Within- and between-laboratory precision in the measurement of body volume using air displacement plethysmography and its effect on body composition assessment.
MedLine Citation:
PMID:  14710169     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
OBJECTIVE: To determine and compare the extent of within- and between-laboratory precision in body volume (BV) measurements using air displacement plethysmography (ADP), the BOD POD body composition system, and to interpret any such variability in terms of body composition estimates.
DESIGN: Repeated test procedures of BV assessment using the BOD POD ADP were reproduced at two laboratories for the estimation of precision, both within and between laboratories.
SUBJECTS: In total, 30 healthy adult volunteers, 14 men (age, 19-48 y; body mass index (BMI), 19.7-30.3 kg/m2) and 16 women (age, 19-40 y; BMI, 16.3-35.7 kg/m2), were each subjected to two test procedures at both laboratories. Two additional volunteers were independently subjected to 10 repeated test procedures at both laboratories.
MEASUREMENTS: Repeated measurements of BV, uncorrected for the effects of isothermal air in the lungs and the surface area artifact, were obtained using the BOD POD ADP, with the identical protocol being faithfully applied at both laboratories. Uncorrected BV measurements were adjusted to give estimates of actual BV that were used to calculate body density (body weight (BWt)/actual BV) from which estimates of body composition were derived. The differences between repeated BV measurements or body composition estimates were used to assess within-laboratory precision (repeatability), as standard deviation (SD) and coefficient of variation; the differences between measurements reproduced at each laboratory were used to determine between-laboratory precision (reproducibility), as bias and 95% limits of agreement (from SD of the differences between laboratories).
RESULTS: The extent of within-laboratory methodological precision for BV (uncorrected and actual) was variable according to subject, sample group and laboratory conditions (range of SD, 0.04-0.13 l), and was mostly due to within-individual biological variability (typically 78-99%) rather than to technical imprecision. There was a significant (P<0.05) bias between laboratories for the 10 repeats on the two independent subjects (up to 0.29 l). Although no significant bias (P=0.077) was evident for the sample group of 30 volunteers (-0.05 l), the 95% limits of agreement were considerable (-0.68 to 0.58 l). The effects of this variability in BV on body composition were relatively greater: for example, within-laboratory precision (SD) for body fat as % BWt was between 0.56 and 1.34% depending on the subject and laboratory; the bias (-0.59%) was not significant between laboratories, but there were large 95% limits of agreement (-3.67 to 2.50%).
CONCLUSION: Within-laboratory precision for each BOD POD instrument was reasonably good, but was variable according to the prevailing conditions. Although the bias between the two instruments was not significant for the BV measurements, implying that they can be used interchangeably for groups of similar subjects, the relatively large 95% limits of agreement indicate that greater consideration may be needed for assessing individuals with different ADP instruments. Therefore, use of a single ADP instrument is apparently preferable when assessing individuals on a longitudinal basis.
Authors:
A L Collins; S Saunders; H D McCarthy; J E Williams; N J Fuller
Related Documents :
24705629 - The sensitivity and specificity of using a computer aided diagnosis program for automat...
1580749 - Analytic sensitivity and specificity of enzyme immunoassay results in testing for human...
21649429 - Peer reviewed: in situ bioremediation: defining the limits.
23981009 - Patient, physician, and procedural factors influencing the use of defibrillation testin...
23645849 - Does empirical clostridium difficile infection (cdi) therapy result in false-negative c...
23159889 - Clinical manifestations and anti-phospholipid antibodies in 712 patients with systemic ...
15461599 - Allergenicity and cross-reactivity of russian olive pollen (eleagnus angustifolia).
7634179 - Evaluation of a rapid microfiltration test for detection of hiv-1 antibody in serum.
1151599 - Psychopathology scale of the hutt adaptation of the bender-gestalt test: reliability.
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  International journal of obesity and related metabolic disorders : journal of the International Association for the Study of Obesity     Volume:  28     ISSN:  -     ISO Abbreviation:  Int. J. Obes. Relat. Metab. Disord.     Publication Date:  2004 Jan 
Date Detail:
Created Date:  2004-01-07     Completed Date:  2004-03-02     Revised Date:  2014-06-03    
Medline Journal Info:
Nlm Unique ID:  9313169     Medline TA:  Int J Obes Relat Metab Disord     Country:  England    
Other Details:
Languages:  eng     Pagination:  80-90     Citation Subset:  IM    
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Body Composition / physiology*
Body Constitution / physiology*
Female
Humans
Laboratories / standards*
Male
Middle Aged
Plethysmography / standards
Reference Standards
Reproducibility of Results
Sensitivity and Specificity

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  High protein intake sustains weight maintenance after body weight loss in humans.
Next Document:  Influence of gender, age and BMI on lower limb muscular power output in a large population of obese ...