Document Detail


A cadaver model evaluating femoral intramedullary reaming: a comparison between new reamer design (Pressure Sentinel) and a novel suction/irrigation reamer (RIA).
MedLine Citation:
PMID:  21144925     Owner:  NLM     Status:  In-Process    
Abstract/OtherAbstract:
OBJECTIVE: To compare the intramedullary pressures developed during reaming of cadaveric femurs with the Synthes Reamer/Irrigator/Aspirator (RIA) reamer and the Zimmer Pressure Sentinel (PS) reamer, controlling for the force and speed of reaming.
METHOD: Fifteen matched pairs of frozen unpreserved femurs were used in the study. Two pressure transducers and two thermocouples were screwed into holes drilled into the femoral shaft. The femurs were stabilized in the vertical position by mechanically attaching them to a custom jig equipped with a load cell to detect the vertical component of force applied to the reamer. Proximally a linear voltage displacement transducer (LVDT) was attached to the reamer to record the continuous position of the reamer. All femurs were reamed to a diameter 2 mm larger than the narrowest point of the canal. Proximal and distal pressures, proximal and distal temperatures, applied vertical force and displacement were measured continuously throughout the reaming process. Maximum and minimum, and average proximal and distal pressures for each reamer were obtained and compared using paired t-tests. Averages were also calculated and compared in the same manner. A p-value of <0.05 was considered statistically significant.
RESULTS: The maximum, minimum and average intramedullary pressures during reaming were significantly lower with the RIA system than the PS reamer. In general, the pressures produced by the RIA system were consistently below atmospheric pressure for the majority of the reaming time. This was not true for the PS system. No appreciable temperature changes were observed during any of the trials.
CONCLUSION: The RIA reaming system significantly reduces the intramedullary pressures produced during the reaming process compared to the PS reaming system.
Authors:
Gordon Goplen; Janie Astephen Wilson; Michael McAffrey; Kevin Deluzio; Ross Leighton
Related Documents :
6641025 - Hemodynamic changes during hemodialysis: role of dialyzate.
22734575 - Wrist skin temperature, motor activity, and body position as determinants of the circad...
15857155 - Oxygenation inhibits the physiological tissue-protecting mechanism and thereby exacerba...
15650865 - Tissue capnometry: does the answer lie under the tongue?
12022525 - Intracoronary shunts reduce transient intraoperative myocardial dysfunction during off-...
1834845 - Reversal of cardiovascular structural changes during long-term antihypertensive therapy.
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Injury     Volume:  41 Suppl 2     ISSN:  1879-0267     ISO Abbreviation:  Injury     Publication Date:  2010 Nov 
Date Detail:
Created Date:  2010-12-14     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  0226040     Medline TA:  Injury     Country:  Netherlands    
Other Details:
Languages:  eng     Pagination:  S38-42     Citation Subset:  IM    
Copyright Information:
Copyright © 2010 Elsevier Ltd. All rights reserved.
Affiliation:
Faculty of Medicine, Division of Orthopedic Surgery, University of Alberta, Edmonton, Alberta, Canada. goplens@shaw.ca
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Unreamed or RIA reamed nailing: an experimental sheep study using comparative histological assessmen...
Next Document:  The biology of intramedullary reaming.