OBJECTIVE: To test the accuracy of the TRANSTEK home blood pressure monitor TMB-986 using the International Protocol of the European Society of Hypertension.
METHOD: Device evaluation was performed on 33 participants (15 men and 18 women) with a mean±standard deviation age of 56.2±12.8 years (range: 35-80 years). Blood pressures (systolic blood pressure; SBP and diastolic blood pressure; DBP) were sequentially measured using mercury sphygmomanometer (two trained observers) and alternately measured by the test device (one supervisor).
RESULTS: In phase 1, 35, 44, 45 of SBP measurements and 41, 45, 45 of DBP measurements were within 5, 10, 15 mmHg. In phase 2.1, 70, 94, 99 of SBP measurements and 78, 93, 99 of DBP measurements were within 5, 10, 15 mmHg. In addition, the difference between the device and the mean of two observers was 2.2±5.0 and 1.7±4.8 mmHg for SBP and DBP, respectively. In phase 2.2, for SBP and DBP, respectively, 25 and 26 of 33 participants had at least two of their three differences with 5 mmHg and there were two and one participants who did not have any difference within 5 mmHg.
CONCLUSION: The TRANSTEK blood pressure monitor TMB-986 passed all the phases of the International Protocol both for SBP and DBP and is recommended for home use.
Authors:
Wen Jun Liu; Su Gang Li; Zhe Song; Wei Gong
Related Documents :
8787956 - Description of a new quotient that may differentiate blood pressure profiles in essenti...
19904656 - Co-exposure to noise, n,n-dimethylformamide, and toluene on 24-hour ambulatory blood pr...
14691196 - Association between physical activity and blood pressure is modified by variants in the...
16814126 - How long shall the patient rest before clinic blood pressure measurement?
4817186 - Blood pressure in women taking oral contraceptives.
967886 - Role of the renin-angiotensin system in dogs with perinephritis hypertension.
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't; Validation Studies    
Journal Detail:
Title:  Blood pressure monitoring     Volume:  15     ISSN:  1473-5725     ISO Abbreviation:  Blood Press Monit     Publication Date:  2010 Oct 
Date Detail:
Created Date:  2010-09-17     Completed Date:  2011-01-05     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9606438     Medline TA:  Blood Press Monit     Country:  England    
Other Details:
Languages:  eng     Pagination:  278-80     Citation Subset:  IM    
Affiliation:
Biomedical Engineering Lab, School of Biomedical Engineering, Southern Medical University, Baiyun District, Guang Zhou City, GuangDong, China. sugang.li@gmail.com
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Aged
Aged, 80 and over
Blood Pressure Monitoring, Ambulatory / instrumentation,  methods*
Female
Humans
Male
Middle Aged

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Increased arterial stiffness in children with Williams syndrome and normal blood pressure.
Next Document:  Acute effects of passive smoking on blood pressure and heart rate in healthy females.