Document Detail


The influence of acetabular cup material on pelvis cortex surface strains, measured using digital image correlation.
MedLine Citation:
PMID:  22236529     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
Acetabular cup loosening is a late failure mode of total hip replacements, and peri-prosthetic bone deterioration may promote earlier failure. Preservation of supporting bone quality is a goal for implant design and materials selection, to avoid stress shielding and bone resorption. Advanced polymer composite materials have closer stiffness to bone than metals, ceramics or polymers, and have been hypothesised to promote less adverse bone adaptation. Computer simulations have supported this hypothesis, and the present study aimed to verify this experimentally. A composite hemi-pelvis was implanted with Cobalt Chromium (CoCr), polyethylene (UHMWPE) and MOTIS(®)carbon-fibre-reinforced polyether etherketone (CFR-PEEK) acetabular cups. In each case, load was applied to the implanted pelvis and Digital Image Correlation (DIC) was used for surface strain measurement. The test was repeated for an intact hemi-pelvis. Trends in implanted vs. intact bone principal strains were inspected to assess the average principal strain magnitude change, allowing comparison of the potential bone responses to implantation with the three cups. The CFR-PEEK cup was observed to produce the closest bone strain to the intact hip in the main load path, the superior peri-acetabular cortex (+12% on average, R(2)=0.84), in comparison to CoCr (+40%, R(2)=0.91) and UHWMPE cups (-26%, R(2)=0.94). Clinical observations have indicated that increased periacetabular cortex loading may result in reduced polar cancellous bone loading, leading to longer term losses in periprosthetic bone mineral density. This study provides experimental evidence to verify previous computational studies, indicating that cups produced using materials with stiffness closer to cortical bone recreate physiological cortical bone strains more closely and could, therefore, potentially promote less adverse bone adaptation than stiffer press-fitted implants in current use.
Authors:
A S Dickinson; A C Taylor; M Browne
Related Documents :
21054639 - Prosthodontic rehabilitation of a shotgun injury: a patient report.
21778669 - Effect of trehalose coating on basic fibroblast growth factor release from tailor-made ...
21365039 - Bone mineral apposition rates at early implantation times around differently prepared t...
22246049 - Shear bond strength and enamel fracture behavior of ceramic brackets fascination® and ...
21279959 - Choosing the optimum swab for canal drying in cemented total hip replacement.
22208909 - Photoelastic analysis of stress distribution with different implant systems.
18717379 - Sinus floor elevation using an osteotome technique without grafting materials or membra...
21725259 - Speech detection in noise and spatial unmasking in children with simultaneous versus se...
264339 - An attempt to correlate dentin and pulp changes in human carious teeth.
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2012-1-9
Journal Detail:
Title:  Journal of biomechanics     Volume:  -     ISSN:  1873-2380     ISO Abbreviation:  -     Publication Date:  2012 Jan 
Date Detail:
Created Date:  2012-1-12     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  0157375     Medline TA:  J Biomech     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Copyright Information:
Copyright © 2011 Elsevier Ltd. All rights reserved.
Affiliation:
University of Southampton, Southampton, UK; Aurora Medical Ltd., Chilworth, UK.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Elastic modulus of hard tissues.
Next Document:  Planar biaxial characterization of diseased human coronary and carotid arteries for computational mo...