Document Detail


The accuracy of non-invasively continuous total hemoglobin measurement by pulse CO-Oximetry undergoing acute normovolemic hemodilution and reinfusion of autologous blood.
MedLine Citation:
PMID:  24972855     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
BACKGROUND: Non-invasively continuous total hemoglobin (SpHb) measurement has not been assessed adequately in acute bleeding and rapid blood transfusion during surgery. Thus, we have assessed the efficacy of SpHb during both acute normovolemic hemodilution (ANH) and autologous blood transfusion (ABT).
METHODS: Twenty-four patients undergoing urological and gynecological surgery were enrolled. ANH was induced by withdrawing blood of 800 g with simultaneous fluid administration. When surgical hemostasis was completed, collected blood was reinfused. Measurement of SpHb, perfusion index (PI) and real total Hb (tHb) were done before and after each 400 ml blood removal (-0, -400, -800 ml) and reinfusion (+0, +400, +800 ml).
RESULTS: A Bland-Altman analysis for repeated measurements showed a bias (precision) g/dl of 1.12 (1.25), 1.43 (1.24) and 1.10 (1.23) for all data, during ANH and during ABT, respectively. Additionally, a bias (precision) increased with a reduction in tHb (g/dl): ≥10.0; 0.74 (1.30), 8.0-10.0; 1.15 (1.12) and <8.0; 1.60 (1.28). Although the difference between SpHb and tHb was almost zero before anesthesia induction, it became significant just before ANH and did not change further by ANH and ABT. Significant correlations between SpHb and tHb for all data (r = 0.75, n = 228, p < 0.001) were observed. PI slightly correlated with the difference between SpHb and tHb (r = 0.38, n = 216, p < 0.001). Furthermore, before and after induction of anesthesia, PI also correlated with the difference between SpHb and tHb (r = 0.42, n = 23, p = 0.048 and r = 0.51, n = 22, p = 0.016, respectively).
CONCLUSIONS: The present data suggest that SpHb may overestimate tHb during ANH and ABT. In addition, PI and tHb levels had an impact on the accuracy of SpHb measurements.
Authors:
Junichi Saito; Masato Kitayama; Masafumi Oishi; Tomoyuki Kudo; Masahiro Sawada; Hiroshi Hashimoto; Kazuyoshi Hirota
Related Documents :
14718645 - Incorporation of the genetic control of alcohol dehydrogenase into a physiologically ba...
11354735 - Co and no2 pollution in a long two-way traffic road tunnel: investigation of no2/nox ra...
18049455 - Coding and non-coding polymorphisms in alcohol dehydrogenase alters protein expression ...
16036765 - Insights into the origin of postmortem ethanol.
168145 - On the distribution of some genetic markers in libya.
1484865 - Garlic in the ovine fetal environment.
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2014-6-28
Journal Detail:
Title:  Journal of anesthesia     Volume:  -     ISSN:  1438-8359     ISO Abbreviation:  J Anesth     Publication Date:  2014 Jun 
Date Detail:
Created Date:  2014-6-28     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  8905667     Medline TA:  J Anesth     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Comparative in vitro activities of dalbavancin and seven comparator agents against 41 Staphylococcus...
Next Document:  Age-related clinicopathologic and molecular features of patients receiving curative hepatectomy for ...