The influence of inserting a Fuji pressure sensitive film between the tibiofemoral joint of knee prosthesis on actual contact characteristics.
MedLine Citation:
PMID:  11222935     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
OBJECTIVE: To investigate the influence of inserting a Fuji pressure sensitive film between the tibiofemoral joint of knee prosthesis on actual contact characteristics. DESIGN: A finite element analysis was used to investigate the alteration of contact characteristics of the tibiofemoral joint due to inserting a pressure sensitive film between the contacting surfaces. BACKGROUND: The discrepancy between actual contact behaviors of tibiofemoral joint of knee prosthesis and the measurement using Fuji pressure sensitive film was not discussed extensively. The change of direct contact circumstance of the tibiofemoral joint due to inserting a pressure sensitive film was not well reported. METHOD: A two-dimensional finite element model of the tibiofemoral joint of knee prosthesis in the sagittal plane was constructed. Four different radii of the femoral component were used to investigate the conformity effect. Two-layers of plane strain element were used to model the pressure sensitive film. The influence of inserting a pressure sensitive film on the actual contact characteristics was analyzed by comparing the results of the change in contact radius. RESULTS: Inserting a pressure sensitive film between contacting surfaces would disturb the original contact behaviors, especially in the lowest conformity design. The maximum difference of contact radius between the model simulating actual contact circumstance and the model with inserting a pressure sensitive film between contacting surface was 77% in the lowest conformity condition at the smallest load. CONCLUSIONS: This study proposes a quantitative analysis of contact characteristics in the tibiofemoral joint of knee prosthesis between the models with and without inserting a pressure sensitive film into the contact surface. The measurement of contact area in artificial tibiofemoral joints by using Fuji pressure sensitive film is always overestimated its true contact area by 14-77%. RELEVANCE: This study revealed the measurement of contact characteristics of artificial tibiofemoral joints by using Fuji pressure sensitive film which depends on not only the applied load, but also the conformity and material properties of the contact surface. Therefore, the information of the conformity and material properties of the contact surface should be provided as well as the applied load wherever a measurement of contact characteristics using Fuji pressure sensitive film is analyzed and interpreted.
Authors:
J J Liau; C C Hu; C K Cheng; C H Huang; W H Lo
Related Documents :
15676575 - Intra-abdominal pressure responses of mineworkers to standard loads.
21914985 - Errors of oscillometric blood pressure measurement as predicted by simulation.
14729985 - Changes in plantar foot pressure with in-shoe varus or valgus wedging.
14558645 - Analysis of contact mechanics in mckee-farrar metal-on-metal hip implants.
19843055 - The effects of urotensin ii and urantide on forearm blood flow and systemic haemodynami...
9328795 - Doxazosin suppresses the morning increase in blood pressure and sympathetic nervous act...
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Clinical biomechanics (Bristol, Avon)     Volume:  16     ISSN:  0268-0033     ISO Abbreviation:  Clin Biomech (Bristol, Avon)     Publication Date:  2001 Feb 
Date Detail:
Created Date:  2001-03-06     Completed Date:  2001-06-07     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  8611877     Medline TA:  Clin Biomech (Bristol, Avon)     Country:  England    
Other Details:
Languages:  eng     Pagination:  160-6     Citation Subset:  IM    
Affiliation:
Orthopaedic Biomechanics Laboratory, Institute of Biomedical Engineering, National Yang Ming University, No. 155, Sec. 2, Li-Nung St., Shih-Pai, ROC, Taipei, Taiwan.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Biomechanics
Finite Element Analysis
Humans
Knee Joint / physiology
Knee Prosthesis*