Document Detail


Validation of a new transit time ultrasound flowmeter in man.
MedLine Citation:
PMID:  8696905     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
OBJECTIVES: To validate the new CardioMed CM 4000 transit time ultrasound apparatus for intraoperative measurement of volume blood flow in vivo in man. DESIGN: Open, prospective series. MATERIALS: Thirteen patients undergoing in situ saphenous vein grafting for chronic critical leg ischaemia and 12 patients subjected to myocardial revascularisation with the internal thoracic artery as coronary bypass. METHODS: During operations, volume blood flows were measured simultaneously by exsanguination from the cut distal end of the in situ saphenous vein graft or the internal thoracic artery and by the transit time flowmeter equipment. In addition, the feasibility to detect arteriovenous fistula during in situ saphenous vein grafting was examined. RESULTS: Within the examined blood flow range, the volume blood flow determined by the transit time method corresponded to the directly measured blood flow. For in situ saphenous vein grafts: y = -2.4 + 0.95.x (r = 0.99; 35 measurements in 13 patients), and for internal thoracic artery grafts: y = -9.6 + 1.1.x (r = 0.99; 21 measurements in 12 patients), where y is blood flow determined by transit time, and x is directly measured blood flow by exsanguination (r = correlation coefficient) as calculated by the least squares regression method. Fistula detection was easy and swift. CONCLUSIONS: The transit time apparatus was simple to use during intraoperative settings and gave fast, precise measurements of volume blood flow.
Authors:
J Laustsen; E M Pedersen; K Terp; D Steinbrüchel; H H Kure; P K Paulsen; H Jørgensen; W P Paaske
Related Documents :
9436565 - Effect of low-dose positive inotropic drugs on human internal mammary artery flow.
10559005 - Focal expression of angiotensin ii type 1 receptor and smooth muscle cell proliferation...
12493505 - Internal thoracic artery flow competition: studies in a canine h-graft model.
17680515 - Single versus sequential saphenous vein grafting of the circumflex system: a flowmetric...
8354725 - Secnidazole concentrations in plasma and crevicular fluid after a single oral dose.
3921365 - Pcb and pcdf congeners in the blood and tissues of yusho and yu-cheng patients.
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  European journal of vascular and endovascular surgery : the official journal of the European Society for Vascular Surgery     Volume:  12     ISSN:  1078-5884     ISO Abbreviation:  Eur J Vasc Endovasc Surg     Publication Date:  1996 Jul 
Date Detail:
Created Date:  1996-09-03     Completed Date:  1996-09-03     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  9512728     Medline TA:  Eur J Vasc Endovasc Surg     Country:  ENGLAND    
Other Details:
Languages:  eng     Pagination:  91-6     Citation Subset:  IM    
Affiliation:
Department of Cardiothoracic and Vascular Surgery T, Aarhus University Hospital, Skejby Sygehus, Denmark.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Aged
Aged, 80 and over
Arteriovenous Fistula / diagnosis
Blood Flow Velocity
Blood Volume
Coronary Artery Bypass
Equipment Design
Feasibility Studies
Female
Hemorheology / instrumentation*
Humans
Ischemia / surgery
Leg / blood supply
Male
Middle Aged
Monitoring, Intraoperative / instrumentation
Prospective Studies
Reproducibility of Results
Rheology / instrumentation*
Saphenous Vein / transplantation
Thoracic Arteries / transplantation
Ultrasonography / instrumentation*