We present the 10- to 17-year results of 112 computer-assisted design computer-assisted manufacture femoral components. The total hip replacements were performed between 1992 and 1998 in 111 patients, comprising 53 men and 58 women. Their mean age was 46.2 years (24.6 to 62.2) with a mean follow-up of 13 years (10 to 17). The mean Harris Hip Score improved from 42.4 (7 to 99) to 90.3 (38 to 100), the mean Oxford Hip Score from 43.1 (12 to 59) to 18.2 (12 to 51) and the mean Western Ontario MacMasters University Osteoarthritis Index score from 57.0 (7 to 96) to 11.9 (0 to 85). There was one revision due to failure of the acetabular component but no failures of the femoral component. There were no revisions for aseptic loosening. The worst-case survival in this cohort of custom femoral components at 13.2 years follow-up was 98.2% (95% confidence interval 95 to 99). Overall survival of this series of total hip replacements was 97.3% (95% confidence interval 95 to 99). These results are comparable with the best medium- to long-term results for femoral components used in primary total hip replacement with any means of fixation.
S K Muirhead-Allwood; N Sandiford; J A Skinner; J Hua; C Kabir; P S Walker
Publication Detail:
Type:  Evaluation Studies; Journal Article; Review    
Journal Detail:
Title:  The Journal of bone and joint surgery. British volume     Volume:  92     ISSN:  0301-620X     ISO Abbreviation:  J Bone Joint Surg Br     Publication Date:  2010 Aug 
Date Detail:
Created Date:  2010-08-02     Completed Date:  2010-08-30     Revised Date:  2010-11-10    
Medline Journal Info:
Nlm Unique ID:  0375355     Medline TA:  J Bone Joint Surg Br     Country:  England    
Other Details:
Languages:  eng     Pagination:  1079-84     Citation Subset:  AIM; IM    
The London Hip Unit, 4th Floor, 30 Devonshire Street, London W1G 6PU, UK.
Export Citation:
APA/MLA Format     Download EndNote     Download BibTex
MeSH Terms
Arthroplasty, Replacement, Hip / methods*
Coated Materials, Biocompatible
Computer-Aided Design*
Follow-Up Studies
Hip Joint / radiography
Hip Prosthesis*
Middle Aged
Osteoarthritis, Hip / surgery
Prosthesis Design / methods*
Prosthesis Failure
Survival Analysis
Treatment Outcome
Young Adult
Reg. No./Substance:
0/Coated Materials, Biocompatible; 1306-06-5/Durapatite

From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine

Previous Document:  Optimal acetabular orientation for hip resurfacing.
Next Document:  Total knee replacement in patients with end-stage haemophilic arthropathy: 25-YEAR RESULTS.