The accuracy of home glucose meters in hypoglycemia.
MedLine Citation:
PMID:  20615103     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
BACKGROUND: Home glucose meters (HGMs) may not be accurate enough to sense hypoglycemia. We evaluated the accuracy and the capillary and venous comparability of five different HGMs (Optium Xceed [Abbott Diabetes Care, Alameda, CA, USA], Contour TS [Bayer Diabetes Care, Basel, Switzerland], Accu-Chek Go [Roche Ltd., Basel, Switzerland], OneTouch Select [Lifescan, Milpitas, CA, USA], and EZ Smart [Tyson Bioresearch Inc., Chu-Nan, Taiwan]) in an adult population.
METHODS: The insulin hypoglycemia test was performed to 59 subjects (56 males; 23.6 +/- 3.2 years old). Glucose was measured from forearm venous blood and finger capillary samples both before and after regular insulin (0.1 U/kg) was injected. Venous samples were analyzed in the reference laboratory by the hexokinase method. In vitro tests for method comparison and precision analyses were also performed by spiking the glucose-depleted venous blood.
RESULTS: All HGMs failed to sense hypoglycemia to some extend. EZ Smart was significantly inferior in critical error Zone D, and OneTouch Select was significantly inferior in the clinically unimportant error Zone B. Accu-Chek Go, Optium Xceed, and Contour TS had similar performances and were significantly better than the other two HGMs according to error grid analysis or International Organization for Standardization criteria. The in vitro tests were consistent with the above clinical data. The capillary and venous consistencies of Accu-Chek Go and OneTouch Select were better than the other HGMs.
CONCLUSIONS: The present results show that not all the HGMs are accurate enough in low blood glucose levels. The patients and the caregivers should be aware of these restrictions of the HGMs and give more credit to the symptoms of hypoglycemia than the values obtained by the HGMs. Finally, these results indicate that there is a need for the revision of the accuracy standards of HGMs in low blood glucose levels.
Authors:
Alper Sonmez; Zeynep Yilmaz; Gokhan Uckaya; Selim Kilic; Serkan Tapan; Abdullah Taslipinar; Aydogan Aydogdu; Mahmut Yazici; Mahmut Ilker Yilmaz; Muhittin Serdar; M Kemal Erbil; Mustafa Kutlu
Related Documents :
16324923 - Effects of glimepiride and glyburide on glucose counterregulation and recovery from hyp...
2645123 - Syndromes of autoimmunity and hypoglycemia. autoantibodies directed against insulin and...
20152743 - Medical management of diabetes after bariatric surgery.
Publication Detail:
Type:  Evaluation Studies; Journal Article    
Journal Detail:
Title:  Diabetes technology & therapeutics     Volume:  12     ISSN:  1557-8593     ISO Abbreviation:  Diabetes Technol. Ther.     Publication Date:  2010 Aug 
Date Detail:
Created Date:  2010-07-09     Completed Date:  2010-10-08     Revised Date:  2011-03-02    
Medline Journal Info:
Nlm Unique ID:  100889084     Medline TA:  Diabetes Technol Ther     Country:  United States    
Other Details:
Languages:  eng     Pagination:  619-26     Citation Subset:  IM    
Affiliation:
Department of Endocrinology and Metabolism, Gulhane School of Medicine, Ankara, Turkey. alpersonmez@yahoo.com
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Blood Glucose / analysis*
Blood Glucose Self-Monitoring / instrumentation*,  standards*
Female
Humans
Hypoglycemia / blood*
Male
Prospective Studies
Chemical
Reg. No./Substance:
0/Blood Glucose
Comments/Corrections
Comment In:
Diabetes Technol Ther. 2011 Feb;13(2):171   [PMID:  21284485 ]

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Effectiveness of exenatide in asian indians in a clinical care setting.
Next Document:  Insulin pump therapy in patients with type 2 diabetes safely improved glycemic control using a simpl...