Unusual Wenckebach upper rate response of an atrial-based DDD pacemaker.
MedLine Citation:
PMID:  1378607     Owner:  NLM     Status:  MEDLINE    
We describe in this report an unusual form of Wenckebach upper rate response produced by a DDD pulse generator with atrial-based lower rate timing. The pacemaker maintained the programmed upper and lower rate intervals at the expense of a prolonged atrial paced-ventricular paced AV interval. This form of upper rate behavior eliminated the longer cycle (containing the unsensed P wave) that occurs at the end of the pacemaker Wenckebach sequence during traditional DDD pacing with ventricular-based lower rate timing.
C S Fredman; P Bjerregaard; D L Janosik; S S Barold
Related Documents :
18617317 - Hyperkalemia induced failure of pacemaker capture and sensing.
1704517 - The effects of percutaneous catheter ablation on preexisting permanent pacemakers.
11104407 - Indications for cardiac pacemaker implantation in myotonic dystrophy.
12042627 - Single coronary artery and sick sinus syndrome.
18387637 - Chemical preconditioning effect of 3-nitropropionic acid in anesthetized rat heart.
10155087 - Complications of cardiac catheterization, coronary angiography, and coronary interventi...
Publication Detail:
Type:  Case Reports; Journal Article    
Journal Detail:
Title:  Pacing and clinical electrophysiology : PACE     Volume:  15     ISSN:  0147-8389     ISO Abbreviation:  Pacing Clin Electrophysiol     Publication Date:  1992 Jul 
Date Detail:
Created Date:  1992-08-14     Completed Date:  1992-08-14     Revised Date:  2004-11-17    
Medline Journal Info:
Nlm Unique ID:  7803944     Medline TA:  Pacing Clin Electrophysiol     Country:  UNITED STATES    
Other Details:
Languages:  eng     Pagination:  975-8     Citation Subset:  IM    
St. Louis University Medical Center, Division of Cardiology, MO 63110-0250.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Cardiac Pacing, Artificial / methods*
Equipment Design
Heart Atria
Heart Block / complications,  therapy*
Lupus Erythematosus, Systemic / complications
Pacemaker, Artificial*
Tachycardia, Sinus / complications,  therapy

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine

Previous Document:  Short-term reproducibility over time of right ventricular pulse pressure as a potential hemodynamic ...
Next Document:  Thermal ablation of perfused porcine left ventricle in vitro with the neodymium-YAG laser hot tip ca...