Validation of the A&D UA-1020 upper-arm blood pressure monitor for home blood pressure monitoring according to the British Hypertension Society Protocol.
MedLine Citation:
PMID:  23571228     Owner:  NLM     Status:  Publisher    
Abstract/OtherAbstract:
OBJECTIVE: The present study aimed to evaluate the accuracy of the automated oscillometric upper-arm blood pressure (BP) monitor A&D UA-1020 with two different-shaped cuffs for home BP monitoring according to the British Hypertension Society (BHS) Protocol. METHODS: We recruited individuals for each of the two cuffs (D-ring and cylindrical) until there were 85 eligible participants (255 pairs of comparisons) and their BP could meet the BP distribution requirements specified by the BHS Protocol. For each participant, we sequentially measured the systolic and diastolic BP using a mercury sphygmomanometer (two observers) and the UA-1020 device (one supervisor). RESULTS: For the D-ring cuff, the device achieved grade A. The percentage of BP differences within 5, 10, and 15 mmHg was 67, 87, and 96%, respectively, for systolic BP, and 70, 90, and 99%, respectively, for diastolic BP. The average (±SD) of the device-observer differences was -0.2±7.3 mmHg (P=0.64) and 1.7±5.8 mmHg (P<0.0001) for systolic and diastolic BP, respectively. For the cylindrical cuff, the device also achieved grade A. The percentage of BP differences within 5, 10, and 15 mmHg was 67, 88, and 97%, respectively, for systolic BP and 64, 89, and 98%, respectively, for diastolic BP. The average of the device-observer differences was -0.1±7.0 mmHg (P=0.89) and 2.0±6.3 mmHg (P<0.0001) for systolic and diastolic BP, respectively. CONCLUSION: The UA-1020 device has passed the requirements of the BHS Protocol with both the D-ring and the cylindrical cuffs, and hence can be recommended for home use in adults.
Authors:
Wei-Fang Zeng; Yuan-Yuan Kang; Ming Liu; Yan Li; Ji-Guang Wang
Related Documents :
2137728 - Postischemic recovery of mechanical performance and energy metabolism in the presence o...
22875028 - Preventing and managing heel pressure ulceration: an overview.
2402298 - Beneficial effect of magnesium on the isolated perfused rat heart during reperfusion af...
11903068 - Effects of nicorandil on myocardial function and metabolism in the post-ischaemic reper...
9472998 - Paroxysmal sympathetic storms ("diencephalic seizures") after severe diffuse axonal hea...
10096978 - Ischemic preconditioning does not acutely improve load-insensitive parameters of contra...
Publication Detail:
Type:  JOURNAL ARTICLE     Date:  2013-4-6
Journal Detail:
Title:  Blood pressure monitoring     Volume:  -     ISSN:  1473-5725     ISO Abbreviation:  Blood Press Monit     Publication Date:  2013 Apr 
Date Detail:
Created Date:  2013-4-10     Completed Date:  -     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9606438     Medline TA:  Blood Press Monit     Country:  -    
Other Details:
Languages:  ENG     Pagination:  -     Citation Subset:  -    
Affiliation:
Centre for Epidemiological Studies and Clinical Trials, The Shanghai Institute of Hypertension, Ruijin Hospital, Shanghai Jiaotong University School of Medicine, Shanghai, China.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  A review of zinc oxide mineral beneficiation using flotation method.
Next Document:  Oscillometric estimation of aortic pulse wave velocity: comparison with intra-aortic catheter measur...