Document Detail

A pilot trial on kinematic magnetic resonance imaging using a superconducting, horizontally opened, 1.2 T magnetic resonance system.
Jump to Full Text
MedLine Citation:
PMID:  22375248     Owner:  NLM     Status:  PubMed-not-MEDLINE    
Abstract/OtherAbstract:
PURPOSE: This study was performed to introduce and evaluate the potential of kinematic magnetic resonance imaging (KMRI) using a high-field open-magnet magnetic resonance (MR) system.
METHODS: We attempted to perform KMRI of healthy volunteers' lumbar spine and knee in the lateral position and ankle in the supine position utilizing the superconducting, horizontally opened, 1.2 T MR system (OASIS, HITACHI, Tokyo, Japan). For the KMRI of the lumbar spine, the volunteer had to lie on one side while maintaining maximally anteflexed, neutral, and maximally retroflexed positions and remain still for the duration of the acquisition time for each posture. In the same way, KMRI of the knee was performed with the volunteer's knee flexed at 0°, 30°, 60°, 90°, and 120° in the lateral position, and KMRI of the ankle was performed with the volunteer's ankle in maximally dorsiflexed, neutral, and maximally plantarflexed positions while lying in the supine position.
RESULTS: We could acquire higher quality kinematic MR images than those acquired using low-field MR systems. The spinal canal, intervertebral discs and foramina, and facet joints in lumbar spine KMRI; the ligaments, menisci and patellofemoral joint in knee KMRI; and the tibiotalar articulation and peroneal tendon in ankle KMRI were clearly depicted.
CONCLUSION: The results of our pilot trial indicated that a superconducting horizontally opened, 1.2 T MR system offers high-quality KMRI images and can be utilized for the kinematic diagnosis and evaluation of sports injuries.
Authors:
Daisuke Shimao; Yoshikazu Shimada; Jiro Kobayashi; Kazuyuki Kato; Tatsuya Misawa; Haruyasu Kato; Michiko Dohi
Related Documents :
22718588 - Experimental thresholds of magnetically induced currents via a figure-of-eight coil up ...
22672578 - Multimodal quantitative analysis of early pulsed-dye laser treatment of scars at a pedi...
22756618 - Su-e-t-439: first experience of three dimensional conformai radiotherapy (3dcrt) planni...
22449928 - Imaging the efficacy of anti-inflammatory liposomes in a rabbit model of atherosclerosi...
7638708 - Mr and spiral/helical ct imaging of lower extremity occlusive disease.
20813778 - Composite mri scores improve correlation with edss in multiple sclerosis.
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Asian journal of sports medicine     Volume:  2     ISSN:  2008-7209     ISO Abbreviation:  Asian J Sports Med     Publication Date:  2011 Dec 
Date Detail:
Created Date:  2012-02-29     Completed Date:  2012-08-23     Revised Date:  2013-05-29    
Medline Journal Info:
Nlm Unique ID:  101528850     Medline TA:  Asian J Sports Med     Country:  Iran    
Other Details:
Languages:  eng     Pagination:  267-74     Citation Subset:  -    
Affiliation:
Japan Football Association (JFA) Medical Center Hospital, Fukushima, Japan.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine

Full Text
Journal Information
Journal ID (nlm-ta): Asian J Sports Med
Journal ID (publisher-id): ASJSM
ISSN: 2008-000X
ISSN: 2008-7209
Publisher: Tehran University of Medical Sciences
Article Information
Download PDF
© 2011 Sports Medicine Research Center, Tehran University of Medical Sciences
open-access:
Received Day: 21 Month: 5 Year: 2011
Accepted Day: 16 Month: 9 Year: 2011
Print publication date: Month: 12 Year: 2011
Volume: 2 Issue: 4
First Page: 267 Last Page: 274
ID: 3289220
PubMed Id: 22375248
Publisher Id: ASJSM-2-267

A Pilot Trial on Kinematic Magnetic Resonance Imaging Using a Superconducting, Horizontally Opened, 1.2 T Magnetic Resonance System
Daisuke Shimao, PhD12*
Yoshikazu Shimada, BMSc1
Jiro Kobayashi, Dip1
Kazuyuki Kato, BMSc3
Tatsuya Misawa, MD1
Haruyasu Kato, MD, PhD14
Michiko Dohi, MD, PhD15
1Japan Football Association (JFA) Medical Center Hospital, Fukushima, Japan
2Department of Radiological Sciences, Ibaraki Prefectural University of Health Sciences, Ibaraki, Japan
3Department of Clinical Application, Hitachi Medical Corporation, Chiba, Japan
4Department of Sport and Wellness, Rikkyo University, Saitama, Japan
5Department of Sports Medicine, Japan Institute of Sports Sciences, Tokyo, Japan
*Corresponding Author:Address: 4669-2 Ami, Ami-machi, Inashiki-gun, Ibaraki 300-0394, Japan. E-mail:shimao@ipu.ac.jp

INTRODUCTION

Kinematic magnetic resonance imaging (KMRI) is a well-known diagnostic method with great potential. It consists of a series of static magnetic resonance (MR) images taken with the target joint flexed at various discrete angles. Investigators can observe structural displacements and distortions of the anatomies of joints or the spinal column including bones, intervertebral discs, neural elements, ligaments, facets, menisci, and tendons by varying their positions. Many studies of KMRI have been performed using permanent or superconducting magnets in horizontally or vertically opened MR systems with relatively low (0.2-0.6 T) magnetic field strengths. Detailed explanations and examples of clinical applications for KMRI were first described in a book edited by Shellock and Powers [1]. Although KMRI has been applied to the lumbar spine [28], cervical spine [612], knee [1321], ankle [2224], shoulder [2529], and wrist [3033] and it offers information that often serves to characterize the underlying abnormality or to complement the information acquired with standard static magnetic resonance imaging (MRI), the deterioration of image quality due to the low magnetic field cannot be avoided. Recently, the superconducting, horizontally opened, 1.2 T MR system (OASIS, HITACHI, Tokyo, Japan) has become available commercially, and our facility, JFA (Japan Football Association) Medical Center Hospital, installed the system under a grant from the FIFA (Fédération Internationale de Football Association) goal program in August 2009. This system utilizes the combination of a vertical magnetic field and a solenoid receiver coil, offering higher sensitivity and greater uniformity than the combination of a horizontal magnetic field and a multiarray receiver coil adopted in the standard tunnel-type MR system. This new system enables us to acquire high–quality images (high spatial resolution, tissue contrast, and signal–to–noise ratio) in acquisition times that are nearly the same as those of standard tunnel-type 1.5 T systems, even though the strength of the magnetic field is lower. Therefore, the purpose of this study was to introduce and evaluate the potential of KMRI utilizing a horizontally opened space by means of this high-field open-magnet MR system.


METHODS AND SUBJECTS
Participants

We focused on KMRI of three parts of the body – lumbar spine, knee, and ankle – where sports injuries commonly occur in soccer players. Three healthy female volunteers who are futsal players were recruited and each of them was assigned one of the parts to be examined for this pilot trial. We performed the study according to the Declaration of Helsinki requirements and obtained informed consent from all subjects.

Procedure

We utilized the superconducting, horizontally opened, 1.2 T MR system (OASIS) (Fig. 1a) for KMRI, which was equipped with a higher magnetic field than a conventional 0.2-0.6 T open-type MR system. The bore was widely opened horizontally (270°) and its height was 44 cm. One can easily recognize that the potential for KMRI is higher than that of the conventional tunnel-type bored MR system (Fig. 1b). We performed a qualitative comparison of the quality of KMRI images with those reported in previous studies.

The designated volunteer was required to lie on her right side and maintain maximum anteflexed, neutral, and maximum retroflexed positions for the duration of the acquisition time (TA) for each posture with a solenoid receiving coil put on her back[34] (Fig. 2). The imaging plane was set in the sagittal section and T2-weighted images (T2WI) were acquired using a fast spin-echo (FSE) sequence, which is routinely ordered in spine MRI. The parameters for the FSE T2WI were: repetition time (TR), 2400 ms; echo time (TE), 96.0 ms; echo train length (ETL), 16; number of phase encodes/frequency encodes (PE/FE), 320/352; slice thickness (T), 4.5 mm; slice gap (G), 0.5 mm; number of excitations (NEX), 4; field of view (FOV), 300 mm; reconstruction matrix (Mx), 512*512; TA, 3 min 14 s.

The designated volunteer was laid on her left side while maintaining her right knee in a 0°, 30°, 60°, 90°, or 120° flexion position fixed by a supporting tool (CHAMCO, Inc., Cocoa, Florida, USA) with her knee through the solenoid receiving coil (Fig. 3) during each TA. The imaging plane was set in the sagittal section and proton density weighted images (PDWI) were acquired using an FSE sequence, which is routinely ordered in joint MRI. The parameters for FSE PDWI were: TR, 1500 ms; TE, 12.0 ms; ETL, 3; PE/FE, 320/448; T, 3.0 mm; G, 0.3 mm; NEX, 1; FOV, 150 mm; Mx, 512*512; TA, 3 min 14 s.

The designated volunteer was laid supine while maintaining her ankle in a maximally dorsiflexed, neutral, or maximally plantarflexed position fixed by a supporting tool (CHAMCO, Inc.) with her ankle through the solenoid receiving coil (Fig. 4) during each TA. The imaging plane was set in the sagittal section and PDWI using an FSE sequence were acquired. The parameters for FSE PDWI were: TR, 1700 ms; TE, 12.6 ms; ETL, 4; PE/FE, 320/384; T, 3.0 mm; G, 0.3 mm; NEX, 1; FOV, 180 mm; Mx, 512*512; TA, 3 min 40 s.


RESULTS

For each of the three volunteers, all KMRI examinations could be performed without any technical failure; there were no issues regarding coil settings, space for examinees to pose, parameter settings for the sequences, and limitations to slice planning.

Representative results of KMRI of the lumbar spine are shown in Figure 5, which includes the slices at the middle plane of the vertebral body and at the level of the intervertebral foramen in maximal anteflexion, neutral, and maximal retroflexion positions. The alignments of the vertebral bodies and spinous processes and the shapes of the intervertebral discs and the vertebral canal according to the differences of the three postures may be observed in the upper images in Figure 5. In the same way, the sizes and shapes of the intervertebral foramina and the kinematics of the facet joints may be observed in the lower images in Figure 5. The kinematic images of the knee at the slice level of the anterior cruciate ligament (ACL), posterior cruciate ligament (PCL), and medial meniscus (MM) are shown with the flexion angle of the knee in Figure 6. The motion of the ACL, PCL, and MM could be depicted clearly with high spatial resolution and adequate signal-to-noise ratio. If one desires, the kinematics of the patellofemoral joint can also be evaluated using these KMRI images.

Kinematic images of the ankle in the maximal dorsiflexion, neutral, and maximal plantarflexion positions at the slice level of the central trochlea of the talus (TT) and peroneal tendon (PT) are shown in Figure 7. The movement of the tibiotalar articulation and the location of PT may be observed.


DISCUSSION

Many studies on KMRI have been performed in the past 20 years. Most of these have been performed using vertically [38,12,14,19,20,28,29] or horizontally [2,1618,21,24,26,27,34] opened MR systems, or a specified system of KMRI [15], excluding the studies under the small ranges of the joint posture within the bore size of conventional tunnel-type MR systems [13,25], because there are sometimes serious limitations on the poses of subjects in conventional tunnel-type MR systems. Conversely, KMRI of the cervical spine, ankle, or wrist, whose range of motions are relatively tiny, has been performed using a tunnel-type MR system [911,22,23,3033]. Generally, the higher the magnetic field of an MR system is, the higher the resulting image qualities such as spatial resolution, tissue contrast, and signal-to-noise ratio are and the shorter the acquisition time. However, MR systems with higher magnetic fields were available only as tunnel-type systems until recently. Thus, it has been impossible to perform KMRI of the lumbar spine, knee, or shoulder, which requires wide space beyond the size of a tunnel-type MR system's bore, with high-quality images due to the high magnetic field. Therefore, an open MR system with a high magnetic field has been desired for a long time. Finally, such a system has become available commercially, which gives clinicians a way to perform KMRI of those parts of the body, resulting in high-quality images in shorter acquisition times. Short acquisition time is invaluable, particularly in KMRI with postures attended with pain. In this study, acquisition times were over 3 min due to the sequence parameters for high spatial resolution and signal-to-noise ratio. If we set the sequence parameters to acquire the images whose qualities are equal to those acquired by a low field type of MR system, we can shorten the acquisition times.

We performed KMRI of the lumbar spine, knee, and ankle, which are closely associated with soccer players’ injuries, because our facility was established mainly for the treatment of soccer players. In the present study, KMRI of the lumbar spine and knee with high-quality images was performed for the first time in the world (Figs. 5 and 6), although KMRI of the ankle (Fig. 7) has been available using a tunnel-type, high magnetic field MR system. These high-quality KMRI images can not only demonstrate the lesions not visible on conventional MRI but also provide further detailed depiction of abnormalities derived from varying joints or spinal column postures. Those abnormalities include spinal canal and/or intervertebral foramen narrowing caused by deformation of an intervertebral disk and ligamentum flavum, dysfunction of a facet joint apparent in KMRI of the lumbar spine, ligament dysfunction and meniscus deformation apparent in KMRI of the knee, and tendon dysfunction, impingement syndromes, osteochondral defects, and loose bodies apparent in KMRI of the ankle. Although we excluded KMRI of the shoulder from this study, the shoulder can be one of the best candidates for KMRI using a high-field, horizontally opened MR system utilizing the large space of the gantry, which can provide high-quality images of tendons and the glenoid labrum for the diagnoses of rotator cuff tear and shoulder dislocation. Furthermore, we can acquire KMRI images with much higher spatial resolution utilizing a microscopic coil [35] under the high magnetic field, which will enable us to observe minute lesions. However, this high-field, horizontally opened MR system is not complete and there is a critical imperfection in a certain type of KMRI. We cannot perform weight-bearing KMRI with a horizontally opened MR system, although it is feasible with a vertically opened MR system [38,12,14,19,20]. The ideal solution for a full KMRI study may be the development of a high-magnetic-field, vertically opened MR system. Nevertheless, the fact that a high-magnetic-field, horizontally opened MR system has become commercially available will lead to progressive improvements in KMRI studies.


CONCLUSION

We performed KMRI of the lumbar spine, knee, and ankle using a high-field, open-magnet MR system. We introduced high-quality KMRI images and evaluated their potential for clinical use. Those KMRI images make possible a detailed depiction of abnormalities derived from varying joints or spinal column postures, including spinal canal and/or intervertebral foramen narrowing and dysfunction of a facet joint apparent in lumbar spine KMRI, ligament dysfunction and meniscus deformation apparent in knee KMRI, and impingement syndromes, osteochondral defects, and loose bodies apparent in ankle KMRI. The results of our pilot trial indicated that a superconducting horizontally opened, 1.2 T MR system offers high-quality KMRI images and can be utilized for the kinematic diagnosis and evaluation of sports injuries.


ACKNOWLEDGMENTS

The experiment was performed under approval by the JFA medical committee and was financially supported in part by research funds for the teaching staff from Ibaraki Prefectural University of Health Sciences.

Conflict of interests: The authors declare that they have no conflict of interests.


REFERENCES
1. Shellock FG,Powers CM. Kinematic MRI of the Joints. Functional Anatomy, Kinesiology, and Clinical Applications.Year: 2001Boca RatonCRC Press
2. Harvey SB,Smith FW,Hukins DWL. Measurement of lumbar spine flexion-extension using a low-field open-magnet magnetic resonance scannerInvest RadiolYear: 199833439439704282
3. Wildermuth S,Zanetti M,Duwell S,et al. Lumbar spine: quantitative and qualitative assessment of positional (upright flexion and extension) MR imaging and myelographyRadiologyYear: 199820739189577486
4. Schmid MR,Stucki G,Duewell S,et al. Changes in cross-sectional measurements of the spinal canal and intervertebral foramina as a function of body position: In vivo studies on an open-configuration MR systemAm J RoentgenolYear: 1999172109510210587155
5. Weishaupt D,Schmid MR,Zanetti M,et al. Positional MR imaging of the lumbar spine: Does it demonstrate nerve root compromise not visible at conventional MR imaging?RadiologyYear: 20002152475310751495
6. Jinkins JR,Dworkin J. Proceedings of the State-of-the-Art Symposium on Diagnostic and Interventional Radiology of the Spine, Antwerp, September 7, 2002 (Part two). Upright, weight-bearing, dynamic-kinetic MRI of the spine: pMRI/kMRIJBR-BTRYear: 2003862869314651085
7. Jinkins JR,Dworkin JS,Damadian RV. Upright, weight-bearing, dynamic-kinetic MRI of the spine: Initial resultEur RadiolYear: 20051518152515906040
8. Morishita Y,Hymanson H,Miyazaki M,et al. Kinematic evaluation of the spine: a kinetic magnetic resonance imaging studyJ Orthop SurgYear: 20081634850
9. Allmann K-H,Uhl M,Uhrmeister P,et al. Functional MR imaging of the cervical spine in patients with rheumatoid arthritisActa RadiolYear: 19983954369755705
10. Muhle C,Weinert D,Falliner A,et al. Dynamic Changes of the spinal canal in patients with cervical spondylosis at flexion and extension using magnetic resonance imagingInvest RadiolYear: 19983344499704283
11. Muhle C,Wiskirchen J,Weinert D,et al. Biomechanical aspects of the subarachnoid space and cervical cord in healthy individuals examined with kinematic magnetic resonance imagingSpineYear: 199823556679530787
12. Daffner SD,Xin J,Taghavi CE,et al. Cervical segmental motion at levels adjacent to disc herniation as determined with kinetic magnetic resonance imagingSpineYear: 20093423899419829252
13. Kawahara Y,Uetani M,Fuchi K,et al. MR assessment of movement and morphologic change in the menisci during knee flexionActa RadiolYear: 199940610410598848
14. Boxheimer L,Lutz AM,Treiber K,et al. MR imaging of the knee: Position related changes of the menisci in asymptomatic volunteersInvest RadiolYear: 2004392546315087719
15. Pfeiffer RP,DeBeliso M,Shea KG,et al. Kinematic MRI assessment of McConnell taping before and after exerciseAm J Sports MedYear: 200432621815090376
16. von Eisenhart-Rothe R,Bringmann C,Siebert M,et al. Femoro-tibial and menisco-tibial translation patterns in patients with unilateral anterior cruciate ligament deficiency – a potential cause of secondary meniscal tearsJ Orthop ResYear: 2004222758215013085
17. Hinterwimmer S,von Eisenhart-Rothe R,Siebert M,et al. Patella kinematics and patello-femoral contact areas in patients with genu varum and mild osteoarthritisClin BiomechYear: 20041970410
18. Komatsu T,Kadoya Y,Nakagawa S,et al. Movement of the posterior cruciate ligament during knee flexion – MRI analysisJ Orthop ResYear: 200523334915734245
19. Besier TF,Draper CE,Gold GE,et al. Patellofemoral joint contact area increases with knee flexion and weight-bearingJ Orthop ResYear: 2005233455015734247
20. Boxheimer L,Lutz AM,Zanetti M,et al. Characteristics of displaceable and nondisplaceable meniscal tears at kinematic MR imaging of the kneeRadiologyYear: 20062382213116373770
21. Takeda Y,Sato R,Ogawa T,et al. In vivo magnetic resonance imaging measurement of tibiofemoral relation with different knee flextion angles after single- and double-bundle anterior cruciate ligament reconstructionsArthroscopyYear: 2009257334119560637
22. Shellock FG,Feske W,Frey C,et al. Peroneal tendons: Use of kinematic MR imaging of the ankle to determine subluxationJ Magn Reson ImagingYear: 1997745149090608
23. Farooki S,Sokoloff RM,Theodorou DJ,et al. Visualization of ankle tendons and ligaments with MR imaging: Influence of passive positioningFoot Ankle IntYear: 200223554912095125
24. Tokuda O,Awaya H,Taguchi K,et al. Kinematic MRI of the normal ankle ligaments using a specially designed passive positioning deviceFoot Ankle IntYear: 2006279354217144956
25. Sans N,Richardi G,Railhac JJ,et al. Kinematic MR imaging of the shoulder: Normal PatternsAm J RoentgenolYear: 19961671517228956589
26. Graichen H,Stammberger T,Bonel H,et al. Glenohumeral translation during active and passive elevation of the shoulder – a 3D open-MRI studyJ BiomechYear: 2000336091310708782
27. Tokuda O,Kunihiro Y,Matsunaga N. Kinematic MRI of the normal shoulder using a specially designed positioning deviceJ Comput Assist TomogrYear: 2002268495412439327
28. Sahara W,Sugamoto K,Murai M,et al. 3D kinematic analysis of the acromioclavicular joint during arm abduction using vertically open MRIJ Orthop ResYear: 20062418233116838361
29. Sahara W,Sugamoto K,Murai M,et al. The three-dimensional motions of glenohumeral joint under semi-loaded condition during arm abduction using vertically open MRIClin BiomechYear: 20072230412
30. Ton ER,Pattynama PM,Bloem JL,et al. Interosseous ligaments: device for applying stress in wrist MR imagingRadiologyYear: 199519686347644657
31. Zanetti M,Hodler J,Gilula LA. Assessment of dorsal or ventral intercalated segmental instability configurations of the wrist: Reliability of sagittal MR imagesRadiologyYear: 1998206339459457184
32. Moritomo H,Goto A,Sato Y,et al. The triquetrum-hamate joint: An anatomic and in vivo three-dimensional kinematic studyJ Hand SurgYear: 200328797805
33. Goto A,Moritomo H,Murase T,et al. In vivo three-dimensional wrist motion analysis using magnetic resonance imaging and volume-based registrationJ Orthop ResYear: 200523750616022986
34. Goto T,Hamada K,Takeo Ito,et al. Interactive scan control for kinematic study in open MRIMgn Reson Med SciYear: 200762418
35. Antonio GE,Griffith JF,Yeung DKW. Small-field-of-view MRI of the knee and ankleAm J RoentgenolYear: 200418324815208102

Figures

[Figure ID: F0001]
Fig. 1 

The configuration of two types of magnetic resonance systems: (a) the superconducting, horizontally opened, 1.2 T magnetic resonance system and (b) the conventional tunnel-type, bored 1.5 T magnetic resonance system.



[Figure ID: F0002]
Fig. 2 

The postures in kinematic magnetic resonance imaging of the lumbar spine: (a) maximally anteflexed, (b) neutral, and (c) maximally retroflexed positions. A solenoid receiving coil was put on the subject's back (arrow).



[Figure ID: F0003]
Fig. 3 

The posture in kinematic magnetic resonance imaging of the knee. (a) The fixation of the knee by a supporting tool (CHAMCO, Inc., Cocoa, Florida, USA) and the arrangement of the solenoid receiving coil (arrow) are shown. The knee flexion can be adjusted from 0° to 120°. (b) The subject's position during the image acquisition is shown.



[Figure ID: F0004]
Fig. 4 

The posture in kinematic magnetic resonance imaging of the ankle. (a) The fixation of the ankle by a supporting tool (CHAMCO, Inc., Cocoa, Florida, USA) and the arrangement of the solenoid receiving coil (arrow) are shown. (b) The subject's position during the image acquisition is shown.



[Figure ID: F0005]
Fig. 5 

Example of kinematic magnetic resonance images of the lumbar spine. Upper row: slices at the middle plane of the vertebral body. Lower row: slices at the level of the intervertebral foramen. Each of them is arranged from left to right in order of maximal anteflexion, neutral, and maximal retroflexion positions.



[Figure ID: F0006]
Fig. 6 

Example of kinematic magnetic resonance images of the knee. Upper row: slices at the level of the anterior cruciate ligament (ACL). Middle row: slices at the level of the posterior cruciate ligament (PCL). Lower row: slices at the level of the medial meniscus (MM). The angle indicated at the lower left of each image is the flexion angle of the knee.



[Figure ID: F0007]
Fig. 7 

Example of kinematic magnetic resonance images of the ankle. Upper row: slices at the central trochlea of the talus (TT). Lower row: slices at the level of the peroneal tendon (PT). Each of them is arranged in order from left to right of maximal dorsiflexion, neutral, and maximal plantarflexion positions.



Article Categories:
  • Original Article

Keywords: Joint Motion, Magnetic Resonance Imaging, Kinematic, Open MRI, KMRI.

Previous Document:  How effective is sun salutation in improving muscle strength, general body endurance and body compos...