The application of a modified proportional-derivative control algorithm to arterial pressure alarms in anesthesiology.
MedLine Citation:
PMID:  9641855     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
OBJECTIVE: We have developed an arterial pressure alarm system based on a modified proportional-derivative (PD) controller algorithm, and prospectively tested its ability to predict significant hypotensive episodes, defined as systolic arterial pressure < 80 mmHg, in comparison to conventional limit alarms. METHODS: The alarm algorithm was tuned to detect hypotension using selected invasive arterial pressure traces taken from ten patients who had large intra-operative arterial pressure changes. The algorithm's performance was then tested prospectively in comparison to conventional limit alarms and median filtered limit alarms, set at 85 mmHg and 90 mmHg, for its ability to predict hypotensive episodes in a further 100 patients who required invasive arterial pressure monitoring. RESULTS: For the PD alarm algorithm, onset times for significant hypotensive episodes were between those of limit alarms set at 85 mmHg and 90 mmHg. Offset times were similar to the 85 mmHg limit alarms. The false positive rate was 34% compared with 45-64% for the other alarms (p < 0.01). Using our definitions, there was one false negative in the PD group, being a 15 second drop in observed arterial pressure, when a non invasive blood pressure cuff was inflated above the arterial line. CONCLUSIONS: An arterial pressure alarm system design based on a closed loop control algorithm offered improved perform ance over conventional limit alarms and in addition provided a graded output of severity of the hypotension.
Authors:
J L Derrick; C L Thompson; T G Short
Related Documents :
1565585 - Segmental blood pressure after total hip replacement.
8936235 - An observational study of student nurses' measurement of arterial blood pressure by sph...
2342835 - Finger-pressure waveforms measured on clynes' sentograph distinguish among emotions.
13460725 - Determining blood pressure in infants: use of the flush technique.
24053215 - Prevalence and risk factors for resistant hypertension among hypertensive patients from...
8424305 - Intra-arterial papaverine and leg vascular resistance during in situ bypass surgery wit...
Publication Detail:
Type:  Comparative Study; Journal Article    
Journal Detail:
Title:  Journal of clinical monitoring and computing     Volume:  14     ISSN:  1387-1307     ISO Abbreviation:  J Clin Monit Comput     Publication Date:  1998 Jan 
Date Detail:
Created Date:  1998-07-28     Completed Date:  1998-07-28     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  9806357     Medline TA:  J Clin Monit Comput     Country:  NETHERLANDS    
Other Details:
Languages:  eng     Pagination:  41-7     Citation Subset:  IM    
Affiliation:
Prince of Wales Hospital and The Chinese University of Hong Kong. jamesderrick@pobox.com
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adolescent
Adult
Aged
Aged, 80 and over
Algorithms*
Anesthesiology / instrumentation*
Blood Pressure
Blood Pressure Determination / instrumentation
Blood Pressure Monitors*
Child
Computer Graphics
Equipment Failure*
False Negative Reactions
False Positive Reactions
Female
Forecasting
Fuzzy Logic
Humans
Hypotension / diagnosis
Male
Middle Aged
Monitoring, Intraoperative / instrumentation
Prospective Studies
Systole

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Transcranial Doppler blood flow velocity versus 133Xe clearance cerebral blood flow during mild hypo...
Next Document:  Pulse oximetry monitoring and late postoperative hypoxemia on the general care floor.