Work of breathing associated with pressure support ventilation in two different ventilators.
MedLine Citation:
PMID:  11416878     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
The purpose of this study was to compare the work of breathing during pressure support ventilation (PSV) with positive end expiratory pressure (PEEP) utilizing the Siemens SV300 (SV300) and Dräger Evita 4 (EV4) ventilators. Our hypothesis was that patients' work of breathing (WOB(P)) would be unchanged in PSV utilizing flow triggering (FT) in both the SV300 and EV4. We compared two ventilators using six healthy, intubated, sedated, spontaneously breathing pigs weighing approximately 10 kg each. WOB(P) (j/L) and ventilator work of breathing (WOB(V)) (j/L) were measured using a portable monitor which utilizes an esophageal balloon and flow transducer. Each breath was further analyzed for duration of inspiratory effort and negative deflection of pressure needed to trigger PSV. Animals were studied with the SV300 and EV4 on a pressure support of 5 cmH(2)O and PEEP settings of 0 and 5 cmH(2)O. Data were analyzed using the Wilcoxon signed rank test with significance set at P <or= 0.05 WOB(P) was 90% (PS 5, PEEP 0) and 52% (PS 5, PEEP 5) lower on the SV300 compared to the EV4. WOB(V) was 94% (PS 5, PEEP 0) and 39% (PS 5, PEEP 5) higher on the SV300 when compared to the EV4. The change in airway pressure (delta p) from baseline and most negative deflection of pressure were greater with the EV4 as compared to the SV300, although delta pressure was not found to be statistically significant on PS of 5 cmH(2)O and PEEP of 5 cmH(2)O. The SV300 also had shorter duration of inspiratory effort from initiation of breath to most negative deflection of pressure and to maximum flow than the EV4. In conclusion, these results suggest there are significant differences in WOB(P) between the SV300 and EV4 ventilators. Response time of the ventilators may explain the differences in duration of inspiratory effort and the patient's work of breathing and thus may have an impact on weaning time for ventilated patients.
Authors:
R C Sanders; T L Thurman; S J Holt; K Taft; M J Heulitt
Related Documents :
20185658 - The ability of pleth variability index to predict the hemodynamic effects of positive e...
9468178 - Acute lung injury and the acute respiratory distress syndrome.
2161598 - Role of the renin-angiotensin system in liver blood flow reduction produced by positive...
3918778 - Relaxation of tracheomotor tone by peep: influence of hypocapnia and acidemia.
12698068 - What is the most important component of blood pressure: systolic, diastolic or pulse pr...
22254488 - Modeling the adaptive pathophysiology of essential hypertension.
Publication Detail:
Type:  Comparative Study; Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Pediatric pulmonology     Volume:  32     ISSN:  8755-6863     ISO Abbreviation:  Pediatr. Pulmonol.     Publication Date:  2001 Jul 
Date Detail:
Created Date:  2001-06-20     Completed Date:  2001-08-30     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  8510590     Medline TA:  Pediatr Pulmonol     Country:  United States    
Other Details:
Languages:  eng     Pagination:  62-70     Citation Subset:  IM    
Copyright Information:
Copyright 2001 Wiley-Liss, Inc.
Affiliation:
Critical Care Medicine, Department of Pediatrics, University of Arkansas for Medical Sciences and Respiratory Care Services, Arkansas Children's Hospital, Little Rock, Arkansas 72202, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Animals
Equipment Design
Models, Animal
Positive-Pressure Respiration / instrumentation*
Pulmonary Ventilation / physiology
Swine
Ventilators, Mechanical*
Work of Breathing / physiology*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Spirometry in children aged 3 to 5 years: reliability of forced expiratory maneuvers.
Next Document:  Patient-ventilator interactions in new modes of patient-triggered ventilation.