| A pressure controller of selective cerebral perfusion with single centrifugal pump. | |
| | |
MedLine Citation:
|
PMID: 12058996 Owner: NLM Status: MEDLINE |
Abstract/OtherAbstract:
|
We are developing an original nonroller extracorporeal circulation system (NRECC). However, this NRECC could not perform selective cerebral perfusion (SCP). Therefore, we added cerebral perfusion lines and an automatic pressure controller to the system. The purpose of this study was to evaluate the stability and response of the pressure controller for the SCP in simulated clinical circulation. The NRECC consists of a centrifugal pump, four isolated vacuum suction lines and a conventional module. The SCP line branches from the main perfusion tubing and divides into three lines. Cerebral perfusion is regulated by a pressure controller, which is composed of an actuator, a pressure transducer, and a personal computer. The mock circuit was primed with normal saline, and the actual SCP pressure and flow were measured when the target pressure settled at 60, 80, and 100 mm Hg. The main perfusion flow was maintained at 4 L/min and the main perfusion pressure was altered from 120 to 300 mm Hg. The pressure and flow data were recorded. The SCP pressure was maintained within +/-1.47 mm Hg when the set pressure was 80 mm Hg. Fluctuation of flow in the SCP line was within the range of +/-2.8%. The time needed to reach the steady state pressure was 8+/-1 seconds when the initial setting of the roller occluder was full-open, and it took only 3+/-1 seconds to reach the next initial set pressure. We have developed the NRECC-SCP system. The SCP pressure is stable and quickly reaches steady state via the pressure controller. This system is useful for extracorporeal circulation during aortic arch operation. |
| | |
Authors:
|
Toshihiro Moriya; Yukio Fukaya; Satomi Sigemitsu; Takamitsu Terasaki; Satoshi Hirose; Jun Amano |
Related Documents
:
|
12911466 - Bedside biochemical monitoring of the penumbra zone surrounding an evacuated acute subd... 9782176 - Reflex control of splanchnic blood volume in anaesthetized dogs. 380266 - Prostacyclin-dependent coronary vasodilation in rabbit and guinea pig hearts. 16523026 - Effects of vascular perfusion on coagulation size in radiofrequency ablation of ex vivo... 18375746 - A new predictive indicator for development of pressure ulcers in bedridden patients bas... 17178776 - Higher plantar pressure on the medial side in four soccer-related movements. |
Publication Detail:
|
Type: Journal Article |
Journal Detail:
|
Title: ASAIO journal (American Society for Artificial Internal Organs : 1992) Volume: 48 ISSN: 1058-2916 ISO Abbreviation: ASAIO J. Publication Date: 2002 May-Jun |
Date Detail:
|
Created Date: 2002-06-12 Completed Date: 2002-11-15 Revised Date: 2006-05-23 |
Medline Journal Info:
|
Nlm Unique ID: 9204109 Medline TA: ASAIO J Country: United States |
Other Details:
|
Languages: eng Pagination: 239-43 Citation Subset: IM |
Affiliation:
|
SN Seiki Co., Ltd., Nagano, Japan. |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
Centrifugation Cerebrovascular Circulation* Extracorporeal Circulation / instrumentation* Humans Perfusion Pressure |
From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine
Previous Document: A new approach to completely autologous cardiovascular tissue in humans.
Next Document: Hemoperfusion with polymyxin B immobilized fibers for urinary albumin excretion in septic patients w...