The effects of electrode misplacement on clinicians' interpretation of the standard 12-lead electrocardiogram.
MedLine Citation:
PMID:  22939805     Owner:  NLM     Status:  In-Data-Review    
Abstract/OtherAbstract:
INTRODUCTION: This study investigates how a particular incorrect electrode configuration affects the 12-lead Electrocardiogram (ECG).
METHODS: A correct and an incorrect 12-lead ECG were extracted from a 192-lead BSPM. This was done for 232 BSPMs yielding 464 12-lead ECGs. The particular incorrect ECG involved displacing electrodes V1 and V2 in the second intercostal space whilst also offsetting the remaining electrodes. These ECGs were examined in two stages: (a) analysis of the effects of electrode misplacement on signal morphology and (b) analysis of how often the incorrect electrode configuration changed the diagnosis of two clinicians in a random sample of 75 patients.
RESULTS: According to the Root Mean Square Error (RMSE) of the difference between PQRST intervals in the correct and incorrect ECGs, lead V2 is the most affected lead (mean: 185μV±82μV), followed by lead V4 (mean: 114μV±59μV) and lead V1 (mean: 100μV±47μV). It was found that if the incorrect electrode configuration is applied, there is a 17% to a 24% chance the diagnostic interpretation will be different. Quantified using Similarity Coefficient (SC) leads V1 and V2 were found to be more alike when misplaced in the second intercostal space. The average SC between these leads when correctly placed was 0.08 (±0.65), however when incorrectly placed, the average SC was 0.43 (±0.3).
CONCLUSION: There is a reasonable chance this particular incorrect electrode configuration will change the diagnosis of the 12-lead ECG. This highlights the importance of developing algorithms to detect electrode misplacement along with better education regarding ECG acquisition.
Authors:
R R Bond; D D Finlay; C D Nugent; C Breen; D Guldenring; M J Daly
Related Documents :
14972635 - The detection of impedance cardiogram characteristic points using wavelet transform.
7334285 - Analysis of pattern recognition by man using detection experiments.
17281825 - A novel design of a portable otoacoustic emissions detecting system based on arm.
9735275 - The coexistence of competing parasites. part ii-hyperparasitism and food chain dynamics.
15305325 - Sampling correction in linkage analysis.
6864315 - Radionuclide kinetics in mird dose calculations.
Publication Detail:
Type:  Journal Article     Date:  2012-04-03
Journal Detail:
Title:  European journal of internal medicine     Volume:  23     ISSN:  1879-0828     ISO Abbreviation:  Eur. J. Intern. Med.     Publication Date:  2012 Oct 
Date Detail:
Created Date:  2012-09-03     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9003220     Medline TA:  Eur J Intern Med     Country:  Netherlands    
Other Details:
Languages:  eng     Pagination:  610-5     Citation Subset:  IM    
Copyright Information:
Copyright © 2012 European Federation of Internal Medicine. Published by Elsevier B.V. All rights reserved.
Affiliation:
Computer Science Research Institute, School of Computing and Mathematics, University of Ulster, Northern Ireland, UK.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Red blood cell distribution width and its relation to cardiac function and biomarkers in a prospecti...
Next Document:  Plasma metalloproteinase-9 and restrictive filling pattern as major predictors of outcome in patient...