| A newly designed arterial monitoring/perfusion cannula for cardiac operations. | |
| | |
MedLine Citation:
|
PMID: 299806 Owner: NLM Status: MEDLINE |
Abstract/OtherAbstract:
|
A newly designed arterial monitoring and perfusion cannula for cardiopulmonary bypass eliminates the need for cannulation of a peripheral artery for pressure monitoring. The double-lumen cannula is designed as follows: the large central lumen (12 to 26f) acts as the arterial inflow conduit from the pump oxygenator, while a second, smaller lumen (18 gauge)constructed in the wall of the first cannula acts as the pressure-monitoring port and the source for blood sampling and drug infusion. This monitoring/perfusion cannula has been used succesfully in more than 250 clinical patients in a variety of settings--total cardiopulmonary bypass, left heart bypass, and when multiple arterial inflow lines were necessary (as in aortic arch replacement). Use of this cannula is advantageous in the infant and pediatric patient or in the emergency setting, when insertion of a peripheral arterial line can be difficult. Used in conjunction with a Doppler system, the cannula provides accurate, dependable blood pressure monitoring. |
| | |
Authors:
|
S J Phillips; J E Okies; R H Zeff |
Related Documents
:
|
3722076 - Vacuum-assisted perfusion of the dog hindlimb. 8513386 - A new sodium-nitroprusside-infusion controller for the regulation of arterial blood pre... 3974126 - Interaction between heart as a pump and artery as a load. 10222916 - Dynamic pressure measurement for detection of blood access stenosis. 12647236 - Chest wall kinematics during chemically stimulated breathing in healthy man. 1885656 - The influence of pressure on actin and myosin interactions in solution and in single mu... |
Publication Detail:
|
Type: Journal Article |
Journal Detail:
|
Title: The Annals of thoracic surgery Volume: 23 ISSN: 0003-4975 ISO Abbreviation: Ann. Thorac. Surg. Publication Date: 1977 Jan |
Date Detail:
|
Created Date: 1977-02-24 Completed Date: 1977-02-24 Revised Date: 2004-11-17 |
Medline Journal Info:
|
Nlm Unique ID: 15030100R Medline TA: Ann Thorac Surg Country: UNITED STATES |
Other Details:
|
Languages: eng Pagination: 80-2 Citation Subset: AIM; IM |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
Blood Pressure Cardiac Surgical Procedures* Cardiopulmonary Bypass / instrumentation* Catheterization / instrumentation* Coronary Artery Bypass / instrumentation Humans Perfusion / instrumentation |