Document Detail


The effects of radiofrequency energy probe speed and application force on chondrocyte viability.
MedLine Citation:
PMID:  17364094     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
OBJECTIVE: To determine the thermal effects of monopolar radiofrequency energy (mRFE) on bovine articular cartilage when it was moved at different speeds and using varying application forces. METHODS: Thirty-six fresh osteochondral sections divided into two groups (18 sections/group) were used in this study. The first group was tested at three speed rates of mRFE probe (1 mm/sec, 5 mm/sec and 10 mm/sec) at a constant force (50 g) applied to the probe tip. In the second group, three application forces of the probe tip were tested (25 g, 50 g and 75 g) at a constant speed (5 mm/sec) (n = 6/test). All tests were performed using a custom-built jig to control the mRFE (Vulcan EAS) probe during a 20-mm pass on each section. After treatment, viability of osteochondral sections was determined by confocal laser microscopy (CLM) combined with vital cell staining. RESULTS: There were not any significant differences in cartilage thickness of tested osteochondral sections among the different speeds or forces. During the mRFE probe treatments at different speeds, CLM demonstrated that probe application at the speed of 1 mm/sec caused significantly greater chondrocyte death than at the speeds of 5 and 10 mm/sec, whereas there were no significant differences in chondrocyte death among the variable application forces (p > 0.05). DISCUSSION: This in vitro study demonstrated that RFE thermal penetration correlated most closely with probe application speed than application force for this mRFE probe. CLINICAL RELEVANCE: Improper use of mRFE may cause thermal injury on articular cartilage.
Authors:
Y Lu; M L Meyer; J J Bogdanske; M D Markel
Related Documents :
3480294 - Validity of a hinged constant force probe and a similar, immobilised probe in untreated...
14506574 - Retrospective review of the internal doppler probe for intra- and postoperative microva...
15152824 - Smoking may affect root coverage outcome: a prospective clinical study in humans.
6575024 - Validity of clinical assessments related to the cemento-enamel junction.
3476584 - Wear and microhardness of glass-ionomer cements.
8833584 - Hardness and young's modulus of human peritubular and intertubular dentine.
Publication Detail:
Type:  Evaluation Studies; Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Veterinary and comparative orthopaedics and traumatology : V.C.O.T     Volume:  20     ISSN:  0932-0814     ISO Abbreviation:  Vet Comp Orthop Traumatol     Publication Date:  2007  
Date Detail:
Created Date:  2007-03-16     Completed Date:  2007-04-17     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  8906319     Medline TA:  Vet Comp Orthop Traumatol     Country:  Germany    
Other Details:
Languages:  eng     Pagination:  34-7     Citation Subset:  IM    
Affiliation:
Comparative Orthopaedic Research Laboratory, Department of Medical Sciences, School of Veterinary Medicine, University of Wisconsin-Madison, Madison, Wisconsin, USA. Luy@svm.vetmed.wisc.edu
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Animals
Arthroscopy / methods
Catheter Ablation / instrumentation,  methods
Cattle / surgery*
Chondrocytes / radiation effects*
Models, Animal
Patella / cytology,  radiation effects*
Radio Waves*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Intradiscal pressure in the degenerated porcine intervertebral disc.
Next Document:  Use of transilial pinning for the treatment of sacroiliac separation in 25 dogs and finite element a...