| Unsteady fluid dynamics of several mechanical prosthetic heart valves using a two component laser Doppler anemometer system. | |
| | |
MedLine Citation:
|
PMID: 9335370 Owner: NLM Status: MEDLINE |
Abstract/OtherAbstract:
|
Five typical mechanical heart valves (Starr-Edwards, Björk-Shiley convexo-concave (c-c), Björk-Shiley monostrut, Bicer-Val, and St. Jude Medical) were tested in the mitral position under the pulsatile flow condition. The test program included measurements of velocity and turbulent stresses at 5 downstream locations. The study was carried out using a sophisticated cardiac simulator in conjunction with a highly sensitive 2 component laser Doppler anemometer (LDA) system. The continuous monitoring of parametric time histories revealed useful details about the complex flow and helped to establish the locations and times of the peak parameter values. Based upon the nondimensional presentation of data, the following general conclusions can be made. First, all the 5 valve designs created elevated turbulent stresses during the accelerating and peak flow phases, presenting the possibility of thromboembolism and perhaps hemolysis. Second, the difference in valve configuration seemed to affect the flow characteristics; third, the bileaflet design of the St. Jude valve appeared to create a lower turbulence stress level. |
| | |
Authors:
|
T Akutsu; V J Modi |
Publication Detail:
|
Type: Comparative Study; Journal Article; Research Support, Non-U.S. Gov't |
Journal Detail:
|
Title: Artificial organs Volume: 21 ISSN: 0160-564X ISO Abbreviation: Artif Organs Publication Date: 1997 Oct |
Date Detail:
|
Created Date: 1998-02-02 Completed Date: 1998-02-02 Revised Date: 2006-11-15 |
Medline Journal Info:
|
Nlm Unique ID: 7802778 Medline TA: Artif Organs Country: UNITED STATES |
Other Details:
|
Languages: eng Pagination: 1110-20 Citation Subset: IM |
Affiliation:
|
Kanto Gakuin University, Yokohama, Japan. |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
Biomechanics Heart Valve Prosthesis* Hemolysis Laser-Doppler Flowmetry / methods* Mitral Valve* Models, Theoretical Pressure Pulsatile Flow Stress, Mechanical Thromboembolism |