Validation of the Welch Allyn SureBP (inflation) and StepBP (deflation) algorithms by AAMI standard testing and BHS data analysis.
MedLine Citation:
PMID:  21412074     Owner:  NLM     Status:  In-Data-Review    
Abstract/OtherAbstract:
OBJECTIVES: We evaluated two new Welch Allyn automated blood pressure (BP) algorithms. The first, SureBP, estimates BP during cuff inflation; the second, StepBP, does so during deflation. We followed the American National Standards Institute/Association for the Advancement of Medical Instrumentation SP10:2006 standard for testing and data analysis. The data were also analyzed using the British Hypertension Society analysis strategy.
METHODS: We tested children, adolescents, and adults. The requirements of the American National Standards Institute/Association for the Advancement of Medical Instrumentation SP10:2006 standard were fulfilled with respect to BP levels, arm sizes, and ages. Association for the Advancement of Medical Instrumentation SP10 Method 1 data analysis was used.
RESULTS: The mean±standard deviation for the device readings compared with auscultation by paired, trained, blinded observers in the SureBP mode were -2.14±7.44 mmHg for systolic BP (SBP) and -0.55±5.98 mmHg for diastolic BP (DBP). In the StepBP mode, the differences were -3.61±6.30 mmHg for SBP and -2.03±5.30 mmHg for DBP. Both algorithms achieved an A grade for both SBP and DBP by British Hypertension Society analysis.
CONCLUSION: The SureBP inflation-based algorithm will be available in many new-generation Welch Allyn monitors. Its use will reduce the time it takes to estimate BP in critical patient care circumstances. The device will not need to inflate to excessive suprasystolic BPs to obtain the SBP values. Deflation is rapid once SBP has been determined, thus reducing the total time of cuff inflation and reducing patient discomfort. If the SureBP fails to obtain a BP value, the StepBP algorithm is activated to estimate BP by traditional deflation methodology.
Authors:
Bruce S Alpert
Related Documents :
16799344 - A comparison of effects of alfentanil, fentanyl, and remifentanil on hemodynamic and re...
7941944 - Central-to-peripheral arterial pressure gradient during cardiopulmonary bypass: relatio...
10945384 - Cardiac function after intermittent antegrade warm blood cardioplegia: contribution of ...
8477024 - Potential risks of high-dose epinephrine for resuscitation from ventricular fibrillatio...
23181124 - Impact of renal denervation on renalase expression in adult rats with spontaneous hyper...
15710784 - Aortic stiffness is an independent predictor of progression to hypertension in nonhyper...
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Blood pressure monitoring     Volume:  16     ISSN:  1473-5725     ISO Abbreviation:  Blood Press Monit     Publication Date:  2011 Apr 
Date Detail:
Created Date:  2011-03-17     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9606438     Medline TA:  Blood Press Monit     Country:  England    
Other Details:
Languages:  eng     Pagination:  96-8     Citation Subset:  IM    
Affiliation:
Department of Pediatrics, University of Tennessee Health Science Center, Memphis, Tennessee, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Different roles of S100P Overexpression in Intrahepatic Cholangiocarcinoma: Carcinogenesis of Perihi...
Next Document:  Recent progress in the therapeutic applications of nanotechnology.