OBJECTIVE: The present study aimed to validate the automated upper arm blood pressure (BP) measuring device BM 44 for home BP monitoring according to the 2002 Protocol of the European Society of Hypertension. The most important new feature of the new device was an integrated 'WHO indicator', which categorizes the patient's individual result within the WHO recommendations for target BP by a coloured scale.
METHODS: Systolic and diastolic BPs were measured sequentially in 35 adult participants (16 men, 19 women) using a standard mercury y-tubed reference sphygmomanometer (two observers) and the BM 44 device (one supervisor). Ninety-nine pairs of comparisons were obtained from 15 participants in phase 1 and a further 18 participants in phase 2 of the validation study.
RESULTS: The BM 44 device passed phase 1 of the validation study successfully with a number of absolute differences between device and observers of 5, 10 and 15 mmHg for at least 28 out of 25, 35 out of 35 and 40 out of 40 measurements, respectively. The device also achieved the targets for phases 2.1 and 2.2, with 23 and 26 participants having had at least two of three device-observers differences within 5 mmHg for systolic and diastolic BP, respectively.
CONCLUSION: The Beurer BM 44 upper arm BP monitor has passed the International Protocol requirements, and hence can be recommended for home use in adults.
Authors:
Stephan Lüders; Ralf Krüger; Claudia Zemmrich; Klaus Forstner; Claus-Dieter Sturm; Peter Bramlage
Related Documents :
24751464 - Social heuristics shape intuitive cooperation.
24899914 - Multiple human pressures and their spatial patterns in european running waters.
24283714 - High-performance nanothermite composites based on aloe vera-directed cuo nanorods.
15045394 - Studies of gas-jet assisted rf-glow discharge atomic absorption spectrometry.
24621044 - Validity and reliability of a pressure ulcer monitoring tool for persons with spinal co...
11085994 - High hydrostatic pressure can probe the effects of functionally related ligands on the ...
Publication Detail:
Type:  Journal Article    
Journal Detail:
Title:  Blood pressure monitoring     Volume:  17     ISSN:  1473-5725     ISO Abbreviation:  Blood Press Monit     Publication Date:  2012 Dec 
Date Detail:
Created Date:  2012-11-13     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9606438     Medline TA:  Blood Press Monit     Country:  England    
Other Details:
Languages:  eng     Pagination:  248-52     Citation Subset:  IM    
Affiliation:
aMedical Clinic, St Josefs Hospital, Cloppenburg bTEMPESTAS - Institute for Medical Research, Cloppenburg cInstitute of Pharmacology and Preventive Medicine GmbH, Potsdam dForschungsinstitut fuer klinische Medizintechnik (Research Institute for Clinical Engineering), Asperg, Germany.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Twenty-four-hour ambulatory blood pressure monitoring in the follow-up of the univentricular heart a...
Next Document:  Validation of the iHealth BP3 upper-arm blood pressure monitor, for clinic use and self-measurement,...