| Validation of the Omron HEM-907 device for blood pressure measurement. | |
| | |
MedLine Citation:
|
PMID: 12198340 Owner: NLM Status: MEDLINE |
Abstract/OtherAbstract:
|
BACKGROUND: The aim of this study was to validate the Omron HEM-907 blood pressure (BP) measuring device according to the international validation protocol. DESIGN: The international validation protocol is divided into two phases: the first phase is performed on 15 selected subjects and if the device passes this phase, 18 more subjects are selected making a total number of 33 subjects on which the final validation is performed. METHODS: For each subject, BP measurements were performed simultaneously by two trained observers using mercury sphygmomanometers alternately with the Omron HEM-907 device. In all, 99 measurements were obtained for comparison. The difference between the BP value given by the device and that obtained by the two observers (mean of the two observers) was calculated for each measure. RESULTS: The difference between the two observers was -1 +/- 2 mmHg for the systolic BP (SBP) and for the diastolic BP (DBP). The Omron HEM-907 passed the first phase of the validation process. For the second phase, the average differences between the device and mercury sphygmomanometer readings were -1 +/- 7 and -5 +/- 6 mmHg for SBP and DBP respectively. Readings for the HEM-907 device differed by less than 5 mmHg for 61 of the systolic readings and 52 of the diastolic readings; by less than 10 mmHg for 85 of the systolic readings and 85 of the diastolic readings; and by less than 15 mmHg for 94 of the systolic readings and 96 of the diastolic readings. CONCLUSIONS: The Omron HEM-907 device passes the two phases of the international validation protocol. |
| | |
Authors:
|
Mohamed A El Assaad; Jirar A Topouchian; Bernadette M Darné; Roland G Asmar |
Publication Detail:
|
Type: Journal Article; Research Support, Non-U.S. Gov't; Validation Studies |
Journal Detail:
|
Title: Blood pressure monitoring Volume: 7 ISSN: 1359-5237 ISO Abbreviation: Blood Press Monit Publication Date: 2002 Aug |
Date Detail:
|
Created Date: 2002-08-28 Completed Date: 2003-01-30 Revised Date: 2006-11-15 |
Medline Journal Info:
|
Nlm Unique ID: 9606438 Medline TA: Blood Press Monit Country: England |
Other Details:
|
Languages: eng Pagination: 237-41 Citation Subset: IM |
Copyright Information:
|
Copyright 2002 Lippincott Williams & Wilkins |
Affiliation:
|
L'Institut CardioVasculaire, Paris, France. |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
Adult Aged Equipment Design Female Humans Male Middle Aged Observer Variation Sphygmomanometers* / standards |
From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine
Previous Document: Assessment of repeatability and correlates of arterial compliance.
Next Document: Validation of the A&D UA-631 (UA-779 Life Source) device for self-measurement of blood pressure and ...