Validation and characterization of transcatheter aortic valve effective orifice area measured by Doppler echocardiography.
MedLine Citation:
PMID:  21999863     Owner:  NLM     Status:  In-Data-Review    
Abstract/OtherAbstract:
OBJECTIVES: The objectives were to compare different Doppler echocardiographic methods for the measurement of prosthetic valve effective orifice area (EOA) following transcatheter aortic valve implantation (TAVI) and to determine the factors influencing the EOA of transcatheter balloon expandable valves.
BACKGROUND: Previous studies have used different methods for the measurement of the valve EOA following TAVI. Factors influencing the EOA of transcatheter valves are unknown.
METHODS: A total of 122 patients underwent TAVI with the use of the Edwards-SAPIEN valve (Edwards Lifesciences, Irvine, California). The EOA was measured by transthoracic echocardiography at hospital discharge, 6 months and 1 year after TAVI with the use of 2 methods as described in previous studies. In Method #1 (EOA(1)), LVOT diameter (LVOTd) entered in the continuity equation was measured at the base of prosthesis cusps whereas, in Method #2 (EOA(2)), LVOTd was measured immediately proximal to the prosthesis stent.
RESULTS: The average EOA(2) (1.57 ± 0.41 cm(2)) was larger (p < 0.01) than the EOA(1) (1.21 ± 0.38 cm(2)). Accordingly, incidence of severe PPM (indexed EOA ≤0.65 cm(2)/m(2)) was 3-fold lower with the use of EOA(2) than with EOA(1) (9% vs. 33%; p < 0.001). Mean transprosthetic gradient correlated better (p = 0.03) with indexed EOA(2) (r = -0.70, p < 0.0001) than with indexed EOA(1) (r = -0.58, p < 0.0001). Intraobserver and interobserver variability were lower for EOA(2) compared to EOA(1) (intra: 5% vs. 7%, p = 0.06; inter: 6% vs. 14%; p < 0.001). Aortic annulus size was the sole independent determinant (p = 0.01) of prosthetic valve EOA(2). The average EOA varied from 1.37 ± 0.23 cm(2) for aortic annulus size <19 mm up to 1.90 ± 0.17 cm(2) for size >23 mm.
CONCLUSIONS: When estimating the EOA of Edwards-SAPIEN valves by Doppler-echocardiography, it is recommended to use the LVOT diameter and velocity measured immediately proximal to the stent. The main determinant of the EOA of transcatheter valves is the patient's annulus size and these valves provide excellent hemodynamics even in patients with a small aortic annulus.
Authors:
Marie-Annick Clavel; Josep Rodés-Cabau; Eric Dumont; Rodrigo Bagur; Sébastien Bergeron; Robert De Larochellière; Daniel Doyle; Eric Larose; Jean G Dumesnil; Philippe Pibarot
Related Documents :
7482673 - Delayed onset of fatal basilar thrombotic embolus after whiplash injury.
2757513 - Iliac arterial transposition.
15791743 - Preoperative localization of the carotid bifurcation for carotid endarterectomy using m...
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  JACC. Cardiovascular imaging     Volume:  4     ISSN:  1876-7591     ISO Abbreviation:  JACC Cardiovasc Imaging     Publication Date:  2011 Oct 
Date Detail:
Created Date:  2011-10-17     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  101467978     Medline TA:  JACC Cardiovasc Imaging     Country:  United States    
Other Details:
Languages:  eng     Pagination:  1053-62     Citation Subset:  IM    
Copyright Information:
Copyright © 2011 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.
Affiliation:
Institut Universitaire de Cardiologie et de Pneumologie de Québec/Québec Heart & Lung Institute, Laval University, Québec, Québec, Canada.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Achieving Optimal Dose Intensity with Adjuvant Chemotherapy in Elderly Breast Cancer Patients: A 10-...
Next Document:  Presence of Extensive LV Remodeling Limits the Benefits of CRT in Patients With Intraventricular Dys...