An improved device for posterior rhinomanometry to measure nasal resistance.
MedLine Citation:
PMID:  16438238     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
Rhinomanometry is a method for measuring nasal resistance for the purpose of providing an objective evaluation of nasal patency. Posterior rhinomanometry is accomplished without the use of a mask, thus allowing the patient to breathe naturally. Here, we report on the improvements we have made to the existing posterior rhinomanometry system. In this system, nasal airflow is measured indirectly by measuring the pressure differential across a small mesh window in the body plethysmograph. We have calibrated this measurement and developed software that automatically provides the correct values for all airflow rates. Also, we have developed software that automatically corrects for the phase shift caused by the plethysmograph structure. These refinements should provide more accurate values for nasal resistance.
Authors:
Joseph Tiran; Nir Ben-Oved; Elvaiah Elan; Youval Slovik; Alberto Leiberman; Aviva Weisel-Eichler; Daniel M Kaplan
Related Documents :
19659518 - Effects of simulated tidal and deep breathing on immature airway contraction to acetylc...
8864428 - Impact of central, obstructive and mixed apnea on cerebral hemodynamics in preterm infa...
371648 - Cardiorespiratory effects of increased airway pressure during controlled and spontaneou...
12832428 - Upper airway extraluminal tissue pressure fluctuations during breathing in rabbits.
6283918 - Coronary blood flow changes following activation of adrenergic receptors in the conscio...
3301358 - Nonpharmacologic treatment of hypertension: does it work?
Publication Detail:
Type:  Clinical Trial; Comparative Study; Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Journal of biomechanical engineering     Volume:  127     ISSN:  0148-0731     ISO Abbreviation:  J Biomech Eng     Publication Date:  2005 Nov 
Date Detail:
Created Date:  2006-01-27     Completed Date:  2006-03-31     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  7909584     Medline TA:  J Biomech Eng     Country:  United States    
Other Details:
Languages:  eng     Pagination:  994-7     Citation Subset:  IM    
Affiliation:
Department of Mechanical Engineering, Pearlstone Center for Aeronautical Engineering Studies. tiran@bgumail.bgu.ac.il
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Airway Resistance / physiology*
Diagnosis, Computer-Assisted / instrumentation*,  methods*
Equipment Design
Equipment Failure Analysis
Female
Humans
Male
Manometry / instrumentation*,  methods
Nose / physiology*
Plethysmography / instrumentation*,  methods
Reproducibility of Results
Sensitivity and Specificity
Transducers, Pressure*