Document Detail


A comparison of three minimally invasive cardiac output devices with thermodilution in elective cardiac surgery.
MedLine Citation:
PMID:  22165352     Owner:  NLM     Status:  In-Process    
Abstract/OtherAbstract:
This study compared the cardiac output responses to haemodynamic interventions as measured by three minimally invasive monitors (Oesophageal Doppler Monitor the VigileoFlotrac and the LiDCOrapid) to the responses measured concurrently using thermodilution, in cardiac surgical patients. The study also assessed the precision and bias of these monitors in relation to thermodilution measurements. After a fluid bolus of at least 250 ml, the measured change in cardiac output was different among the devices, showing an increase with thermodilution in 82% of measurements, Oesophageal Doppler Monitor 68%, VigileoFlotrac 57% and LiDCOrapid 41%. When comparing the test devices to thermodilution, the kappa statistic showed at best only fair agreement, Oesophageal Doppler Monitor 0.34, LiDCOrapid 0.28 and VigileoFlotrac -0.03. After vasopressor administration, there was also significant variation in the change in cardiac output measured by the devices. Using Bland-Altman analysis, the precision of the devices in comparison to thermodilution showed minimal bias, but wide limits of agreement with percentage errors of Oesophageal Doppler Monitor 64.5%, VigileoFlotrac 47.6% and LiDCOrapid 54.2%. These findings indicate that these three devices differ in their responses, do not always provide the same information as thermodilution and should not be used interchangeably to track cardiac output changes.
Authors:
T D Phan; R Kluger; C Wan; D Wong; A Padayachee
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Anaesthesia and intensive care     Volume:  39     ISSN:  0310-057X     ISO Abbreviation:  Anaesth Intensive Care     Publication Date:  2011 Nov 
Date Detail:
Created Date:  2011-12-13     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  0342017     Medline TA:  Anaesth Intensive Care     Country:  Australia    
Other Details:
Languages:  eng     Pagination:  1014-21     Citation Subset:  IM    
Affiliation:
Department of Anaesthesia, St Vincent's Hospital, Fitzroy, Victoria, Australia. tuong.phan@svhm.org.au
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Comments/Corrections
Comment In:
Anaesth Intensive Care. 2011 Nov;39(6):999-1000   [PMID:  22165348 ]

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Laryngeal mask airway for airway control during percutaneous dilatational tracheostomy.
Next Document:  Comparison of fluid compartments and fluid responsiveness in septic and non-septic patients.