Validation of the Kingyield BP101H blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol.
MedLine Citation:
PMID:  20978435     Owner:  NLM     Status:  In-Process    
Abstract/OtherAbstract:
OBJECTIVE: This study aimed to evaluate the accuracy of the automated oscillometric upper arm blood pressure monitor BP101H (Kingyield Technology, Shenzhen, China) for home blood pressure monitoring according to the International Protocol of the European Society of Hypertension.
METHOD: Systolic and diastolic blood pressures were sequentially measured in 33 adult Chinese participants (13 women, 52.6 years of mean age) using a mercury sphygmomanometer (two observers) and the BP101H device (one supervisor). Ninety-nine pairs of comparisons were obtained from 15 participants in phase 1 and further 18 participants in phase 2 of the validation study. Data analysis was done using the ESHIP Analyzer.
RESULTS: The BP101H device successfully passed phase 1 of the validation study with a number of absolute differences between device and observers within 5, 10, and 15 mmHg for at least 34/45, 42/45, and 43/45 measurements, respectively. The device also achieved the targets for phase 2.1, with 79/99, 94/99, and 96/99 differences within 5, 10, and 15 mmHg, respectively, for systolic blood pressure, and with 81/99, 94/99, and 98/99 within 5, 10, and 15 mmHg, respectively, for diastolic blood pressure. In phase 2.2, 27 and 28 participants had at least two of the three device-observers differences within 5 mmHg (required ≥22) for systolic blood pressure and diastolic blood pressure, respectively.
CONCLUSION: The Kingyield upper arm blood pressure monitor BP101H has passed the International Protocol requirements, and hence can be recommended for home use in adults.
Authors:
Wei-Fang Zeng; Qi-Fang Huang; Chang-Sheng Sheng; Yan Li; Ji-Guang Wang
Related Documents :
2312145 - System for mechanical measurements during cardiopulmonary resuscitation in humans.
2656925 - Understanding natural frequency and damping and how they relate to the measurement of b...
438425 - Evaluation of a simple device for regulating intravenous infusions.
19884295 - Beyond office sphygmomanometry: ways to better assess blood pressure.
10320155 - Treatment of intraoperative refractory hypotension with terlipressin in patients chroni...
2884855 - An open one-year comparison of doxazosin and prazosin for mild to moderate essential hy...
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Blood pressure monitoring     Volume:  15     ISSN:  1473-5725     ISO Abbreviation:  Blood Press Monit     Publication Date:  2010 Dec 
Date Detail:
Created Date:  2010-11-18     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9606438     Medline TA:  Blood Press Monit     Country:  England    
Other Details:
Languages:  eng     Pagination:  329-31     Citation Subset:  IM    
Affiliation:
Centre for Epidemiological Studies and Clinical Trials, Shanghai Institute of Hypertension, Ruijin Hospital, Shanghai Jiaotong University School of Medicine, Ruijin 2nd Road 197, Shanghai, China.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Mayne's sign is mainly related to younger age and not to aortic regurgitation.
Next Document:  Effectiveness of an audience response system in teaching pharmacology to baccalaureate nursing stude...