| The newly designed univalved artificial heart. | |
| | |
MedLine Citation:
|
PMID: 1751128 Owner: NLM Status: MEDLINE |
Abstract/OtherAbstract:
|
A univalved artificial heart (AH) powered electromagnetically in the frequency range of 1-30 Hz was developed to obtain a totally implantable AH. This small sized AH consisted of a vibrating tube, coils, magnets, and a jellyfish valve as an outlet AH valve. The fluid mechanical, hemodynamic, and hematologic properties were evaluated in a mock circulation and 10 animal experiments using adult goats. This vibrating electromagnetic AH could generate more than 10 L/min as an output volume, with 10 Hz vibration using 20 volts supplied voltage. It could also provide two kinds of flow and pressure patterns, in constant-peak and periodic-peak patterns. The values of free hemoglobin remained within acceptable limits. The authors concluded that this new type of pump was useful as a totally implantable AH, ventricular assist, or organ assist system. |
| | |
Authors:
|
S Nitta; H Hashimoto; T Sonobe; Y Katahira; T Yambe; S Naganuma; M Tanaka; N Sato; M Miura; H Mohri |
Related Documents
:
|
11482478 - Continuous venovenous renal replacement therapy using a conventional infusion pump. 1751178 - Comparison of an extraaortic counterpulsation device versus intraaortic balloon pumping... 1640528 - Pitfalls in determination of leak point pressure. 19416768 - Assessment of arterial blood pressure during support with an axial flow left ventricula... 10234108 - Exercise blood pressure predicts cardiovascular death and myocardial infarction. 23150518 - Pulsatile hemodynamics and microcirculation: evidence for a close relationship in hyper... |
Publication Detail:
|
Type: Journal Article |
Journal Detail:
|
Title: ASAIO transactions / American Society for Artificial Internal Organs Volume: 37 ISSN: 0889-7190 ISO Abbreviation: ASAIO Trans Publication Date: 1991 Jul-Sep |
Date Detail:
|
Created Date: 1992-01-29 Completed Date: 1992-01-29 Revised Date: 2008-11-21 |
Medline Journal Info:
|
Nlm Unique ID: 8611947 Medline TA: ASAIO Trans Country: UNITED STATES |
Other Details:
|
Languages: eng Pagination: M240-1 Citation Subset: IM |
Affiliation:
|
Department of Medical Engineering and Cardiology, Tohoku University, Sendai, Japan. |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
Blood Flow Velocity
/
physiology Blood Pressure / physiology Electric Power Supplies Electromagnetic Phenomena / instrumentation Heart Valve Prosthesis* Heart, Artificial* Heart-Assist Devices* Humans Prosthesis Design |
From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine
Previous Document: A unique, efficient, implantable, electromechanical, total artificial heart.