Validation and use of the Finometer for blood pressure measurement in normal, hypertensive and pre-eclamptic pregnancy.
MedLine Citation:
PMID:  14597848     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
OBJECTIVE: Although a large variety of automated blood pressure devices are available, only some have been validated for use in clinical practice. The British Hypertension Society (BHS) recommends separate validation of automated devices in special subgroups, e.g. the elderly and pregnant women. The aim of this study was to compare the Finometer (FM) and the earlier validated SpaceLabs 90207 (SL) with standard auscultatory blood pressure measurements in normal, pre-eclamptic and hypertensive pregnancy, following the guidelines of the BHS and the Association for the Advancement of Medical Instrumentation (AAMI). METHODS: The total study group consisted of 123 pregnant women, of whom were 54 normotensive, 31 pre-eclamptic and 38 hypertensive. Automated readings with the FM and SL were compared with auscultatory blood pressure measurements. Bland-Altman plots, BHS grades, mean pressure differences and 95% limits of agreement were used for analysis. RESULTS: Bland-Altman plots showed a wide scatter of the pressure differences between auscultatory and automated measurements. FM achieved BHS grades C/D, C/B, D/D and D/D in the total, normotensive, pre-eclamptic and hypertensive group, respectively. The AAMI criteria were only met for diastolic blood pressure in the normotensive group. For SL almost identical BHS grades and 95% limits of agreement as compared to our earlier study were found. CONCLUSIONS: The accuracy and precision of the Finometer are not sufficient for determination of absolute blood pressure levels in individual pregnant women. Our present findings on the SpaceLabs 90207 reconfirm our earlier results.
Authors:
Ayten Elvan-Taşpinar; Leonore A Uiterkamp; J Marko Sikkema; Michiel L Bots; Hein A Koomans; Hein W Bruinse; Arie Franx
Related Documents :
22457098 - Pressure applied during surgery alters the biomechanical properties of human saphenous ...
10097588 - Effect of age on urinary excretion of n-acetyl-beta-d-glucosaminidase.
11909568 - Changes in hemodynamics and left ventricular structure after menopause.
9099398 - Employment status and high blood pressure in women: variations by time and by sociodemo...
22887048 - End tidal carbon dioxide is as reliable as transcutaneous monitoring in ventilated post...
2525568 - Complete reversibility of physiological coronary vascular abnormalities in hypertrophie...
Publication Detail:
Type:  Clinical Trial; Journal Article; Research Support, Non-U.S. Gov't; Validation Studies    
Journal Detail:
Title:  Journal of hypertension     Volume:  21     ISSN:  0263-6352     ISO Abbreviation:  J. Hypertens.     Publication Date:  2003 Nov 
Date Detail:
Created Date:  2003-11-04     Completed Date:  2004-07-12     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  8306882     Medline TA:  J Hypertens     Country:  England    
Other Details:
Languages:  eng     Pagination:  2053-60     Citation Subset:  IM    
Affiliation:
Department of Perinatology and Gynaecology, University Medical Centre Utrecht, The Netherlands.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Auscultation
Automation
Blood Pressure Determination / instrumentation*
Female
Humans
Hypertension / physiopathology*
Oscillometry / instrumentation
Pre-Eclampsia / physiopathology*
Pregnancy
Pregnancy Complications, Cardiovascular / physiopathology*

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  A comparison of awake-sleep blood pressure variation between normotensive Japanese-American and Cauc...
Next Document:  The G-217A variant of the angiotensinogen gene affects basal transcription and is associated with hy...