The effect of very small air gaps on small field dosimetry.
MedLine Citation:
PMID:  23044638     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
The purpose of this study was to investigate the effect of very small air gaps (less than 1 mm) on the dosimetry of small photon fields used for stereotactic treatments. Measurements were performed with optically stimulated luminescent dosimeters (OSLDs) for 6 MV photons on a Varian 21iX linear accelerator with a Brainlab µMLC attachment for square field sizes down to 6 mm × 6 mm. Monte Carlo simulations were performed using EGSnrc C++ user code cavity. It was found that the Monte Carlo model used in this study accurately simulated the OSLD measurements on the linear accelerator. For the 6 mm field size, the 0.5 mm air gap upstream to the active area of the OSLD caused a 5.3% dose reduction relative to a Monte Carlo simulation with no air gap. A hypothetical 0.2 mm air gap caused a dose reduction >2%, emphasizing the fact that even the tiniest air gaps can cause a large reduction in measured dose. The negligible effect on an 18 mm field size illustrated that the electronic disequilibrium caused by such small air gaps only affects the dosimetry of the very small fields. When performing small field dosimetry, care must be taken to avoid any air gaps, as can be often present when inserting detectors into solid phantoms. It is recommended that very small field dosimetry is performed in liquid water. When using small photon fields, sub-millimetre air gaps can also affect patient dosimetry if they cannot be spatially resolved on a CT scan. However the effect on the patient is debatable as the dose reduction caused by a 1 mm air gap, starting out at 19% in the first 0.1 mm behind the air gap, decreases to <5% after just 2 mm, and electronic equilibrium is fully re-established after just 5 mm.
Authors:
P H Charles; S B Crowe; T Kairn; J Kenny; J Lehmann; J Lye; L Dunn; B Hill; R T Knight; C M Langton; J V Trapp
Related Documents :
22501518 - Vibration energy of metamaterials with negative effective parameters.
22463308 - Optimal phase description of chaotic oscillators.
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2012-10-09
Journal Detail:
Title:  Physics in medicine and biology     Volume:  57     ISSN:  1361-6560     ISO Abbreviation:  Phys Med Biol     Publication Date:  2012 Oct 
Date Detail:
Created Date:  2012-10-9     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  0401220     Medline TA:  Phys Med Biol     Country:  -    
Other Details:
Languages:  ENG     Pagination:  6947-6960     Citation Subset:  -    
Affiliation:
School of Chemistry, Physics and Mechanical Engineering, Queensland University of Technology, GPO Box 2434, Brisbane, Qld 4001, Australia.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Assessment of left ventricular 2D flow pathlines during early diastole using spatial modulation of m...
Next Document:  Inappropriate Prescribing and Related Hospital Admissions in Frail Older Persons According to the ST...