Document Detail


A fatal device-device interaction between a wearable automated defibrillator and a unipolar ventricular pacemaker.
MedLine Citation:
PMID:  18684292     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
We report a fatal device-device interaction between a wearable automated defibrillator (WAD; LifeVest - LifeCor, Inc., Pittsburgh, PA, USA) and a unipolar pacemaker that occurred in an 18-year-old patient listed for cardiac transplantation due to his failing Fontan. The patient developed ventricular tachycardia that was initially detected by the WAD. However, large unipolar pacing artifacts and specific WAD arrhythmia detection algorithms caused the WAD to revert to nonrecognition of the arrhythmia, which lead to the patient's death. We identify likely causes of the failure and suggest methods of preventing such occurrences in the future.
Authors:
Martin J LaPage; Charles E Canter; Edward K Rhee
Publication Detail:
Type:  Case Reports; Journal Article    
Journal Detail:
Title:  Pacing and clinical electrophysiology : PACE     Volume:  31     ISSN:  1540-8159     ISO Abbreviation:  Pacing Clin Electrophysiol     Publication Date:  2008 Jul 
Date Detail:
Created Date:  2008-08-07     Completed Date:  2008-10-30     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  7803944     Medline TA:  Pacing Clin Electrophysiol     Country:  United States    
Other Details:
Languages:  eng     Pagination:  912-5     Citation Subset:  IM    
Affiliation:
Washington University School of Medicine, Department of Pediatrics, Division of Pediatric Cardiology, St. Louis, Missouri 63110, USA. lapage_m@kids.wustl.edu
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adolescent
Defibrillators / adverse effects*
Equipment Failure*
Fatal Outcome
Humans
Male
Medical Errors / adverse effects*,  instrumentation*,  prevention & control
Pacemaker, Artificial / adverse effects*
Tachycardia, Ventricular / etiology*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Cardiac resynchronization after left ventricular lead extraction: usefulness of angioplasty in coron...
Next Document:  A case of catecholaminergic polymorphic ventricular tachycardia caused by two calsequestrin 2 mutati...