Document Detail


Valvuloplasty balloon detaching from the stem of a catheter.
MedLine Citation:
PMID:  11428479     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
A 12-year-old boy underwent pulmonary balloon valvotomy for isolated critical pulmonary stenosis. Following valvotomy, blood was found to be drawn into the syringe during deflation of the balloon, so a provisional diagnosis of a burst balloon was made. However, when the balloon catheter was withdrawn, the balloon got detached from the stem of the catheter at the level of the right atrium and was retained over the exchange guidewire. The balloon, when retrieved with a snare, was found to be intact. The balloon may have been partially detached at the junction of the proximal end of the balloon and the catheter; hence, blood was drawn from the catheter during deflation. In our institution balloons are reused following sterilization with ethylene oxide gas. We conclude that any balloon presumed to have burst inside the heart must be removed with great caution. In a third world country like India, where cost is an important factor, balloons can be reused, but with caution, keeping in mind complications such as in this case.
Authors:
D Prabhakar; P Thirumalai; R Alagesan
Related Documents :
17635469 - Infusion of 50% glucose solution before injection of ethanolamine oleate during balloon...
18807219 - Posttraumatic vasospasm detected by continuous brain tissue oxygen monitoring: treatmen...
20719729 - Balloon occlusion of the ascending aorta without hypothermic circulatory arrest in valv...
17125639 - Alpha(2)-blocker helps to avoid systemic to pulmonary shunt in a prostaglandin dependen...
19490359 - Drug delivery at the aortic valve tissues of healthy domestic pigs with a paclitaxel-el...
487419 - Massive hemoptysis secondary to flow-directed thermodilution catheters.
8084039 - Infrainguinal reconstruction with arm vein, lesser saphenous vein, and remnants of grea...
3520649 - Endotracheal tubes in neonates: sonographic positioning.
517769 - Extent of vascularization by the two branches of the renal artery in the brown bear (ur...
Publication Detail:
Type:  Case Reports; Journal Article    
Journal Detail:
Title:  Indian heart journal     Volume:  53     ISSN:  0019-4832     ISO Abbreviation:  Indian Heart J     Publication Date:    2001 Mar-Apr
Date Detail:
Created Date:  2001-06-28     Completed Date:  2001-07-26     Revised Date:  2004-11-17    
Medline Journal Info:
Nlm Unique ID:  0374675     Medline TA:  Indian Heart J     Country:  India    
Other Details:
Languages:  eng     Pagination:  206-7     Citation Subset:  IM    
Affiliation:
Department of Cardiology, Madras Medical College, Chennai. prabhud@vsnl.com
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Balloon Dilatation / adverse effects*,  instrumentation*
Child
Device Removal / methods*
Echocardiography
Equipment Failure
Follow-Up Studies
Humans
Male
Pulmonary Valve Stenosis / therapy*,  ultrasonography
Risk Assessment