| Ventricular cross stimulation using a pacing system analyzer. | |
| | |
MedLine Citation:
|
PMID: 1697964 Owner: NLM Status: MEDLINE |
Abstract/OtherAbstract:
|
Cross stimulation refers to the situation in which a pacing lead positioned in one cardiac chambers results in stimulation of another chamber. We report three cases of ventricular cross stimulation observed at the time of pacing system implantation. The cross stimulation occurred through bipolar atrial and ventricular leads connected in bipolar configuration to a dual chamber pacing systems analyzer. Analysis of events during cross stimulation indicated that the current pathway was occurring within the pacing system analyzer itself, due to low resistance internal connections between the ring electrodes of the atrial and ventricular pacing systems analyzer channels. Cross stimulation within the pacing system analyzer can be eliminated by making the ventricular channel unipolar. Awareness of the phenomenon reported here should assist in the avoidance of the mistaken diagnosis of pacing lead migration and malposition during pacing system implant procedures. |
| | |
Authors:
|
N Goldschlager; R Francoz |
Related Documents
:
|
14613854 - Preservation of no production by statins in the treatment of heart failure. 3769174 - Characteristics of initiation and termination of catecholamine-induced triggered activi... 17445924 - Right ventricular outflow tract pacing: relation of high to low synchronous ventricular... 15683484 - An algorithm for verifying biventricular capture based on evoked-response morphology. 3698244 - Relationship of angiographically defined coronary artery disease to serum lipoproteins ... 7546794 - Right ventricular diastolic collapse without hemodynamic compromise in a patient with l... |
Publication Detail:
|
Type: Case Reports; Journal Article |
Journal Detail:
|
Title: Pacing and clinical electrophysiology : PACE Volume: 13 ISSN: 0147-8389 ISO Abbreviation: Pacing Clin Electrophysiol Publication Date: 1990 Aug |
Date Detail:
|
Created Date: 1990-10-16 Completed Date: 1990-10-16 Revised Date: 2004-11-17 |
Medline Journal Info:
|
Nlm Unique ID: 7803944 Medline TA: Pacing Clin Electrophysiol Country: UNITED STATES |
Other Details:
|
Languages: eng Pagination: 986-92 Citation Subset: IM |
Affiliation:
|
Division of Cardiology, San Francisco General Hospital, CA 94110. |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
Aged Aged, 80 and over Cardiac Pacing, Artificial Electric Conductivity Electrocardiography Electrodes, Implanted Female Heart Atria Heart Ventricles Humans Male Pacemaker, Artificial* |
From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine
Previous Document: Reversible myocardial depression in survivors of cardiac arrest.
Next Document: Ventricular cross stimulation in a dual chamber pacing system: phenomenon analysis.