Document Detail


Voltage dip in pacemaker battery supply: a new cause of pacemaker mediated tachycardia.
MedLine Citation:
PMID:  7683809     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
A new cause of pacemaker mediated tachycardia was observed in a patient equipped with a Vitatron Quintech 931 DDD pacemaker. In this type of pacemaker the microprocessor is switched off during the atrial refractory period. Beyond the recommended replacement time, the internal resistance of the battery may increase to such an extent that switching on and off the microprocessor may cause voltage dips. Those voltage dips are erroneously interpreted as P waves by the atrial sensing amplifier, which may cause self-triggering of the pacemaker and initiate a pacemaker mediated tachycardia.
Authors:
R Stroobandt; F Vandenbulcke; H Falleyn; A Sinnaeve
Related Documents :
15121079 - Capture management efficacy in children and young adults with endocardial and unipolar ...
7999589 - Cheveux incoiffables--diagnostic, clinical and hair microscopic findings, and pathogeni...
19081339 - Transvenous demand pacemaker treatment for intermittent complete heart block in a cat.
21611889 - Recombinant human angiotensin-converting enzyme 2 as a new renin-angiotensin system pep...
2419839 - Tachycardia mediated by an av universal (ddd) pacemaker triggered by a ventricular depo...
25219759 - Improving risk stratification in non-st-segment elevation myocardial infarction with co...
132199 - Challenge to conventional treatment for myocardial failure-mechanical assist.
9659449 - The low molecular weight heparin, enoxaparin, limits infarct size at reperfusion in the...
20469439 - Occult cardiac arrhythmias as risk factors for falls in the elderly.
Publication Detail:
Type:  Case Reports; Journal Article    
Journal Detail:
Title:  Pacing and clinical electrophysiology : PACE     Volume:  16     ISSN:  0147-8389     ISO Abbreviation:  Pacing Clin Electrophysiol     Publication Date:  1993 Apr 
Date Detail:
Created Date:  1993-06-09     Completed Date:  1993-06-09     Revised Date:  2005-11-17    
Medline Journal Info:
Nlm Unique ID:  7803944     Medline TA:  Pacing Clin Electrophysiol     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  806-11     Citation Subset:  IM    
Affiliation:
Department of Cardiology, St.-Jozef Hospital, Ostend, Belgium.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Aged
Electric Power Supplies
Electrocardiography
Equipment Failure
Female
Humans
Pacemaker, Artificial / adverse effects*
Tachycardia / diagnosis,  etiology*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine