Document Detail


Validation of continuous thermodilution cardiac output in patients implanted with a left ventricular assist device.
MedLine Citation:
PMID:  12486654     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
OBJECTIVE: To assess the accuracy of a continuous cardiac output (CCO) monitor against an independent, intravascular measurement of flow as can be performed in patients fitted with a left ventricular assist device (LVAD). DESIGN: A prospective cohort study. SETTING: Academic tertiary-care center. PARTICIPANTS: Adult patients (n = 15) presenting for LVAD placement. INTERVENTIONS: Consenting patients presenting for LVAD placement for end-stage cardiac failure were anesthetized, and a CCO pulmonary artery catheter was placed (OptiQ, CCO/SvO(2); Abbott Critical Care, North Chicago, IL). Patients were monitored with transesophageal echocardiography and excluded from analysis if aortic regurgitation was found. Cardiac output was determined using a Q-Vue, CCO/SvO(2) computer with digital readout (Abbott Critical Care, North Chicago, IL). The LVAD was placed in standard fashion during cardiopulmonary bypass. The Thoratec vented electric Heartmate (Thoratec Co, Pleasanton, CA) incorporates an LVAD flow console, which computes LVAD flow within +/- 5% (range, 1.8 to 10 L/min). MEASUREMENTS AND MAIN RESULTS: Cardiac output flow measurements were made from both systems at the following time points: 5 minutes and 30 minutes after protamine administration, at chest closure, and after skin closure. Mean cardiac output for each device did not differ at any time point. Regression analysis (Pearson's) showed acceptable correlation (r(2) = 0.79, p < 0.0001), whereas a bias of 529 mL with limits of agreement of 1,208 mL were shown for CCO measurement compared with LVAD flow. CONCLUSION: The data indicate that the CCO system tends to overestimate cardiac output by approximately 500 mL/min when compared with LVAD flow. Nevertheless, this bias is within the range found by other less-invasive studies done to assess the accuracy of this system and further serves to confirm its relative accuracy.
Authors:
B Mets; R J Frumento; E Bennett-Guerrero; Y Naka
Related Documents :
3901944 - The evolution of the management of penetrating wounds of the heart.
21721194 - Cardiac centre of excellence: duke heart center.
14689414 - A comparison of cardioq and thermodilution cardiac output during off-pump coronary arte...
8335834 - Diagnostic value of chest radiography for pericardial effusion.
19357504 - Changes in plasma c-reactive protein and hemostatic factors prior to and after a first ...
1869744 - Morphologic comparison of frequency and types of acute lesions in the major epicardial ...
Publication Detail:
Type:  Journal Article; Validation Studies    
Journal Detail:
Title:  Journal of cardiothoracic and vascular anesthesia     Volume:  16     ISSN:  1053-0770     ISO Abbreviation:  J. Cardiothorac. Vasc. Anesth.     Publication Date:  2002 Dec 
Date Detail:
Created Date:  2002-12-17     Completed Date:  2003-04-02     Revised Date:  2004-11-17    
Medline Journal Info:
Nlm Unique ID:  9110208     Medline TA:  J Cardiothorac Vasc Anesth     Country:  United States    
Other Details:
Languages:  eng     Pagination:  727-30     Citation Subset:  IM    
Copyright Information:
Copyright 2002, Elsevier Science (USA). All rights reserved.
Affiliation:
Department of Anesthesiology, Columbia University, New York, NY, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Cardiac Output*
Cardiac Output, Low / complications,  surgery
Catheterization, Swan-Ganz
Echocardiography, Transesophageal
Female
Heart-Assist Devices*
Humans
Male
Middle Aged
Monitoring, Intraoperative*
Prospective Studies
Prostheses and Implants
Thermodilution / instrumentation*,  methods