Document Detail

An improved delivery system for bladder irrigation.
Jump to Full Text
MedLine Citation:
PMID:  20957138     Owner:  NLM     Status:  PubMed-not-MEDLINE    
Abstract/OtherAbstract:
INTRODUCTION: Occasionally, urologists may see patients requiring temporary bladder irrigation at hospitals without stocks of specialist irrigation apparatus. One option is to transfer the patient to a urology ward, but often there are outstanding medical issues that require continued specialist input. Here, we describe an improved system for delivering temporary bladder irrigation by utilizing readily available components and the novel modification of a sphygmomanometer blub. This option is good for bladder irrigation in patients with moderate or severe gross hematuria due to various causes.
MATERIALS AND METHODS: In this prospective study from March 2007 to April 2009, we used our new system in eligible cases. In this system, an irrigant bag with 1 L of normal saline was suspended 80 cm above the indwelled 3-way Foley catheter, and its drainage tube was inserted into the irrigant port of the catheter. To increase the flow rate of the irrigant system, we inserted a traditional sphygmomanometer bulb at the top of the irrigant bag. This closed system was used for continuous bladder irrigation (CBI) in patients who underwent open prostatectomy, transurethral resection of the prostate (TURP), or transurethral resection of the bladder (TURB). This high-pressure system is also used for irrigation during cystourethroscopy, internal urethrotomy, and transurethral lithotripsy. Our 831 eligible cases were divided into two groups: group 1 were endourologic cases and group 2 were open prostatectomy, TURP, and TURB cases. The maximum and average flow rates were evaluated. The efficacy of our new system was compared prospectively with the previous traditional system used in 545 cases.
RESULTS: In group 1, we had clear vision at the time of endourologic procedures. The success rate of this system was 99.5%. In group 2, the incidence of clot retention decreased two fold in comparison to traditional gravity-dependent bladder flow system. These changes were statistically significant (P = 0.001). We did not observe any adverse effects such as bladder perforation due to our high-pressure, high-flow system.
CONCLUSION: A pressurized irrigant system has better visualization during endourologic procedures, and prevents clot formation after open prostatectomy, TURP, and TURB without any adverse effects.
Authors:
Mohammad K Moslemi; Mojtaba Rajaei
Related Documents :
6876268 - Ileal replacement of the bilharzial ureter: is it worthwhile?
12846628 - A pilot study of the effect of the queen's square external bladder stimulator on urinar...
12900008 - Orthotopic bladder reconstruction in women--what we have learned over the last decade.
11743338 - Congenital adrenal hyperplasia: preliminary observations of the urethra in 9 cases.
9195618 - Postradiation osteonecrosis of the mandible: a long-term follow-up study.
1213658 - Extended somatostatin treatment of a patient with bleeding ulcer.
Publication Detail:
Type:  Journal Article     Date:  2010-10-05
Journal Detail:
Title:  Therapeutics and clinical risk management     Volume:  6     ISSN:  1178-203X     ISO Abbreviation:  Ther Clin Risk Manag     Publication Date:  2010  
Date Detail:
Created Date:  2010-10-19     Completed Date:  2011-03-17     Revised Date:  2013-05-29    
Medline Journal Info:
Nlm Unique ID:  101253281     Medline TA:  Ther Clin Risk Manag     Country:  New Zealand    
Other Details:
Languages:  eng     Pagination:  459-62     Citation Subset:  -    
Affiliation:
Department of Urology, Kamkar Hospital, School of Medicine, Qom University of Medical Sciences, Bajak Ave, Qom, 3715694978, Iran. moslemi_urologist@yahoo.com; moslemi@muq.ac.ir
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine

Full Text
Journal Information
Journal ID (nlm-ta): Ther Clin Risk Manag
Journal ID (publisher-id): Therapeutics and Clinical Risk Management
ISSN: 1176-6336
ISSN: 1178-203X
Publisher: Dove Medical Press
Article Information
Download PDF
© 2010 Moslemi and Rajaei, publisher and licensee Dove Medical Press Ltd.
License:
Electronic publication date: Day: 5 Month: 10 Year: 2010
collection publication date: Year: 2010
Print publication date: Year: 2010
Volume: 6First Page: 459 Last Page: 462
PubMed Id: 20957138
DOI: 10.2147/TCRM.S13525
Publisher Id: tcrm-6-459

An improved delivery system for bladder irrigation
Mohammad K Moslemi1
Mojtaba Rajaei2
1 Department of Urology, Kamkar Hospital, School of Medicine, Qom University of Medical Sciences, Qom, Iran
2 Kamkar Hospital, School of Medicine, Qom University of Medical Sciences, Qom, Iran
Correspondence: Correspondence: Mohammad K Moslemi, Department of Urology, Kamkar Hospital, School of Medicine, Qom University of Medical Sciences, Bajak Ave, Qom, 3715694978, Iran, Tel +98 912 252 1646, Fax +98 251 771 3473, Email moslemi_urologist@yahoo.com; moslemi@muq.ac.ir

Introduction

Three-way Foley catheters are used when irrigation of the bladder is anticipated to prevent or manage blood clots in the bladder. Effective bladder irrigation influences patient safety and clinical outcomes.1,2 Continuous bladder irrigation (CBI) is commonly used after transurethral resection of the prostate (TURP). The incidence of clot retention after TURP was reported to be 3.3% in a large, multicenter retrospective study.3 A high risk of postoperative bleeding, and thus clot retention, has been correlated with a greater size of the resected prostate, longer operative time, and positive preoperative urine culture.46 CBI is also used for intravesical delivery of pharmacologic agents, such as amphotericin B irrigation for candida cystitis.79 CBI with a 3-way Foley catheter is an important method of preventing and treating the hematuria associated with chemotherapy, vesical tumors, hemorrhagic cystitis, and radiation cystitis.


Materials and methods

In this 2-year prospective study conducted between March 2007 and April 2009, our new delivery system was used routinely for our admitted patients. Before this period, the traditional gravity system was used routinely for all 545 admitted cases, of which, 234 cases were underwent open prostatectomy, TURP, and transurethral resection of the bladder (TURB), and 311 cases underwent an endourologic procedure such as cystoscopy, ureteroscopy, or transurethral lithotripsy (TUL). In this system, a plastic connector was inserted into the irrigant bag with 1 L of normal saline, and its drainage terminal was connected to a Foley catheter or an endourologic instrument such as a cystoscope or ureteroscope. This system works only with gravity. It is clear that the drainage velocity of the irrigant fluid depends on the height of the irrigant bag.

The 831 eligible cases for our new system were divided into two groups who used our instrument in the same way in two situations: the first group underwent endourologic procedures such as cystourethroscopy, internal urethrotomy, and ureteroscopy, or some situations in the TUL procedures (458 cases); the second group used the system to increase flow rate and bladder irrigation after open prostatectomy, TURP, and TURB (373 cases). The duration of use for the new system depends on the severity of bleeding. When bleeding is increasingly severe, or for a more clear vision, we increased flow rate by hand pumping a sphygmomanometer bulb to increase pressure in the irrigant bag ( Figures 13). This hand pumping is performed by the scrub nurse in the operating theater or recovery room. After clearing the drainage fluid, the irrigant system was changed to the gravity flow system. Analysis of data was performed with SPSS software (version 16; SPSS Inc, Chicago, IL).


Results

The major issue in the endourologic procedures is clear vision. This new system was used in 458 endourologic cases. We had two cases (0.43%) of bloody or unclear vision in group 1, while in group 2 t this problem was seen in 93 (29.9%) of 311 cases. Clot retention episodes in group 2 decreased from 10 of 234 (4%) to 6 of 321 (1.8%) cases (Table 1). These differences were statistically significant (P = 0.001). The maximum and average flow rate of the system were evaluated. The gravity-dependent flow rate of an irrigant system suspended at 80 cm is 3 mL/sec, whereas with the use of our new delivery system the flow rate increased to 7.5 mL/sec at maximal hand pressure.


Discussion

Using 3-way Foley catheters is an essential part of routine urological practice. They are available in different sizes: 20, 22, or 24 F. In our daily practice and in the presence of significant bleeding, a size of 22 F catheter or greater is used. Three-way catheters require good irrigation and drainage flow characteristics to optimize a continuous flow system. All catheters with the size of 22 and 24 F have equivalent irrigation and drainage properties. The flow rate of the various 3-way catheters assessed is proportional to the size of the catheter when the drainage port is used for irrigation. This does not hold true when the irrigation channel is used. Larger catheter size does not equate to better irrigation or drainage when continuous irrigation is performed.10 The disadvantage of 3-way catheters is that they provide a smaller lumen than do 2-way catheters. 11 However, in a CBI system, the flow of irrigation should determine the drainage directly. For this reason, we designed a new high-flow, high pressure, hand pumped delivery system. Whitaker12 emphasized the flawed design of many available 3-way catheters, claiming that the input tube is compromised to achieve a wider output tube. He stressed that more efficient drainage is rooted in the irrigation flow rate because faster flow prevents blood clots from forming and accumulating in the bladder. Only small increases in the size of the irrigation lumen can double the irrigation flow rate without a significant decrease in the drainage tube lumen. As such, increases in flow are governed by Poiseuille’s law, which states that there is a strong (fourth power) dependence of flow on the radius of the channel. Large-bore catheters are considered the best choice for effective removal of blood clots.13 Because 3-way catheters establish a closed system with the bladder, drainage cannot be maximized if irrigation is severely lagging behind drainage. As such, it is meaningless to have good drainage properties if irrigation is insufficient. CBI typically takes place over the course of several hours or days, which makes average irrigation and drainage flow data more relevant than maximal flow data.2


Conclusion

The early hours after prostate surgery (TURP and open prostatectomy) are critical hours that need close observation because of bleeding and the risk of clot formation. Our new irrigation device generates superior flow in comparison to a traditional gravity-assisted irrigation system. This device is useful in situations in which efficient bladder irrigation is essential for patient safety. Our new device optimizes continuous bladder irrigation. It is an easily applicable system for bladder irrigation in patients with gross hematuria, and it may be used in remote or rural hospitals where special urologic equipment is not available.


Notes

Disclosure

Mojtaba Rajaei is the designer and inventor of the system.

References
1. Walton TJ,Heads S,Parkinson RJ,Griffiths TRL. An improvised delivery system for irrigationAnn R Coll Surg EnglYear: 200890542943018642424
2. Braasch M,Antolac C,Hendlin K,et al. Irrigation and drainage properties of three-way urethral cathetersUrologyYear: 2006671404416413329
3. Mebust WK,Holtgrewe HL,Cockett AT,Peters PC. Transurethral prostatectomy: immediate and postoperative complications: a cooperative study of 13 participating institutions evaluating 3,885 patients. 1989J UrolYear: 20021412432472643719
4. Mobb GE,Farrar DJ. Is planned continuous irrigation indicated for haemorrhage following transurethral resection of the prostateBr J UrolYear: 1993717077108343899
5. Olapade-Olaopa EO,Solomon LZ,Carter CJ,Ahiaku EK,Chiverton SG. Haematuria and clot retention after transurethral resection of the prostate: a pilot studyBr J UrolYear: 1998826246279839574
6. El Malik EM,Ibrahim AI,Gahli AM,Saad MS,Bahar YM. Risk factors in prostatectomy bleeding: preoperative urinary infection is the only reversible factorEur UrolYear: 20003719920410705199
7. Kauffman CA,Vazquez JA,Sobel JD,et al. Prospective multicenter surveillance study of funguria in hospitalized patients. The National Institute for Allergy and Infectious Diseases (NIAID) Mycoses Study GroupClin Infect DisYear: 200030141810619726
8. Goldman HJ,Littman ML,Oppenheimer GD,et al. Monilial cystitis-effective treatment with instillations of amphotericin BJAMAYear: 196017435936213706544
9. Jacobs LG,Skidmore EA,Freeman K,et al. Oral fluconazole compared with bladder irrigation with amphotericin B for treatment of fungal urinary tract infections in elderly patientsClin Infect DisYear: 19962230358824962
10. Manikandan R,Selvaratnam V,Philip J,Hanlon A,Machin DG,Williamson M. Evaluation of flow characteristics of 3-way cathetersJ UrolYear: 200918141922192519237172
11. Cro DJ. A simpler method for continuous bladder irrigationUrology Year: 198220110
12. Whitaker RH. A look at the three-way catheterBr J Urol Year: 1975471031131493
13. Choong SKS,Walkden M,Kirby R. The management of intractable haematuriaBJU IntYear: 20008695195911119085

Article Categories:
  • Original Research

Keywords: cystoscopy, drainage, pressure, open prostatectomy, sphygmomanometer, transurethral lithotripsy, transurethral resection, prostate, bladder.

Previous Document:  Gastrointestinal stromal tumor (GIST) recurrence following surgery: review of the clinical utility o...
Next Document:  Pegylated liposomal doxorubicin in the management of ovarian cancer.