An in vitro study of silk stent morphology.
MedLine Citation:
PMID:  20981414     Owner:  NLM     Status:  In-Data-Review    
Abstract/OtherAbstract:
INTRODUCTION: Morphology of the Silk stent (Balt, Montmorency, France) after deployment is not fully understood, especially in tortuous vessels. An in vitro study was conducted to study morphology and flow-diverting parameters of this stent.
METHODS: Two sets of different-sized and curved polytetrafluoroethylene tubes were studied. To simulate the aneurysm neck, a small hole was created in a tube. A stent was placed in each of the different tubes. Angiographic computerized tomography and macroscopic photography were then obtained. The images were analyzed to calculate a Percentage of Area Coverage (PAC).
RESULTS: Good stent conformability was observed. The PAC was 21% in the straight model with matched stent and vessel diameter. In the straight model with an oversized stent, the PAC was increased. In the curved models, dynamic wire repositioning occurred. The repositioning was affected by the size of the stent and the angle of the vessel curve. Compared to the straight model, this increased the PAC in two instances: on the convexity (oversized stent), and on the concavity (matched stent and vessel diameter). The PAC did not significantly change at the sides of the curve.
CONCLUSIONS: By design, the wires of the silk stent move relative to each other. In a curved model, the PAC is different at the convexity, concavity, and lateral walls. The stent diameter affects the PAC. These results are clinically relevant because it is desirable to maximize and minimize the PAC across the aneurysm neck and branch vessel orifice, respectively.
Authors:
Thaweesak Aurboonyawat; Raphaël Blanc; Paul Schmidt; Michel Piotin; Laurent Spelle; Amir Nakib; Jacques Moret
Related Documents :
8993814 - Intravascular ultrasound identification of stent entrapment in vivo with in vitro confi...
15597014 - Flat-panel detector computed tomography for the assessment of coronary artery stents: p...
18580804 - Preliminary experience with the enterprise stent for endovascular treatment of complex ...
1615944 - Fungemia with torulopsis glabrata after endoscopic biliary stent replacement.
20502414 - Expression and localization of vascular endothelial growth factor in normal abdominal a...
21328684 - Acute coronary occlusion following tavi.
Publication Detail:
Type:  Journal Article     Date:  2010-10-28
Journal Detail:
Title:  Neuroradiology     Volume:  53     ISSN:  1432-1920     ISO Abbreviation:  Neuroradiology     Publication Date:  2011 Sep 
Date Detail:
Created Date:  2011-08-16     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  1302751     Medline TA:  Neuroradiology     Country:  Germany    
Other Details:
Languages:  eng     Pagination:  659-67     Citation Subset:  IM    
Affiliation:
Department of Functional and Interventional Neuroradiology, Fondation Rothschild Hospital, University of Paris 7 Bichat School of Medicine, Paris, France, aurboonyawat@gmail.com.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Cellular correlates of anxiety in CA1 hippocampal pyramidal cells of 5-HT1A receptor knockout mice.
Next Document:  Development of laminar organization of the fetal cerebrum at 3.0T and 7.0T: a postmortem MRI study.