Document Detail


A hand-held device to measure oxygen uptake: performance characteristics, patient selection and the propagation of its measurement error into fick cardiac output determinations.
MedLine Citation:
PMID:  17341778     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
Measured oxygen consumption rates (VO2) are being used less frequently for direct Fick cardiac output determinations during cardiac catheterization because the instruments that measure VO2 are usually cumbersome, expensive and difficult to use. A small, inexpensive VO2 device recently became available. Therefore, the aims of this report are: (1) to delineate the advantages and limitations that would affect its use during cardiac catheterization; (2) to define the patient population in which its use would be safe and effective; and (3) to estimate the accuracy of cardiac output determinations based on its VO2 measurements. The MedGem is a fist-size device that measures VO2 as a patient breathes through it using either a facemask or mouthpiece. We evaluated it by measuring the external deadspace it imposes on the patient, by measuring the time required for each measurement, by determining its precision from repeated measurements on healthy volunteers and animals, and by developing a statistical model that calculates cardiac output error as a function of the random errors in the measurements of arterial and venous oxygen concentrations and VO2. Our experience indicates the MedGem is easy to operate and can be used for a wide spectrum of patients at rest including children with the ventilatory capacity to compensate for the added deadspace. Cardiac output determinations made with the MedGem's measurements of oxygen uptake are clinically acceptable if the oximeter used to measure oxygen in arterial and mixed venous blood is also sufficiently accurate. Our statistical model illustrates strategies to minimize cardiac output error.
Authors:
Albert P Shepherd; Beth M Terpolilli; John M Steinke
Related Documents :
1269098 - Plasma renin activity during exercise in the dog.
15487268 - Comparison of hemodynamics during hyperthermal immersion and exercise testing in appare...
1729108 - Comparison of cardiac output determinants in response to upright and supine exercise in...
6930878 - Evaluation of cardiac function in patients with thalassemia major.
23901268 - The 6-minute pegboard and ring test is correlated with upper extremity activity of dail...
2382898 - Effects of humid air breathing during arm or treadmill exercise on exercise-induced bro...
Publication Detail:
Type:  Evaluation Studies; Journal Article    
Journal Detail:
Title:  The Journal of invasive cardiology     Volume:  19     ISSN:  1557-2501     ISO Abbreviation:  J Invasive Cardiol     Publication Date:  2007 Mar 
Date Detail:
Created Date:  2007-03-07     Completed Date:  2007-03-30     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  8917477     Medline TA:  J Invasive Cardiol     Country:  United States    
Other Details:
Languages:  eng     Pagination:  113-22     Citation Subset:  IM    
Affiliation:
Department of Physiology, University of Texas Health Science Center, San Antonio, TX 78229-3900, USA. Shepherd@uthscsa.edu
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Cardiac Output*
Diagnostic Errors*
Diagnostic Techniques, Cardiovascular / instrumentation*
Equipment Design
Equipment and Supplies / adverse effects,  standards
Humans
Oxygen Consumption*
Patient Selection*
Reproducibility of Results
Respiratory Dead Space
Time Factors

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Examination of anti-intima hyperplastic effect on cilostazol-eluting stent in a porcine model.
Next Document:  Impact of gender on in-hospital percutaneous coronary interventional outcomes in African-Americans.