| The application of a modified proportional-derivative control algorithm to arterial pressure alarms in anesthesiology. | |
| | |
MedLine Citation:
|
PMID: 9641855 Owner: NLM Status: MEDLINE |
Abstract/OtherAbstract:
|
OBJECTIVE: We have developed an arterial pressure alarm system based on a modified proportional-derivative (PD) controller algorithm, and prospectively tested its ability to predict significant hypotensive episodes, defined as systolic arterial pressure < 80 mmHg, in comparison to conventional limit alarms. METHODS: The alarm algorithm was tuned to detect hypotension using selected invasive arterial pressure traces taken from ten patients who had large intra-operative arterial pressure changes. The algorithm's performance was then tested prospectively in comparison to conventional limit alarms and median filtered limit alarms, set at 85 mmHg and 90 mmHg, for its ability to predict hypotensive episodes in a further 100 patients who required invasive arterial pressure monitoring. RESULTS: For the PD alarm algorithm, onset times for significant hypotensive episodes were between those of limit alarms set at 85 mmHg and 90 mmHg. Offset times were similar to the 85 mmHg limit alarms. The false positive rate was 34% compared with 45-64% for the other alarms (p < 0.01). Using our definitions, there was one false negative in the PD group, being a 15 second drop in observed arterial pressure, when a non invasive blood pressure cuff was inflated above the arterial line. CONCLUSIONS: An arterial pressure alarm system design based on a closed loop control algorithm offered improved perform ance over conventional limit alarms and in addition provided a graded output of severity of the hypotension. |
| | |
Authors:
|
J L Derrick; C L Thompson; T G Short |
Related Documents
:
|
1565585 - Segmental blood pressure after total hip replacement. 8936235 - An observational study of student nurses' measurement of arterial blood pressure by sph... 2342835 - Finger-pressure waveforms measured on clynes' sentograph distinguish among emotions. 13460725 - Determining blood pressure in infants: use of the flush technique. 8424305 - Intra-arterial papaverine and leg vascular resistance during in situ bypass surgery wit... 12102445 - Effects of lung volume and chemoreceptor activity on blood pressure and r-r interval du... |
Publication Detail:
|
Type: Comparative Study; Journal Article |
Journal Detail:
|
Title: Journal of clinical monitoring and computing Volume: 14 ISSN: 1387-1307 ISO Abbreviation: J Clin Monit Comput Publication Date: 1998 Jan |
Date Detail:
|
Created Date: 1998-07-28 Completed Date: 1998-07-28 Revised Date: 2006-11-15 |
Medline Journal Info:
|
Nlm Unique ID: 9806357 Medline TA: J Clin Monit Comput Country: NETHERLANDS |
Other Details:
|
Languages: eng Pagination: 41-7 Citation Subset: IM |
Affiliation:
|
Prince of Wales Hospital and The Chinese University of Hong Kong. jamesderrick@pobox.com |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
Adolescent Adult Aged Aged, 80 and over Algorithms* Anesthesiology / instrumentation* Blood Pressure Blood Pressure Determination / instrumentation Blood Pressure Monitors* Child Computer Graphics Equipment Failure* False Negative Reactions False Positive Reactions Female Forecasting Fuzzy Logic Humans Hypotension / diagnosis Male Middle Aged Monitoring, Intraoperative / instrumentation Prospective Studies Systole |