...
751724 - Synthesis of myosin heavy and light chains in the afterloaded guinea pig right ventricle.
3322604 - Bradykinin does not contribute to hypotension in early canine endotoxemia.
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2012-3-26
Journal Detail:
Title:  International journal of cardiology     Volume:  -     ISSN:  1874-1754     ISO Abbreviation:  -     Publication Date:  2012 Mar 
Date Detail:
Created Date:  2012-3-30     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  8200291     Medline TA:  Int J Cardiol     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Copyright Information:
Copyright © 2012 Elsevier Ireland Ltd. All rights reserved.
Affiliation:
International Centre for Circulatory Health, National Heart and Lung Institute, Imperial College London, UK.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine

Full Text
Journal Information
Journal ID (nlm-ta): Int J Cardiol
Journal ID (iso-abbrev): Int. J. Cardiol
ISSN: 0167-5273
ISSN: 1874-1754
Publisher: Elsevier
Article Information
© 2013 Elsevier Ireland Ltd.
License:
Received Day: 8 Month: 9 Year: 2011
Revision Received Day: 8 Month: 2 Year: 2012
Accepted Day: 3 Month: 3 Year: 2012
pmc-release publication date: Day: 10 Month: 8 Year: 2013
Print publication date: Day: 10 Month: 8 Year: 2013
Volume: 167 Issue: 3
First Page: 954 Last Page: 964
PubMed Id: 22459364
ID: 3744806
Publisher Id: IJCA14575
DOI: 10.1016/j.ijcard.2012.03.086

A systematic approach to designing reliable VV optimization methodology: Assessment of internal validity of echocardiographic, electrocardiographic and haemodynamic optimization of cardiac resynchronization therapy
Andreas Kyriacoua⁎1 Email: andreas.kyriacou@imperial.ac.uk
Matthew E. Li Kam Waa1
Punam A. Pabaria
Beth Unswortha
Resham Baruaha
Keith Willsona
Nicholas S. Petersab
Prapa Kanagaratnamab
Alun D. Hughesab
Jamil Mayetab
Zachary I. Whinnettab
Darrel P. Francisab
aInternational Centre for Circulatory Health, National Heart and Lung Institute, Imperial College London, UK
bSt. Mary's Hospital, Paddington, London, UK
Corresponding author at: International Centre for Circulatory Health, 59-61 North Wharf Road, Paddington, London, W2 1LA, UK. Tel.: + 44 20 7594 1264; fax: + 44 20 7594 1706. andreas.kyriacou@imperial.ac.uk
1These authors contributed equally to this work.

Introduction

Of all the aspects of optimization, VV optimization of biventricular pacing is the most challenging. The term “Cardiac Resynchronization Therapy” itself implies that dyssynchrony between regional ventricular contraction is the disease being treated: if this is true then it must matter what interventricular delay is programmed. Although the PROSPECT trial [1] showed no worthwhile prediction of echocardiographic response from echocardiographic markers of dyssynchrony, clinical progress over months in patients with heart failure is dependent on a very wide spectrum of intercurrent environmental, neurohormonal, compliance, dietary, infectious, arrhythmic, ischaemic, and psychological phenomena, and the measurement techniques for dyssynchrony have very large beat-to-beat variability [2,3] so their failure to correlate with anything is unsurprising [4] and therefore uninfluential on the specific question of whether ventricular timings make any detectable difference in an environment where all else is unchanged.

Even this tightly-described question is difficult to answer. First, the effect of changing the VV delay is less than that of changing the AV delay [5], making it even more difficult to separate the genuine effect (signal) of changing a setting, from random variability (noise) [6]. Because the effect of VV optimization may be 5–10 times smaller than that of CRT implantation, an endpoint study would have to be 25–100 times larger than, for example, CARE-HF, to give reliable results. Second, changing the VV delay in patients with sinus rhythm inevitably affects the AV delay on one side of the heart or the other, so it is impossible to know for certain if any effects seen come from the AV or VV changes, or both [7].

For these reasons, whilst the benefits of AV optimization are well established [8–10], the benefits of VV optimization are less well understood. Although a few small studies have suggested that optimization of the VV delay may provide benefits beyond AV optimization alone [8,11,12], clinical trials have so far failed to show significant clinical improvements at 6 months [13–15].

In atrial fibrillation however, there is no AV delay to confound VV optimization (Fig. 1). Therefore any impact of a VV delay change is a direct consequence of that change in the VV delay alone. In this study, we recruited patients in AF as a model for ‘pure’ VV optimization.

As it is not yet realistic to look for differences in clinical outcomes between different markers of optimization, a practical way to compare these is to evaluate the relative performance of each method, head to head, in an identical patient group. Any marker of optimality used to select a pacemaker setting must fulfil three essential features that can be efficiently assessed:

  • 1)  Singularity; there should be only one region of optimality, with progressively poorer function as settings are changed away from this region. If the optimal region is at one extreme of settings, then there could be one rather than two regions of optimality, but it is not possible for there to be two regions of optimality separated by a region of non-optimality.
  • 2)  Reproducibility; if the optimization process is repeated immediately with a new operator blinded to the previously found optimal setting, the newly-found optimum should be very similar.
  • 3)  Plausibility; the distribution of optimum settings should not contradict established physiological principles. For example, frequently finding apparent VV optima with large RV pre-excitation would contradict the principle in resynchronization that the LV wall should usually be paced simultaneously with, or earlier than, the RV.

In this study, we tested three different optimization markers that might be used [16–18] to identify a VV delay optimum in the AF cohort, in which there was no possibility of VV delay alteration causing confounding changes in the AV delay. The markers tested were ECG (QRS duration minimization), LVOT VTI maximization (flow) and non-invasive arterial BP (pressure).


Methods
2.1  Patients

Twenty patients (16 men, mean age 75 ± 7 years) in atrial fibrillation, with a CRT device were enrolled in the study. Average NYHA class was 2.4 ± 0.5. The underlying cause of heart failure was ischemic heart disease in 10, dilated cardiomyopathy in 8 and valvular disease in 2 patients.

The study was approved by the Imperial College Healthcare NHS Trust ethics committee, and all patients provided written informed consent.

2.2  Study design

Measurements, using transthoracic echocardiography, finger photoplethysmography (Finometer) and electrocardiography, were made at six VV delays (− 40 ms, − 20 ms, 0 ms, 20 ms, 40 ms and 60 ms, where a negative VV delay indicates right ventricular pre-excitation). All patients were optimized at 2 heart rates; at a slow paced rate just above the rate of intrinsic conduction (always ensuring 100% biventricular pacing); and at a faster paced rate (mean of 27 ± 7 bpm above slow). Optimization was repeated within one hour using all three optimization modalities, at both heart rates.

Because beat-to-beat biological noise exceeds between-setting variability, often by a large margin, it is not possible to reliably define the optimum as the setting which generates the highest observed pressure or flow (or smallest QRS duration) [3,18,19]. Instead, for each variable a curve-fitting approach was used which minimises the impact of noise and makes maximal use of scarce signal [6].

2.3  Finger photoplethysmography

Beat-to-beat blood pressure was recorded using a Finometer (Finapres Medical Systems, Amsterdam, Netherlands). Optimization of the VV delay was performed using a protocol described in previous work [5,6,20–23] and algorithmically similar that used by others [24] Transitions to the tested VV delay from a fixed reference VV delay (of 0 ms) were performed. A minimum of eight transition replicates were recorded for each tested VV delay. Within each replicate, the 10 beats after the transition were compared with the 10 beats before the transition, to calculate the relative systolic blood pressure. Analysis was performed using software based on the Matlab platform (MathWorks, Natick, MA, USA) but could equally have been carried out with a standard spreadsheet. The relationship was fitted to a parabola by standard least-squares regression, and the optimum VV delay defined by interpolation as the position of the peak of the parabola [6].

2.4  Echocardiography

Images were obtained using a ProSound SSD-5500SV system (Aloka, Tokyo, Japan), with the operator blinded to the VV delays which were randomly programmed by a second operator. Six consecutive beats, of left ventricular outflow tract (LVOT) and mitral valve (MV) flow were acquired with the patient positioned in the dorsal decubitus or left lateral decubitus position, at passive end expiration. The average of the 6 beats was taken as the value for that setting. Medcon software (McKesson, San Francisco, USA) was used for offline analysis. The myocardial performance index (MPI) [25] was calculated by measuring the time from cessation to the start of mitral flow (A) and left ventricular ejection time (ET). Therefore, IVCT + IVRT = A − ET and as a result MPI = (A − ET) / ET.

The VTI optimum was defined by parabolic fitting of the average values for each setting, as the interpolated position of the greatest VTI.

2.5  Electrocardiography

12 lead ECG traces were obtained using a MAC 1200 ST system (GE Healthcare, Chalfont St Giles, UK) at each VV delay and at both heart rates for all patients. In this system, the QRS width is calculated using synchronously sampled and time aligned data across 12 leads, measuring the time from first deflection of the median QRS complex in any lead to the latest deflection in any lead. The average QRS duration from 2 recordings was used for each setting. Recordings which contained ventricular ectopics were excluded and the 12 lead ECG repeated. The optimum was defined by the same principle as above, as the interpolated position of the narrowest QRS.

2.6  Statistical analysis

Reproducibility and agreement of the parabola-determined [5] optima by different optimization methods were assessed by using Bland–Altman plots and the standard deviation of the difference (SDD) between the optimal VV delays. Paired comparisons of normally-distributed continuous variables were made using Student's paired t test. Paired comparisons of non-normally-distributed continuous variables were made using Wilcoxon's matched-pairs signed rank test. A p value of < 0.05 was taken as statistically significant. Statview 5.0 (SAS Institute Inc., Cary, NC) was used for statistical analysis.

All studied patients are presented. All data are presented. We believe these findings are representative of patients with CRT and AF and know of no undisclosed biases in recruitment, analysis or presentation. DPF is guarantor of scientific integrity [34].


Results

20 patients were included in this study. Their characteristics are shown in Table 1.

3.1  Criterion 1: Singularity—one optimum region

All 20 patients’ optimization curves for LVOT VTI maximization, pressure maximization and QRS minimization are shown in Figs. 2, 3 and 4, respectively.

Only instances in excess of 50% were evidence of the physiological validity of a measure, as shown in Table 2. The percentage of singularity was 85% for pressure, 63% for QRS and 45% for flow. The extent of singularity was significantly different between modalities (Chi2 = 10.9, p < 0.005). The percentage of singularity for all parameters tested is shown in Fig. 5.

3.2  Criterion 2: Reproducibility—the same optimum when retested

Reproducibility of optimization was quantified by the standard deviation of the difference (SDD). This is shown for all measures at the slow and fast heart rates in Tables 3 and 4, respectively.

Optimization by LVOT VTI showed a wide scatter between successive optima (SDD 35.4 ms). Other echo markers also performed poorly if considered as potential methods of optimization. The most reproducible methods of optimization were SBP maximization (SDD 9.4 ms) and QRS minimization (SDD 6 ms). Bland–Altman plots for the three key modalities being studied (flow, pressure and QRS) are shown in Fig. 6.

3.3  Criterion 3: Plausibility of the distribution of optima

The distribution of optimal VV delays at the fast heart rate, of all 3 optimization methods is shown in Fig. 7. By pressure, the optimum was often near a 0 ms VV delay: 75% were within 20 ms of zero. A significant majority (80%, p = 0.007) of pressure optima had a degree of LV pre-excitation.

In contrast, the flow and QRS optima tended to be further away from zero: only 60% and 50%, respectively were within 20 ms of zero. Moreover, the proportion of optima by these modalities that had a degree of LV pre-excitation was not significantly different from chance (45% for flow and 55% for QRS, p = NS in each case).

3.4  Agreement of the optimal VV delay between LVOT VTI, QRS width and beat-to-beat systolic blood pressure

Incidentally available was agreement between modalities, also displayed in Tables 3 and 4. Agreement between methods was poor, typically with a SDD of the order of 40 ms, meaning that if a method gave an optimum of X, the other method would be expected 95% of the time to give an optimum between X-80 ms and X + 80 ms, a very wide range.

We should be mindful that no irreproducible method can ever agree with any other method [4]. Thus, it could be argued that the only scientifically valuable numbers on the table are the principal diagonal (reproducibility of each modality) and the twin values of “43.9” in the bottom right corner, which shows the between-modality test for the only two modalities (QRS minimization and SBP maximization) that show good reproducibility.


Discussion

This study evaluated the performance of three potential approaches for VV optimization, using the three key criteria for internal validity of an optimization. Of the approaches tested, optimization by pressure appears to offer all 3 criteria: singularity, reproducibility and biological plausibility. QRS minimization offers singularity and reproducibility, but the distribution of optimum settings obtained appears biologically implausible. Optimization by flow performed disappointingly on all 3 criteria of internal validity.

4.1  Three criteria for internal validity

Reproducibility is a key criterion for any method of optimization. If an individual patient is given multiple discrepant proposed ‘optima’ in rapid succession by a single method, most of these apparent optima must be incorrect. While reproducibility alone is not sufficient criterion to judge an optimization approach to be internally valid, it is a necessary one.

Singularity indicates that there is only a single region which appears to be optimal, rather than two optimal regions separated by non-optimality. This is relevant because the principle of CRT is to improve cardiac function by bringing the walls of the heart into more closely coordinated times of contraction (although not necessarily to VV 0 ms). Moving away from this optimal timing of contraction in either direction should worsen cardiac function. In some cases (when the true optimal region is at the extremes of the range of tested settings) there might reasonably be only one non-optimal region, but singularity can still be tested by the shape of the curve. Because the shape of the curve should be almost flat at the optimal region, becoming steep as one moves away, fitting a parabola will still verify singularity (i.e. the orientation of curvature would indicate that there would have been 2 non-optimal regions, if more settings beyond the optimal had been tested).

Biological plausibility of the distribution of the obtained optima is the final criterion. It is less fundamental than the others because it relies on separately-acquired beliefs of what distribution of optima is expected. CRT is generally accepted to exert benefits by decreasing rather than increasing dyssynchrony of the myocardial walls. It is therefore rational to expect the distribution of the optima to show that for many patients, their optimum setting is near to the VV delay of 0 ms, with only few patients being at an optimum with marked pre-activation of one lead. It is left, rather than right, bundle branch block in which the salutary effects of biventricular pacing are most prominent, and invasive studies [10,26–28] have shown that left ventricular pacing can achieve most of the effect of biventricular pacing but right ventricular pacing cannot. These prior observations are reasons to expect the VV delay optimum to more often involve pre-activation of the left ventricle than the right.

4.2  Internal validity as an endpoint in a “basic science” of optimization

This study specifically focussed on the three key characteristics of internal validity of an optimization approach. We did this because these are essential criteria that may be definitively established in controlled circumstances, and which enables early identification of potentially unsuitable optimization approaches so that any large clinical event trials can focus resources appropriately.

In this study, only optimization by pressure fulfilled all 3 criteria.

4.3  Does VV delay matter at all?

The term “resynchronization”—which can only mean interventricular because atrioventricular timing is never synchronous in health—implies that VV timing matters, but appears contradicted by data [29]. For example, the prospectively-enrolled PROSPECT study appeared to contradict the belief that mechanical dyssynchrony is relevant; and several well-conducted studies of VV optimization appear to contradict the belief that interventricular delay matters at all.

However, the fact that various imaging markers in PROSPECT were mutually contradictory guarantees that most of them must fail to predict benefit. That they all failed highlights the fact that the study's process of choosing markers was arbitrary, based on availability rather than any detectable scientific process (which would screen out those with poor test–retest reproducibility). PROSPECT therefore casts no light on whether interventricular delay matters.

The 3 VV optimization studies are also less decisive than might be assumed. First, the Decrease-HF [14] study demonstrated that programming VV delay according to a formula obtained purely during intrinsic conduction gave no benefit, but it was not a process of measuring values at different settings to select an optimum. Second, the Rhythm II ICD [13] study used echocardiographic maximization of LVOT VTI, and found no benefit, did not report whether the method of optimization had good test–retest reproducibility: without knowing this we cannot tell whether it was VV delay that was unimportant or the optimization process that was unreliable [3]. Third, the Insync III [15] study was not a randomized trial but a comparison with historical controls. Moreover, although it conducted more than one optimization per patient, the test–retest reproducibility data remains at present undisclosed. Without this information on whether patients retested had the same optima as before, or only just fitting the same overall distribution, conclusions cannot yet be drawn.

4.4  Need for, and plausibility of, reports of clinical endpoint effects of optimization

Some interventions, such as device implantation, impose a large cost and substantial risk for the patient, and therefore judgements on whether to carry them out universally are dependent on the results of randomised trials of clinical outcomes. The financial investment required for these trials typically runs into many millions of dollars, so they are mainly carried out when an industrial organisation, hoping to profit from the conclusions, is willing to invest.

But clinicians make many choices. Not all are “whether” to do something extra which might have a cost or impose risk. Others are “which” of many options: selection of pacemaker settings is an example of such a choice. For such decisions, awaiting an adequately-powered industry-supported clinical endpoint trial may be a forlorn vigil.

In this study we did not attempt to acquire clinical endpoint data from these differences in VV delay. We made this choice because the effect was not as large as the effect of simply switching on a CRT device, and so the number of patients required for the clinical endpoints to be realistically informative—which rises with 1/(effect size) 2—would be very large.

Instead we concentrated on haemodynamic responses whose reproducibility is easily verifiable by any reader and which avoid the illusions of response arising from using optimization data itself to measure response [3], and accidentally generated correlations arising from confounding exclusions or optimistic reanalysis [30]. We are confident our findings will stand the test of time.

4.5  LVOT VTI as a modality for VV optimization

There is a strong a priori rationale for the use of LVOT VTI (flow) as a marker of cardiac function, for optimization. Given that the diameter of the LVOT remains constant, changes in stroke volume are solely dependent on velocity and duration of blood flow. Therefore, the LVOT VTI is a perfect index of stroke volume which in turn is, in principle, a very plausible marker of cardiac function since other characteristics are kept constant.

In this study, conducted in a research setting with less time constraint than occurs in routine clinical practice, we measured a number of LVOT VTIs (6 beats) that would be considered large by routine clinical standards. The reason was to improve the signal-to-noise ratio and therefore improve the reproducibility of the optimal VV delay [3]. Resource constraints often limit routine clinical practice to only 3, 2 or even 1 beat.

Despite taking this step, LVOT VTI optimization performed poorly on all counts of internal validity. This may be an explanation for the inconsistent findings reported by other studies testing clinical outcomes following VV optimization [13–15].

Singularity, (test–retest) reproducibility and plausibility of the optimum have rarely been commented upon in studies of LVOT VTI optimization.

4.6  QRS minimization as a modality for VV optimization

QRS width is a highly reproducible measurement that is quickly and cheaply acquired, potentially automatically, and it avoids the problems of needing good echocardiographic windows and a trained operator with enough time to perform and analyse a lengthy series of measurements. There are studies showing some association between a reduction in the QRS width and clinical benefits [31,32].

However, our findings showed that although reproducible and generally singular, QRS minimization frequently proposed optimal VV delays that had marked right ventricular pre-activation (Fig. 7). This is not biologically plausible if the current understanding of cardiac resynchronization process is correct.

Moreover, in our study, QRS and pressure optima agreed poorly. Because we have quantitative data on reproducibility of these techniques in the same patients at the same session, we know that the discrepancy between methods is not due to imprecision of our ability to establish each optimum, but rather that the two optima are providing fundamentally contradictory information.

It is inescapable from our data that QRS minimisation definitely results in a lower blood pressure. Forced to choose between a pressure maximum and a QRS minimum, our interpretation is that the pressure maximum is both intrinsically a better physiological target as well as showing a biologically more plausible distribution of VV optima. The numerous elements contributing to mechanical dyssynchrony, and cardiac function overall, go far beyond QRS width, as has been documented in several studies [33,34]. Therefore, even though wide QRS may predict benefit for CRT [1,35–37], first principles do not require that QRS minimization must be the ideal approach to VV optimization.

4.7  VV optimum unchanged by heart rate, if measured reproducibly

Using pressure optimization, the internally valid method, we concluded that heart rate had no discernible effect on optimal VV delay. This not to say that the observed optimum was the same at slow as fast heart rate. On the contrary, in each patient, the observed optimum was different between slow and fast. However, this change was as frequently an increase as a decrease, and its scatter was the same as the test–retest scatter of individual patients’ optimal settings at slow or fast heart rates. This means that no mechanism other than natural biological variability (unrelated to heart rate change) is necessary to explain the changes observed between heart rates. While we cannot exclude a small heart rate effect, any such effect must be small.

Judgement of reports of change in optimum of a pacemaker with the passage of time or after an intervention should always be reserved until it has a context of blinded short-term test–retest reproducibility data for that optimization process.

4.8  Clinical implications

Optimization techniques vary in their ability to detect the optimum reliably as we have seen in this study and have reviewed with a generalizable framework previously [3].

If there is insufficient time to perform a reliable optimization [3], unreliable optimization should not be performed except perhaps where mandated as part of a randomized controlled trial, or as part of an evaluation such as this study. It consumes resources and attention; moreover it may cause clinical harm by shifting patients away from default settings (near which the true optimum may lie in most patients) to VV delays far away which could worsen cardiac function. This concern is sound grounds for a clinician to leave default VV settings unchanged until optimization methods with good reliability become available locally.

Disobeying guidelines systematically takes courage if they are well-founded [37]. However, this study provides quantitative reasoning and measurements—both of which can be tested afresh by any party—to hold guideline recommendations to account. For example, if a clinical service uses LVOT VTI to optimize the VV delay, at the slow heart rate the scatter between repeated optimization using flow is 34 ms. This means after one optimization, the next identical procedure will, 95% of the time, report an optimization within ± 1.96 × 34, i.e. ± 67 ms; in other words a range as wide as all the settings tested. Clinical wisdom may prevail, preventing extreme values being selected, but if so, what is the net effect other than random selection amongst clinically-reasonable VV delays?

4.9  Study limitations

This study was not designed to test the clinical benefit of VV optimization, nor to attempt to separate responders from non-responders. It aimed to evaluate the qualities of non-invasive candidate modalities for VV optimization relevant to internal validity.

Although measurements were performed in a teaching hospital by experienced clinical research staff, there may be centres with greater skills in, for example, echocardiography which might be able to achieve better performance from flow optimization (LVOT VTI) than what we achieved. Superficially, that would appear to be the case from the fact that guidelines recommend flow optimization of the VV delay [37]. Nevertheless, we note that the guidelines do not cite independent data on the 3 key principles of internal validity (singularity, reproducibility and plausibility) of flow optimization, or any comparison of these principles between flow and any other alternatives, and therefore it is possible that flow optimization was recommended only because there appeared to be no alternative.

Ventricular ectopy was completely excluded from echocardiography images and from QRS measurements, but not eliminated from the semi-automated analysis of pressure optimization. This may have adversely affected the precision of the pressure data. However, results in terms of singularity, reproducibility and plausibility, were still no worse for pressure than for flow and QRS. The practical advantage of pressure over flow is that it could be monitored by automated systems without manual beat-by-beat calculation, and therefore it is fitting that pressure measurement was calculated in this study in the same way that might happen in a real-world clinical system operating without beat-by-beat supervision.

Optimal VV delays were calculated using data taken with the patient lying down, and this may not necessarily reflect the optimal VV delay during other physiological states such as exercise. However, for the purpose of genuine head-to-head comparison of several optimization methods we had to select a posture suitable to all modalities including echo, and which did not introduce unnecessary noise.

We performed pressure optimization by measuring the changes that occurred in pressure only for a short interval after a change in the VV delay was programmed and did not measure pressure during steady state pacing. We have recently shown [23] that any change in pressure that occurs due to a change in AV delay decays to a lower level only a few seconds later. This phenomenon means that signal-to-noise ratio for pressure optimization is highest soon after a transition, and therefore systematically measured changes just shortly after the VV delay change.

Moreover, we assessed the test–retest reproducibility of the optimum over a short of period of time (within an hour) to guarantee that any differences could not be attributed to large physiological or volume changes between optimization sessions.

Finally, this is a study of AF only. Sinus rhythm may be different, not least because changes in the VV delay always change the AV delay on either the left or right side of the heart.


Conclusions

Singularity, reproducibility and plausibility are sine qua nons for an optimization modality to undergo the next steps of mutual comparison between high-quality modalities, trialling for impact on clinical endpoints, or adoption in clinical practice. Methods that do not have these characteristics may be, unintentionally, a random selection amongst settings. At best this could be useless. At worst, since the patients would be moving away from the vicinity of the optimum in which they begin by factory default (0 ms), optimization might be on average harmful.

Although LVOT VTI is the “final common pathway” of all processes within the heart and—being an index of cardiac output—is in principle a perfect marker for pacemaker optimization, it has a poor singularity and reproducibility even in an ideal research setting where time and resources are artificially abundant. Its singularity and reproducibility could be improved, but only by massively increasing the number of replicates to a level intolerable even in a research setting and certainly in a routine clinical setting.

In our study, only pressure optimization provided singularity, reproducibility and plausibility. This study cannot truly compare pressure optimization against flow or QRS optimization, because neither flow nor QRS optimization were suitable candidates since they both failed to fulfil criteria that are rudimentary.


Funding sources and disclosures

Dr Kyriacou (FS/08/027/24763) and Dr Francis (FS/10/038) are supported by fellowships from the British Heart Foundation. Our institution has filed a patent on some of the methods described in this manuscript. The authors are grateful to the UK's NIHR Biomedical Research Centre scheme and the BHF Research Excellence Award scheme.


References
1. Chung E.S.,Leon A.R.,Tavazzi L.. Results of the predictors of response to CRT (PROSPECT) trialCirculation117Year: 20082608261618458170
2. Vesely M.R.,Li S.,Kop W.J.. Test–retest reliability of assessment for intraventricular dyssynchrony by tissue Doppler imaging echocardiographyAm J Cardiol1015Year: Mar 1 200864565018308014
3. Pabari P.A.,Willson K.,Stegemann B.. When is an optimization not an optimization? Evaluation of clinical implications of information content (signal-to-noise ratio) in optimization of cardiac resynchronization therapy, and how to measure and maximize itHeart Fail RevYear: Nov 26 2010 [Epub ahead of print].
4. Francis D.P.,Coats A.J.S.,Gibson D.G.. How high can a correlation coefficient be? Effect of limited reproducibility on common cardiological variablesInt J Cardiol69Year: 199918519910549842
5. Whinnett Z.I.,Davies J.E.R.,Willson K.. Haemodynamic effects of changes in atrioventricular and interventricular delay in cardiac resynchronisation therapy show a consistent pattern: analysis of shape, magnitude and relative importance of atrioventricular and interventricular delayHeart92Year: 20061628163416709698
6. Francis D.P.. Precision of a parabolic optimum calculated from noisy biological data, and implications for quantitative optimization of biventricular pacemakers (cardiac resynchronization therapyAppl Math212Year: 201114971506
7. Bogaard M.D.,Kirkels J.H.,Hauer R.N.,Loh P.,Doevendans P.A.,Meine M.. Should we optimize cardiac resynchronization therapy during exercise?J Cardiovasc Electrophysiol21Year: 20101307131620550613
8. Bordachar P.,Lafitte S.,Reuter S.. Echocardiographic parameters of ventricular dyssynchrony validation in patients with heart failure using sequential biventricular pacingJ Am Coll Cardiol44Year: 20042157216515582313
9. Sogaard P.,Egeblad H.,Pedersen A.K.. Sequential versus simultaneous biventricular resynchronization for severe heart failure: evaluation by tissue Doppler imagingCirculation106Year: 20022078208412379577
10. Auricchio A.,Stellbrink C.,Block M.. Effect of pacing chamber and atrioventricular delay on acute systolic function of paced patients with congestive heart failure. The Pacing Therapies for Congestive Heart Failure Study Group. The Guidant Congestive Heart Failure Research GroupCirculation99Year: 19992993300110368116
11. van Gelder B.M.,Bracke F.A.,Meijer A.,Lakerveld L.J.M.,Pijls N.H.J.. Effect of optimizing the VV interval on left ventricular contractility in cardiac resynchronization therapyAm J Cardiol93Year: 20041500150315194020
12. Perego G.B.,Chianca R.,Facchini M.. Simultaneous vs. sequential biventricular pacing in dilated cardiomyopathy: an acute hemodynamic studyEur J Heart Fail5Year: 200330531312798828
13. Boriani G.,Muller C.P.,Seidl K.H.. Randomized comparison of simultaneous biventricular stimulation versus optimized interventricular delay in cardiac resynchronization therapy. The Resynchronization for the Hemodynamic Treatment for Heart Failure Management II implantable cardioverter defibrillator (RHYTHM II ICD) studyAm Heart J151Year: 20061050105816644335
14. Rao R.K.,Kumar U.N.,Schafer J.,Viloria E.,De Lurgio D.,Foster E.. Reduced ventricular volumes and improved systolic function with cardiac resynchronization therapy: a randomized trial comparing simultaneous biventricular pacing, sequential biventricular pacing, and left ventricular pacingCirculation115Year: 20072136214417420340
15. Leon A.R.,Abraham W.T.,Brozena S.. Cardiac resynchronization with sequential biventricular pacing for the treatment of moderate-to-severe heart failureJ Am Coll Cardiol46Year: 20052298230416360062
16. Tamborero D.,Mont L.,Sitges M.. Optimization of the interventricular delay in cardiac resynchronization therapy using the QRS widthAm J Cardiol104Year: 20091407141219892059
17. Bertini M.,Ziacchi M.,Biffi M.. Interventricular delay interval optimization in cardiac resynchronization therapy guided by echocardiography versus guided by electrocardiographic QRS interval widthAm J Cardiol102Year: 20081373137718993158
18. Turcott R.G.,Witteles R.M.,Wang P.J.,Vagelos R.H.,Fowler M.B.,Ashley E.A.. Measurement precision in the optimization of cardiac resynchronization therapyCirc Heart Fail3Year: 201039540420176716
19. Cuoco F.A.,Gold M.R.. Optimization of cardiac resynchronization therapy: importance of programmed parametersJ Cardiovasc ElectrophysiolYear: Dec 21 2011
20. Whinnett Z.I.,Nott G.,Davies J.E.. Maximizing efficiency of alternation algorithms for hemodynamic optimization of the AV delay of cardiac resynchronization therapyPacing Clin Electrophysiol342Year: Feb 201121722521029129
21. Whinnett Z.I.,Davies J.E.,Nott G.. Efficiency, reproducibility and agreement of five different hemodynamic measures for optimization of cardiac resynchronization therapyInt J Cardiol1292Year: Sep 26 200821622617881070
22. Whinnett Z.I.,Davies J.E.,Willson K.. Determination of optimal atrioventricular delay for cardiac resynchronization therapy using acute non-invasive blood pressureEuropace8Year: 200635836616635996
23. Manisty C.H.,Al-Hussaini A.,Unsworth B.. The acute effects of changes to AV delay on blood pressure and stroke volume: potential implications for design of pacemaker optimization protocolsCirc Arrhythm ElectrophysiolYear: Nov 17 2011 [Epub ahead of print].
24. van Geldorp I.E.,Delhaas T.,Hermans B.. Comparison of a non-invasive arterial pulse contour technique and echo Doppler aorta velocity-time integral on stroke volume changes in optimization of cardiac resynchronization therapyEuropace131Year: Jan 2011879520880954
25. Tei C.,Ling L.H.,Hodge D.O.. New index of combined systolic and diastolic myocardial performance: a simple and reproducible measure of cardiac function—a study in normals and dilated cardiomyopathyJ Cardiol26Year: 19953573668558414
26. Auricchio A.,Ding J.,Spinelli J.C.. Cardiac resynchronization therapy restores optimal atrioventricular mechanical timing in heart failure patients with ventricular conduction delayJACC39Year: 20021163116911923041
27. Kass D.A.,Chen C.H.,Curry C.. Improved left ventricular mechanics from acute VDD pacing in patients with dilated cardiomyopathy and ventricular conduction delayCirculation99Year: 19991567157310096932
28. Butter C.,Auricchio A.,Stellbrink C.. Effect of resynchronization therapy stimulation site on the systolic function of heart failure patientsCirculation10425Year: 200130263029 18. 11748094
29. Kyriacou A.,Pabari P.A.,Francis D.P.. Cardiac resynchronization therapy is certainly cardiac therapy, but how much resynchronization and how much atrioventricular delay optimization?Heart Fail Rev176Year: 201272773621796453
30. Francis D.P.. How easily can omission of patients, or selection amongst poorly-reproducible measurements, create artificial correlations? Methods for detection and implications for observational research design in cardiologyInt J Cardiol1671Year: 201310211322285446
31. Pitzalis M.V.,Iacoviello M.,Romito R.. Ventricular asynchrony predicts a better outcome in patients with chronic heart failure receiving cardiac resynchronization therapyJ Am Coll Cardiol45Year: 2005656915629375
32. Lecoq G.,Leclercq C.,Leray E.. Clinical and electrocardiographic predictors of a positive response to cardiac resynchronization therapy in advanced heart failureEur Heart J26Year: 20051094110015728648
33. Ghio S.,Constantin C.,Klersy C.. Interventricular and intraventricular dyssynchrony are common in heart failure patients, regardless of QRS durationEur Heart J25Year: 200457157815120054
34. Bleeker G.B.,Schalij M.J.,Molhoek S.G.. Frequency of left ventricular dyssynchrony in patients with heart failure and a narrow QRS complexAm J Cardiol95Year: 200514014215619414
35. Linde C.,Abraham W.T.,Gold M.R.. Randomized trial of cardiac resynchronization in mildly symptomatic heart failure patients and in asymptomatic patients with left ventricular dysfunction and previous heart failure symptomsJ Am Coll Cardiol52Year: 20081834184319038680
36. Moss A.J.,Hall W.J.,Cannom D.S.. Cardiac-resynchronization therapy for the prevention of heart-failure eventsN Engl J Med361Year: 20091329133819723701
37. Gorcsan J. III,Abraham T.,Agler D.A.. Echocardiography for cardiac resynchronization therapy recommendations for performance and reporting—a report from the American Society of Echocardiography Dyssynchrony Writing Group endorsed by the Heart Rhythm SocietyJ Am Soc Echocardiogr213Year: Mar 200819121318314047

Acknowledgement

The authors of this manuscript have certified that they comply with the Principles of Ethical Publishing in the International Journal of Cardiology.


Article Categories:
  • Article

Keywords: Keywords Biventricular pacemakers, Pressure, Echocardiography, Electrocardiography.

Previous Document:  Novel devices and specialized techniques in recanalization of peripheral artery chronic total occlus...
Next Document:  Comparison of various methods for quantitative evaluation of myocardial infarct volume from magnetic...