Validation of the Nihon Kohden PVM-2701/Impulse-1 automated device by both AAMI (2002) and ISO standards testing.
MedLine Citation:
PMID:  22968109     Owner:  NLM     Status:  In-Data-Review    
Abstract/OtherAbstract:
OBJECTIVES: We evaluated a new algorithm, the Nihon Kohden PVM-2701/Impulse-1, in an automated sphygmomanometer against the requirements of both the Association for the Advancement of Medical Instrumentation (AAMI) SP10 and International Standards Organization (ISO) standards. The Impulse-1 algorithm primarily utilizes a 1 pulse/step deflation. The advantage of the system is to reduce the time of deflation and improve patient comfort.
METHODS: We tested 110 participants (41 children) to fulfill the Standards requirements. We used the more up-to-date statistical analyses in the ISO standard.
RESULTS: The mean±SD for method 1 analyses for systolic blood pressure was 0.3±6.5 mmHg. For diastolic blood pressure the value was -0.3±7.7 mmHg. Method 2 analyses gave values of 0.3±5.2 mmHg for systolic blood pressure and -0.3±6.7 mmHg for diastolic blood pressure. All of the data were within the Standards requirements to pass testing. There was no trend for overestimation or underestimation of BP at either low or high BP levels.
CONCLUSION: The Impulse-1 algorithm passed both the AAMI and ISO standards testing in children and adults. The novelties of the 1 pulse/step deflation algorithm are the significant reduction in deflation time and the manufacturer's claim concerning artifact rejection for both dysrhythmia and motion.
Authors:
Bruce S Alpert
Related Documents :
17600159 - Effect of hypobaric air, oxygen, heliox (50:50), or heliox (80:20) breathing on air bub...
16708919 - Paranasal sinus ventilation by humming.
24269249 - Effectiveness of rosmarinus officinalis essential oil as antihypotensive agent in prima...
16234949 - Gravity-induced reorientation of the interface between two liquids of different densiti...
24032339 - Attentional processes and choking under pressure.
16325189 - Linear voltage profiles and flow homogeneity in pressurized planar electrochromatography.
16163089 - Interface dressings influence the delivery of topical negative-pressure therapy.
6627959 - Clinical use of the arterial/alveolar oxygen tension ratio.
17022669 - Effect of temperature and high pressure on the activity and mode of action of fungal pe...
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Blood pressure monitoring     Volume:  17     ISSN:  1473-5725     ISO Abbreviation:  Blood Press Monit     Publication Date:  2012 Oct 
Date Detail:
Created Date:  2012-09-12     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9606438     Medline TA:  Blood Press Monit     Country:  England    
Other Details:
Languages:  eng     Pagination:  207-9     Citation Subset:  IM    
Affiliation:
Department of Pediatrics, Division of Cardiology, University of Tennessee Health Science Center, Memphis, Tennessee, USA.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Validation of the Rossmax S150 wrist blood pressure monitor for home blood pressure monitoring accor...
Next Document:  Mammary epithelial-restricted expression of activated c-src rescues the block to mammary gland morph...