Validation of the A&D UA-705 device for self-measurement of blood pressure according to the British Hypertension Society protocol.
MedLine Citation:
PMID:  16810033     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
OBJECTIVES: We tested the accuracy of the UA-705 blood pressure semi-automatic monitor. METHODS: Device evaluation was performed according to the modified British Hypertension Society protocol released in 1993. Eighty-five patients with characteristics outlined in the British Hypertension Society protocol were recruited among those attending our out-patient clinic. Sequential readings were taken for the main validation test. Outcome was classified according to the British Hypertension Society criteria, which are based on four zones of accuracy differing from the mercury standard by 5, 10 and 15 mmHg, or more. RESULTS: The mean blood pressure difference (+ or -1 SD) between device and observers was 0.4 mmHg (SD 7.7) for systolic blood pressure and 0.3 mmHg (SD 8.2) for diastolic blood pressure. Overall, 96% (observer 1) and 95% (observer 2) of readings between device and observers differed by 15 mmHg or less for both systolic and diastolic blood pressure. The device achieved a grade A for both systolic and diastolic blood pressure. CONCLUSIONS: This study shows that the A&D UA-705 device satisfies the British Hypertension Society standard for accuracy by achieving a grade A for both systolic and diastolic blood pressure.
Authors:
Paolo Verdecchia; Fabio Angeli; Roberto Gattobigio; Enrica Angeli; Silvia Pede; Loretta Pittavini
Related Documents :
15636523 - Polydimethylsiloxane connection for quartz microchips in a high-pressure system.
8912223 - Artifact of blood pressure recording using heparin-filled catheter: effects on blood pr...
8484623 - Effect of electrical stimulation of the hypoglossal nerve on airflow mechanics in the i...
Publication Detail:
Type:  Journal Article; Research Support, Non-U.S. Gov't; Validation Studies    
Journal Detail:
Title:  Blood pressure monitoring     Volume:  11     ISSN:  1359-5237     ISO Abbreviation:  Blood Press Monit     Publication Date:  2006 Aug 
Date Detail:
Created Date:  2006-06-30     Completed Date:  2006-09-15     Revised Date:  2006-11-15    
Medline Journal Info:
Nlm Unique ID:  9606438     Medline TA:  Blood Press Monit     Country:  England    
Other Details:
Languages:  eng     Pagination:  223-7     Citation Subset:  IM    
Affiliation:
Department of Cardiology, Hospital R. Silvestrini, Perugia, Italy. verdec@tin.it
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Automation
Blood Pressure Monitoring, Ambulatory / standards*
Diastole
Female
Humans
Male
Middle Aged
Sphygmomanometers
Systole

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Clinical use of indices determined non-invasively from the radial and carotid pressure waveforms.
Next Document:  Validation of the Omron 705 IT oscillometric device for home blood pressure measurement in children ...