A simple criterion for gas chromatography/mass spectrometric analysis of thermally unstable compounds, and reassessment of the by-products of alkyl diazoacetate synthesis.
MedLine Citation:
PMID:  23280978     Owner:  NLM     Status:  In-Data-Review    
Abstract/OtherAbstract:
RATIONALE: A principal limitation of gas chromatography (GC) and gas chromatography/mass spectrometry (GC/MS) is the thermal instability of analytes. We propose that the injector and column temperatures should not exceed the atmospheric pressure boiling point, without decomposition, of the highest homologue of the series being analyzed, instead of the time-consuming procedure of obtaining chromatograms using different temperatures.
METHODS: A series of thermally unstable diazocarbonyl compounds, alkyl diazoacetates (predicted limit of stability approx. 140 °C, the boiling point of ethyl diazoacetate), was selected for GC/MS analysis using standard equipment. Different GC separation conditions were selected so that the retention temperatures of target compounds were both below and above 140 °C.
RESULTS: Analyzing alkyl diazoacetates within their thermal stability range permitted reanalysis of their typical synthesis by-products. No dialkyl fumarate or maleate impurities, principal decomposition products which have often been reported previously, were found. Instead, alkyl esters of glycolic acid nitrate, O(2) NOCH(2) CO(2) R, and 'pseudo-dimeric' products, ROCO[C(2) H(3) NO]CO(2) R, were discovered for the first time.
CONCLUSIONS: Avoiding the decomposition of thermally unstable organic compounds during GC and/or GC/MS analysis requires estimating their degradation temperature limits. This limit can be estimated as being equal to the atmospheric pressure boiling point of the highest homologue in the homologous series under consideration that does not decompose on boiling. Copyright © 2012 John Wiley & Sons, Ltd.
Authors:
Tatiana A Kornilova; Anton I Ukolov; Rafael R Kostikov; Igor G Zenkevich
Related Documents :
24879538 - Measurement of meropenem concentration in different human biological fluids by ultra-pe...
10845378 - A multicomponent mobile phase for ion chromatography applied to the separation of anion...
22993048 - Mass spectrometry parameters optimization for the 46 multiclass pesticides determinatio...
19886638 - Single droplet separations and surface partition coefficient measurements using laser a...
17451908 - Automated sample preparation and purification of homogenized brain tissues.
23298628 - Reusable light-emitting-diode induced chemiluminescence aptasensor for highly sensitive...
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Rapid communications in mass spectrometry : RCM     Volume:  27     ISSN:  1097-0231     ISO Abbreviation:  Rapid Commun. Mass Spectrom.     Publication Date:  2013 Feb 
Date Detail:
Created Date:  2013-01-02     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  8802365     Medline TA:  Rapid Commun Mass Spectrom     Country:  England    
Other Details:
Languages:  eng     Pagination:  461-6     Citation Subset:  IM    
Copyright Information:
Copyright © 2012 John Wiley & Sons, Ltd.
Affiliation:
Department of Chemistry, St. Petersburg State University, Universitetskii prosp. 26, St. Petersburg, 198504, Russia.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Fragmentation characteristics and isomeric differentiation of flavonol O-rhamnosides using negative ...
Next Document:  Direct analysis in real time high-resolution mass spectrometry for high-throughput analysis of antip...