Validation of the ANDON KD-5915 blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol.
MedLine Citation:
PMID:  20658765     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
OBJECTIVE: This study aimed to evaluate the accuracy of the automated oscillometric upper arm blood pressure monitor KD-5915 (ANDON Health, Tianjin, China) for home blood pressure monitoring according to the International Protocol. METHOD: Systolic and diastolic blood pressures were sequentially measured in 33 adult Chinese individuals using a mercury sphygmomanometer (two observers) and the KD-5915 device (one supervisor). Ninety-nine pairs of comparisons were obtained from 15 participants in phase 1 and the remaining 18 participants in phase 2 of the validation study. Data analysis was performed using the ESHIP Analyzer. RESULTS: The KD-5915 device successfully passed phase 1 of the validation study with a number of absolute differences between the device and observers within 5, 10, and 15 mmHg for at least 27 of 45, 34 of 45, and 43 of 45 measurements, respectively. The device also achieved the targets for phase 2.1, with 66 of 99, 81 of 99, and 95 of 99 device-observer differences within 5, 10, and 15 mmHg, respectively, for systolic blood pressure, and with 79 of 99, 95 of 99, and 99 of 99 differences within 5, 10, and 15 mmHg, respectively, for diastolic blood pressure. In phase 2.2, 23, and 27 participants had at least two of the three device-observer differences within 5 mmHg (required Z22) for systolic blood pressure and diastolic blood pressure, respectively. CONCLUSION: The ANDON upper arm blood pressure monitor KD-5915 has passed the International Protocol requirements, and hence can be recommended for use at home in adults.
Authors:
Qi-Fang Huang; Jie Wang; Chang-Sheng Sheng; Na-Na Zhang; Yan Li; Ji-Guang Wang
Related Documents :
20457825 - A small, silent, low friction, linear actuator for mechanical nociceptive testing in ve...
18641425 - Non-invasive measurement of intra-abdominal pressure: a preliminary study.
8051785 - A new device for indirect blood pressure measurement in rabbits.
10234115 - Validation of the omron r3 blood pressure self-measuring device through simultaneous co...
17315605 - The genetic basis for altered blood vessel function in disease: large artery stiffening.
3383355 - Confounding factors in the hemorrhage of conscious swine: a retrospective study of phys...
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't    
Journal Detail:
Title:  Blood pressure monitoring     Volume:  15     ISSN:  1473-5725     ISO Abbreviation:  Blood Press Monit     Publication Date:  2010 Aug 
Date Detail:
Created Date:  2010-07-23     Completed Date:  2010-11-10     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9606438     Medline TA:  Blood Press Monit     Country:  England    
Other Details:
Languages:  eng     Pagination:  232-4     Citation Subset:  IM    
Affiliation:
Centre for Epidemiological Studies and Clinical Trials, Ruijin Hospital, Shanghai Jiaotong University School of Medicine, China.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Aged
Blood Pressure Monitoring, Ambulatory / instrumentation*,  methods
Female
Humans
Hypertension / diagnosis*
Male
Middle Aged
Oscillometry / instrumentation

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Evaluation of the Mobil-O-Graph new generation ABPM device using the ESH criteria.
Next Document:  In Process Citation