Uniform vascular enhancement of lower-extremity artery on CT angiography using test-injection monitoring at the central level of the scan range: a simulation flow phantom study with clinical correlation.
MedLine Citation:
PMID:  20605732     Owner:  NLM     Status:  MEDLINE    
Abstract/OtherAbstract:
RATIONALE AND OBJECTIVES: To evaluate the efficacy of variable contrast injection durations and scanning delay determined by test injection analysis of computed tomography angiography (CTA) of peripheral arteries.
MATERIALS AND METHODS: We used a flow phantom that simulates the hemodynamics in a lower extremity artery. We set the flow rate at the pump to 2.0 or 5.0 L/minute. In protocol 1, we adopted a variable contrast injection duration based on the peak enhancement time of the test injection monitoring at the central level of the scan range. In protocol 2, we adopted a fixed contrast injection duration. The scanning delay was determined with a conventional bolus-tracking technique monitoring at the top of the scan range. Mean arterial attenuation and difference between the maximum and minimum attenuation values were calculated. To verify the phantom study results, clinical study, including 16 patients was performed under protocol 1.
RESULTS: The mean attenuation values under protocols 1 and 2 were comparable (563.6 Hounsfield units [HU] and 535.0 HU, respectively) at a pump flow rate of 2.0 L/minute; at 5.0 L/minute, they were 289.4 HU and 328.8 HU. The difference between the maximum and minimum attenuation values was smaller under protocol 1 than protocol 2 (76.8 HU vs. 184.9 HU) at a pump flow of 2.0 L/minute and also smaller under protocol 1 than protocol 2 (79.7 HU vs. 203.8 HU) at 5.0 L/minute. In clinical study, the mean attenuation value was 332.6 +/- 51.9 HU, and the difference between the maximum and minimum attenuation values was 55.1 +/- 24.4 HU.
CONCLUSION: The object-specific injection duration based on test injection at the central level of the scan range provides sufficient and constant vascular enhancement at CTA.
Authors:
Kie Shimizu; Daisuke Utsunomiya; Takeshi Nakaura; Kazuo Awai; Seitaro Oda; Yumi Yanaga; Yoshinori Funama; Toshinori Hirai; Masahiro Hashida; Yasuyuki Yamashita
Publication Detail:
Type:  Journal Article     Date:  2010-06-01
Journal Detail:
Title:  Academic radiology     Volume:  17     ISSN:  1878-4046     ISO Abbreviation:  Acad Radiol     Publication Date:  2010 Sep 
Date Detail:
Created Date:  2010-08-09     Completed Date:  2011-01-10     Revised Date:  -    
Medline Journal Info:
Nlm Unique ID:  9440159     Medline TA:  Acad Radiol     Country:  United States    
Other Details:
Languages:  eng     Pagination:  1153-7     Citation Subset:  IM    
Copyright Information:
Copyright 2010 AUR. Published by Elsevier Inc. All rights reserved.
Affiliation:
Central Radiology, Kumamoto University, Kumamoto-shi, Kumamoto, 860-8556, Japan.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Descriptor/Qualifier:
Angiography / methods*
Contrast Media / administration & dosage
Humans
Iopamidol / administration & dosage*,  diagnostic use*
Lower Extremity / blood supply*,  radiography*
Models, Biological
Peripheral Arterial Disease / radiography
Phantoms, Imaging
Radiographic Image Enhancement / methods*
Radiographic Image Interpretation, Computer-Assisted / methods
Reproducibility of Results
Sensitivity and Specificity
Tomography, X-Ray Computed / methods*
Chemical
Reg. No./Substance:
0/Contrast Media; 62883-00-5/Iopamidol

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine


Previous Document:  Frequency and Diagnostic Utility of Intralabyrinthine FLAIR Hyperintensity in the Evaluation of Inte...
Next Document:  The Degree of Stenosis on Cardiac Catheterization Compared to Calcified Coronary Segments on Multi-D...