Document Detail


Validation of human whole blood oximetry, using a hyperspectral fundus camera with a model eye.
MedLine Citation:
PMID:  21220553     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
PURPOSE: To assess the accuracy of human blood oximetry measurements in a model eye with a hyperspectral fundus camera.
METHODS: Seven human whole blood samples (two arterial, five venous) were obtained, the oxygen saturations measured with a CO oximeter, and the samples inserted into quartz tubes with internal diameters of 100 and 150 μm. The tubes (n = 20; ten 100 μm and ten 150 μm) were placed within a model eye in front of a background reflectance surface with reflectivities of 20%, 60%, and 99%. Spectral images at wavelengths between 500 and 650 nm were acquired with a hyperspectral fundus camera and analyzed with an oximetric model to calculate the oxygen saturation of blood within the tubes. The calculated oxygen saturations were compared with the measured oxygen saturations. The effects of the background reflectivity and tube size on the accuracy of the calculated oxygen saturations were evaluated.
RESULTS: Background reflectivity and tube size had no significant effect on the mean oxygen saturation difference (P = 0.18 and P = 0.99, respectively; repeated-measures, two-way ANOVA). The mean differences (SD) between the measured and calculated oxygen saturations in segments of the 100 and 150 μm tubes overlying the 20%, 60%, and 99% background reflectivities were (100 μm) -4.0% (13.4%), -6.4% (9.9%), and -5.5% (10.2%) and (150 μm) -5.3% (10.8%), -5.2% (10.7%), and -5.2% (10.9%), respectively.
CONCLUSIONS: There was reasonable agreement between the measured oxygen saturation values and those calculated by the oximetry model. The oximetry model could be used to determine the functional health of the retina.
Authors:
David J Mordant; Ied Al-Abboud; Gonzalo Muyo; Alistair Gorman; Ahmed Sallam; Paul Rodmell; John Crowe; Steve Morgan; Peter Ritchie; Andrew R Harvey; Andrew I McNaught
Related Documents :
21443643 - Wings or winds: inferring bat migration in a stepping-stone archipelago.
21240403 - Design, modeling and characterization of microfluidic architectures for high flow rate,...
2689253 - Prenatal intracranial haemorrhage.
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't; Validation Studies     Date:  2011-04-27
Journal Detail:
Title:  Investigative ophthalmology & visual science     Volume:  52     ISSN:  1552-5783     ISO Abbreviation:  Invest. Ophthalmol. Vis. Sci.     Publication Date:  2011 Apr 
Date Detail:
Created Date:  2011-04-28     Completed Date:  2011-07-08     Revised Date:  2011-08-24    
Medline Journal Info:
Nlm Unique ID:  7703701     Medline TA:  Invest Ophthalmol Vis Sci     Country:  United States    
Other Details:
Languages:  eng     Pagination:  2851-9     Citation Subset:  IM    
Affiliation:
Ophthalmology Department, Gloucestershire Eye Unit, Cheltenham General Hospital, Gloucestershire Hospitals NHS (National Health Service) Foundation Trust, Gloucestershire, United Kingdom.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Humans
Models, Biological*
Oximetry / instrumentation,  methods*
Oxygen / blood*
Photography / instrumentation*
Reproducibility of Results
Retinal Artery / physiology
Retinal Vein / physiology
Grant Support
ID/Acronym/Agency:
//Department of Health
Chemical
Reg. No./Substance:
7782-44-7/Oxygen

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Retinal nerve fiber layer assessment: area versus thickness measurements from elliptical scans cente...