Document Detail


Voltage dip in pacemaker battery supply: a new cause of pacemaker mediated tachycardia.
MedLine Citation:
PMID:  7683809     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
A new cause of pacemaker mediated tachycardia was observed in a patient equipped with a Vitatron Quintech 931 DDD pacemaker. In this type of pacemaker the microprocessor is switched off during the atrial refractory period. Beyond the recommended replacement time, the internal resistance of the battery may increase to such an extent that switching on and off the microprocessor may cause voltage dips. Those voltage dips are erroneously interpreted as P waves by the atrial sensing amplifier, which may cause self-triggering of the pacemaker and initiate a pacemaker mediated tachycardia.
Authors:
R Stroobandt; F Vandenbulcke; H Falleyn; A Sinnaeve
Related Documents :
8634429 - Morphological and functional changes of coronary vasculature caused by transcellular bi...
3910239 - Pacemaker syndrome: definition and evaluation.
12808269 - Changes in evoked qt intervals according to variations in atrioventricular delay and ca...
1712949 - Fusion or confusion on holter recording.
7260699 - Preservation of fowl semen in liquid nitrogen: application to breeding programmes.
17582769 - Estimating dominant upper-limb segments during daily activity.
6297419 - Perioperative myocardial infarction and changes in left ventricular performance related...
2909779 - Effect of smoking on the activity of ischemic heart disease.
11903359 - Association of a polymorphism of the ecnos gene with myocardial infarction in a subgrou...
Publication Detail:
Type:  Case Reports; Journal Article    
Journal Detail:
Title:  Pacing and clinical electrophysiology : PACE     Volume:  16     ISSN:  0147-8389     ISO Abbreviation:  Pacing Clin Electrophysiol     Publication Date:  1993 Apr 
Date Detail:
Created Date:  1993-06-09     Completed Date:  1993-06-09     Revised Date:  2005-11-17    
Medline Journal Info:
Nlm Unique ID:  7803944     Medline TA:  Pacing Clin Electrophysiol     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  806-11     Citation Subset:  IM    
Affiliation:
Department of Cardiology, St.-Jozef Hospital, Ostend, Belgium.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Aged
Electric Power Supplies
Electrocardiography
Equipment Failure
Female
Humans
Pacemaker, Artificial / adverse effects*
Tachycardia / diagnosis,  etiology*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Adaptive rate pacing controlled by right ventricular preejection interval for severe refractory orth...