Use of conventional ECG electrodes for depth of anaesthesia monitoring using the cerebral state index: a clinical study in day surgery.
MedLine Citation:
PMID:  17371776     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
BACKGROUND: The cost-benefit relationship for depth of anaesthesia monitors is complicated by the high cost of specially designed EEG electrodes. The cerebral state index (CSI) monitor will accept regular ECG electrodes with snap connectors. The purpose of this study was to determine if generic ECG electrodes could replace the more expensive proprietary EEG electrodes for the CSI monitor. METHODS: Two identical cerebral state monitors were used simultaneously during sevoflurane anaesthesia for knee arthroscopy in 14 ASA I-II patients. One monitor used proprietary (Danmeter) EEG electrodes and the other used ECG electrodes (3M Red Dot Diagnostic ECG Electrodes). Paired CSI values were recorded every other minute. Anaesthetic depth was titrated clinically. Sedation depth was scored according to the Observer's Assessment of Alertness/Sedation (OAAS) scale. RESULTS: The agreement between the two measures was found to be high, mean difference--0.23, and the overall repeatability mean bias was 6.6 and 153/163 pairs (94%) were located within the 95% limits of agreement. No major difference was noted in impedance, noise, or artifacts. A large overlap in CSI was noted for each level of the OAAS scale; patients with CSI values as low as 40-50 responded whereas patients not responding to surgical stimulation had CSI values as high as 75. The direct cost of disposables decreased from 4euro to 0.50euro per patient by using ordinary ECG electrodes. CONCLUSIONS: Switching from proprietary EEG electrodes to ordinary generic ECG electrodes maintains the same accuracy at about a 10th of the cost when measuring CSI during day surgery with sevoflurane anaesthesia.
Authors:
R E Anderson; U Sartipy; J G Jakobsson
Related Documents :
10212716 - Sub-tenon's block: the effect of hyaluronidase on speed of onset and block quality.
8672376 - Effect of suxamethonium on the auditory evoked response in humans.
23133756 - Blood flow restricted exercise and vascular function.
23113926 - Investigation of a prolonged aptt. different approaches taken by laboratories to achiev...
11323536 - Effect of ketoprofen on muscle function and semg activity after eccentric exercise.
11300546 - Heat in the treatment of patients with anorexia nervosa.
Publication Detail:
Type:  Evaluation Studies; Journal Article     Date:  2007-03-19
Journal Detail:
Title:  British journal of anaesthesia     Volume:  98     ISSN:  0007-0912     ISO Abbreviation:  Br J Anaesth     Publication Date:  2007 May 
Date Detail:
Created Date:  2007-04-25     Completed Date:  2007-06-18     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  0372541     Medline TA:  Br J Anaesth     Country:  England    
Other Details:
Languages:  eng     Pagination:  645-8     Citation Subset:  IM    
Affiliation:
Department of Molecular Medicine and Surgery, Karolinska Institute, Stockholm, Sweden.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Adult
Ambulatory Surgical Procedures
Anesthesia, Inhalation / methods*
Anesthetics, Inhalation
Arthroscopy
Costs and Cost Analysis
Electrocardiography / instrumentation*
Electrodes* / economics
Electroencephalography / economics,  instrumentation*
Female
Humans
Knee Joint / surgery
Male
Methyl Ethers
Middle Aged
Monitoring, Intraoperative / economics,  instrumentation*,  methods
Reproducibility of Results
Chemical
Reg. No./Substance:
0/Anesthetics, Inhalation; 0/Methyl Ethers; 28523-86-6/sevoflurane

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Effects of sevoflurane and propofol on ischaemia-reperfusion injury after thoracic-aortic occlusion ...
Next Document:  Laparoscopic cholecystectomy under segmental thoracic spinal anaesthesia: a feasibility study.