Document Detail


Validation of the Medipro MediCare 100f upper arm blood pressure monitor, for self-measurement, according to the European Society of Hypertension International Protocol revision 2010.
MedLine Citation:
PMID:  21543970     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
OBJECTIVE: The validation of sphygmomanometer is important in accurate blood pressure measurement. This study presents the validation results by the Medipro MediCare 100f upper arm blood pressure monitor according to the European Society of Hypertension International Protocol (ESH-IP) revision 2010. METHODS: The ESH-IP revision 2010 for the validation of blood pressure measuring devices in adults was followed precisely. A total of 99 couples of test device and reference blood pressure measurements were obtained during the study (three pairs for each of the 33 participants). RESULTS: The device produced 73, 93, and 98 measurements within 5, 10, and 15 mmHg for systolic blood pressure (SBP) and 79, 93, and 96 for diastolic blood pressure (DBP), respectively. The mean standard deviation device-observer difference was 1.4±5.2 mmHg for SBP and 0.02±5.8 mmHg for DBP. The number of participants with two or three of the device-observer differences within 5 mmHg was 24 for SBP and 30 for DBP, whereas there was no participant with none of the device-observer differences within 5 mmHg. CONCLUSION: According to the results of the validation study based on the ESH-IP revision 2010, the Medipro MediCare 100f can be recommended for self-measurement in an adult population.
Authors:
Jun Yi; Yi Wan; Feng Pan; Xiaorong Yu; Huadong Zhao; Fujun Shang; Yongyong Xu
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2011-5-3
Journal Detail:
Title:  Blood pressure monitoring     Volume:  -     ISSN:  1473-5725     ISO Abbreviation:  -     Publication Date:  2011 May 
Date Detail:
Created Date:  2011-5-5     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9606438     Medline TA:  Blood Press Monit     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Affiliation:
aXijing Hospital bDepartment of Health Statistics, Institute for Health Informatics and School of Preventive Medicine cTangdu Hospital, Fourth Military Medical University dThe Fourth Hospital of Xi'an, Xi'an, China.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Beta-lactam antibiotic decreases acquisition of and motivation to respond for cocaine, but not sweet...
Next Document:  Exploring Nursing E-Learning Systems Success Based on Information System Success Model.