| Uncalibrated radial and femoral arterial pressure waveform analysis for continuous cardiac output measurement: an evaluation in cardiac surgery patients. | |
| | |
MedLine Citation:
|
PMID: 19700349 Owner: NLM Status: MEDLINE |
Abstract/OtherAbstract:
|
OBJECTIVES: Arterial pressure waveform analysis is a less invasive alternative to the pulmonary artery catheter for continuous cardiac output (CO) measurement. Uncalibrated and calibrated systems are actually available (ie, the FloTrac/Vigileo system [Edwards Lifesciences, Irvine, CA] and the PiCCOplus system [Pulsion Medical Systems, Munich, Germany]). According to the FloTrac/Vigileo manufacturer, reliable measurements can be performed using any existing arterial catheter. The aim of this study was to evaluate CO determined by the FloTrac/Vigileo system using a radial (FCO(radial)) and femoral arterial catheter (FCO(femoral)) as well as the PiCCOplus system (PCO). Intermittent pulmonary artery thermodilution (ICO) was used as primary reference technique. DESIGN: A prospective clinical study. SETTING: A teaching hospital, single center. PARTICIPANTS: Twenty-six cardiac surgery patients. INTERVENTIONS: Perioperative CO measurements. MEASUREMENTS AND MAIN RESULTS: CO was assessed at predefined measurement points. FCO(radial), FCO(femoral), and PCO were recorded after the induction of anesthesia, after sternotomy, at skin closure, after intensive care unit transfer, and during intensive care unit stay 12 and 24 hours after study initiation. ICO was determined as the mean of 3 bolus injections. Bland-Altman analysis revealed comparable mean bias and limits of agreement for FCO(radial), FCO(femoral), and PCO when compared with ICO. There was a decreased agreement for all devices in the postoperative period. However, a consistently close agreement was observed for the direct comparison between FCO(radial) and FCO(femoral). CONCLUSIONS: Performance of the FloTrac/Vigileo system via radial as well as femoral access and the PiCCOplus monitoring for cardiac output measurement were comparable when tested against intermittent thermodilution in cardiac surgery patients. |
| | |
Authors:
|
Christoph K Hofer; Danny Button; Luc Weibel; Michele Genoni; Andreas Zollinger |
Related Documents
:
|
16425159 - External diameters of the crural arteries in patients with chronic critical limb ischae... 6449899 - An arterial medial arm flap--vascular anatomy and clinical applications. 8877199 - Doppler ultrasonography as a pre-operative aid to base the forearm flap on the radial o... 1561269 - A free arterialized venous loop flap. 9894309 - The reversed-flow medio-distal fasciocutaneous island thigh flap: anatomic basis and cl... 19718239 - Superficial vein thrombosis 21 cm distant from the venous anastomosis of a free flap ca... 61709 - Ionic composition and cardiotoxicity of dimeric contrast media at injection into the co... 17426219 - Pleural tuberculosis in the united states: incidence and drug resistance. 21183749 - Discriminating carotid atherosclerotic lesion severity by luminal stenosis and plaque b... |
Publication Detail:
|
Type: Comparative Study; Evaluation Studies; Journal Article; Research Support, Non-U.S. Gov't Date: 2009-08-22 |
Journal Detail:
|
Title: Journal of cardiothoracic and vascular anesthesia Volume: 24 ISSN: 1532-8422 ISO Abbreviation: J. Cardiothorac. Vasc. Anesth. Publication Date: 2010 Apr |
Date Detail:
|
Created Date: 2010-04-05 Completed Date: 2010-11-08 Revised Date: - |
Medline Journal Info:
|
Nlm Unique ID: 9110208 Medline TA: J Cardiothorac Vasc Anesth Country: United States |
Other Details:
|
Languages: eng Pagination: 257-64 Citation Subset: IM |
Copyright Information:
|
Copyright (c) 2010 Elsevier Inc. All rights reserved. |
Affiliation:
|
Institute of Anesthesiology and Intensive Care Medicine, Triemli City Hospital, Zurich, Switzerland. christoph.hofer@triemli.stzh.ch |
Export Citation:
|
APA/MLA Format Download EndNote Download BibTex |
| MeSH Terms | |
Descriptor/Qualifier:
|
Aged Blood Pressure / physiology* Calibration / standards Cardiac Output / physiology* Cardiac Surgical Procedures / methods, standards* Catheterization, Peripheral / methods, standards Female Femoral Artery / physiology* Humans Male Middle Aged Monitoring, Intraoperative / methods, standards* Prospective Studies Radial Artery / physiology* Thermodilution / methods, standards |
From MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine
Previous Document: Validity of an Arterial Pressure Waveform Analysis Device: Does the Puncture Site Play a Role in the...
Next Document: Reducing mortality in cardiac surgery with levosimendan: a meta-analysis of randomized controlled tr...