Document Detail


A comparison between human magnetostimulation thresholds in whole-body and head/neck gradient coils.
MedLine Citation:
PMID:  11477644     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
Gradient coil magnetostimulation thresholds were measured in a group of 20 volunteers in both a whole-body gradient coil and a head/neck gradient coil. Both coils were operated using both x and y axes simultaneously (xy oblique mode). The waveform applied was a 64-lobe trapezoidal train with 1-ms flat-tops and varying rise times. Thresholds were based on the subjects' perception of stimulation, and painful sensations were not elicited. Thresholds were expressed in terms of the total gradient excursion required to cause stimulation as a function of the duration of the excursion. Thresholds for each subject were fit to a linear model, and values for the threshold curve slope (SR(min)) and vertical axis intercept (DeltaG(min)) were extracted. For the body coil, the mean values were: SR(min) = 62.2 mT/m/ms, DeltaG(min) = 44.4 mT/m. For the head/neck coil, the mean values were: SR(min) = 87.3 mT/m/ms, DeltaG(min) = 78.9 mT/m. These curve parameters were combined with calculated values for the induced electric field as a function of position within the coil to yield the tissue specific parameters E(r) (electric field rheobase) and tau(c) (chronaxie). For tissue stimulated within the body coil, the mean values were: E(r) = 1.8 V/m, tau(c) = 770 micros. For tissue stimulated within the head/neck coil, the mean values were: E(r) = 1.3 V/m, tau(c) = 1100 micros. Scalar potential contributions were not included in the calculation of induced electric fields. The mean threshold curves were combined with the gradient system performance curves to produce operational limit curves. The operational limit curves for the head/neck coil system were verified to be higher than those of the whole-body coil; however, the head/neck system was also found to be physiologically limited over a greater range of its operation than was the body coil. Subject thresholds between the two coils were not well correlated.
Authors:
B A Chronik; B K Rutt
Related Documents :
19528834 - The effect of worksite physical activity intervention on physical capacity, health, and...
22062174 - Effect of pre-slaughter physiological conditions on the oxidative stability of colour a...
2023144 - Hyperventilation provocation in patients with chest pain and a negative treadmill exerc...
16406834 - To treat or not to treat postpartum pelvic girdle pain with stabilizing exercises?
6189164 - Cyclic amp-dependent and -independent effects of beta-adrenergic stimulation on the con...
24744904 - Gdnf content and nmj morphology are altered in recruited muscles following high-speed a...
Publication Detail:
Type:  Comparative Study; Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Magnetic resonance in medicine : official journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine     Volume:  46     ISSN:  0740-3194     ISO Abbreviation:  Magn Reson Med     Publication Date:  2001 Aug 
Date Detail:
Created Date:  2001-07-30     Completed Date:  2001-10-04     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  8505245     Medline TA:  Magn Reson Med     Country:  United States    
Other Details:
Languages:  eng     Pagination:  386-94     Citation Subset:  IM    
Affiliation:
Department of Physics and Astronomy, University of Western Ontario, Ontario, Canada.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Female
Humans
Magnetic Resonance Imaging / adverse effects,  instrumentation*
Male
Pain Threshold / physiology
Sensation / physiology*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Radio frequency magnetic field mapping of a 3 Tesla birdcage coil: experimental and theoretical depe...
Next Document:  Restoration of motion-related signal loss and line-shape deterioration of proton MR spectra using th...